← Product Code [JIX](/productcode/JIX) · K963533

# SIGMA DIAGNOSTIC CX -3 CALIBRATION STANDARD LEVEL 1 (K963533)

_Sigma Diagnostics, Inc. · JIX · Oct 29, 1996 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K963533

## Device Facts

- **Applicant:** Sigma Diagnostics, Inc.
- **Product Code:** [JIX](/productcode/JIX.md)
- **Decision Date:** Oct 29, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

Sigma Diagnostics CX®3 Calibration Standard Level 1 is used to calibrate the Beckman SYNCHRON CX®3 System. This reagent is used for the quantitative determination of BUN/urea, calcium, and total carbon dioxide in serum or plasma; of creatinine, potassium, and sodium in serum, plasma, or urine; and of chloride and glucose in serum, plasma, urine, or cerebrospinal fluid. Calibration Standard Level 1 should be used with CX®3 Calibration Standard Level 2.

## Device Story

Calibration reagent for Beckman SYNCHRON CX®3 clinical chemistry analyzer; contains defined concentrations of BUN/urea, calcium, total CO2, creatinine, potassium, sodium, chloride, and glucose; used in clinical laboratory settings by trained personnel; ensures system accuracy for quantitative diagnostic testing; functions as reference standard for instrument calibration; supports clinical decision-making by maintaining analytical precision of patient sample measurements.

## Clinical Evidence

Bench testing only; no clinical data provided.

## Technological Characteristics

Liquid reagent containing specified concentrations of analytes (Calcium, Chloride, Creatinine, Glucose, Potassium, Sodium, Urea, BUN); intended for use with Beckman SYNCHRON CX®3 System; non-reactive components included; no specific materials or software algorithms described.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Beckman Calibration Standard 1 (Part No. 443360)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K963533
OCT 29 1996

510(K) NOTIFICATION

Sigma Diagnostics
545 South Ewing Avenue
St. Louis, MO 63103

CX®3 Calibration Standard, Level 1
Procedure Number C7308
August 31, 1996

# SUMMARY OF SAFETY AND EFFECTIVENESS

Sigma Diagnostics CX®3 Calibration Standard Level 1 is used to calibrate the Beckman SYNCHRON CX®3 System. This reagent is used for the quantitative determination of BUN/urea, calcium, and total carbon dioxide in serum or plasma; of creatinine, potassium, and sodium in serum, plasma, or urine; and of chloride and glucose in serum, plasma, urine, or cerebrospinal fluid. Calibration Standard Level 1 should be used with CX®3 Calibration Standard Level 2.

CX®3 Calibration Standard Level 1 contains the following analytes and other non-reactive components:

|  ANALYTE | CONVENTIONAL UNITS | SI UNITS  |
| --- | --- | --- |
|  Calcium | 10.0 mg/dL | 2.495 mmol/L  |
|  Chloride | 109 mmol/L | 109 mmol/L  |
|  Creatinine | 5.0 mg/dL | 442 umol/L  |
|  Glucose | 150 mg/dL | 8.34 mmol/L  |
|  Potassium | 4.0 mmol/L | 4.0 mmol/L  |
|  Sodium | 150 mmol/L | 150 mmol/L  |
|  Urea | 107 mg/dL | 17.8 mmol/L  |
|  BUN (Urea Nitrogen) | 50 mg/dL | 17.8 mmol/L  |

The safety and effectiveness of Sigma Diagnostics CX®3 Calibration Standard, Level 1, Procedure Number C7308, are demonstrated by its substantial equivalency to Beckman Calibration Standard 1, Part No. 443360.

---

**Source:** [https://fda-staging.innolitics.com/device/K963533](https://fda-staging.innolitics.com/device/K963533)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).