SIGMA DIAGNOSTIC CX -3 CALIBRATION STANDARD LEVEL 1

K963533 · Sigma Diagnostics, Inc. · JIX · Oct 29, 1996 · Clinical Chemistry

Device Facts

Record IDK963533
Device NameSIGMA DIAGNOSTIC CX -3 CALIBRATION STANDARD LEVEL 1
ApplicantSigma Diagnostics, Inc.
Product CodeJIX · Clinical Chemistry
Decision DateOct 29, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

Sigma Diagnostics CX®3 Calibration Standard Level 1 is used to calibrate the Beckman SYNCHRON CX®3 System. This reagent is used for the quantitative determination of BUN/urea, calcium, and total carbon dioxide in serum or plasma; of creatinine, potassium, and sodium in serum, plasma, or urine; and of chloride and glucose in serum, plasma, urine, or cerebrospinal fluid. Calibration Standard Level 1 should be used with CX®3 Calibration Standard Level 2.

Device Story

Calibration reagent for Beckman SYNCHRON CX®3 clinical chemistry analyzer; contains defined concentrations of BUN/urea, calcium, total CO2, creatinine, potassium, sodium, chloride, and glucose; used in clinical laboratory settings by trained personnel; ensures system accuracy for quantitative diagnostic testing; functions as reference standard for instrument calibration; supports clinical decision-making by maintaining analytical precision of patient sample measurements.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Liquid reagent containing specified concentrations of analytes (Calcium, Chloride, Creatinine, Glucose, Potassium, Sodium, Urea, BUN); intended for use with Beckman SYNCHRON CX®3 System; non-reactive components included; no specific materials or software algorithms described.

Indications for Use

Indicated for calibration of the Beckman SYNCHRON CX®3 System for quantitative determination of BUN/urea, calcium, total CO2, creatinine, potassium, sodium, chloride, and glucose in serum, plasma, urine, or cerebrospinal fluid.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} K963533 OCT 29 1996 510(K) NOTIFICATION Sigma Diagnostics 545 South Ewing Avenue St. Louis, MO 63103 CX®3 Calibration Standard, Level 1 Procedure Number C7308 August 31, 1996 # SUMMARY OF SAFETY AND EFFECTIVENESS Sigma Diagnostics CX®3 Calibration Standard Level 1 is used to calibrate the Beckman SYNCHRON CX®3 System. This reagent is used for the quantitative determination of BUN/urea, calcium, and total carbon dioxide in serum or plasma; of creatinine, potassium, and sodium in serum, plasma, or urine; and of chloride and glucose in serum, plasma, urine, or cerebrospinal fluid. Calibration Standard Level 1 should be used with CX®3 Calibration Standard Level 2. CX®3 Calibration Standard Level 1 contains the following analytes and other non-reactive components: | ANALYTE | CONVENTIONAL UNITS | SI UNITS | | --- | --- | --- | | Calcium | 10.0 mg/dL | 2.495 mmol/L | | Chloride | 109 mmol/L | 109 mmol/L | | Creatinine | 5.0 mg/dL | 442 umol/L | | Glucose | 150 mg/dL | 8.34 mmol/L | | Potassium | 4.0 mmol/L | 4.0 mmol/L | | Sodium | 150 mmol/L | 150 mmol/L | | Urea | 107 mg/dL | 17.8 mmol/L | | BUN (Urea Nitrogen) | 50 mg/dL | 17.8 mmol/L | The safety and effectiveness of Sigma Diagnostics CX®3 Calibration Standard, Level 1, Procedure Number C7308, are demonstrated by its substantial equivalency to Beckman Calibration Standard 1, Part No. 443360.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...