DOUBLE SUPPORT PONCHO/SLIDER PONCHO

K963447 · Skil-Care Corp. · FMQ · Oct 7, 1996 · General Hospital

Device Facts

Record IDK963447
Device NameDOUBLE SUPPORT PONCHO/SLIDER PONCHO
ApplicantSkil-Care Corp.
Product CodeFMQ · General Hospital
Decision DateOct 7, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6760
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The Double Support Poncho is used for restraint and to provide upper body posture support. The Slider Poncho provides restraint, posture support, and slider control.

Device Story

Poncho-style upper body support devices for wheelchair users; constructed from woven polyester or polyester mesh; secured via chest and waist straps to wheelchair kickspurs. Slider Poncho includes additional crotch pad with polyurethane foam lining and strap secured to wheelchair cross brace to prevent sliding. Operated by healthcare staff in nursing homes or clinical facilities. Provides physical restraint and posture support; intended to assist in medical treatment under appropriate supervision. Benefits include improved patient positioning and prevention of sliding/falls from wheelchairs.

Clinical Evidence

No clinical data provided. Devices are currently marketed and exempt from 510(k) procedures; safety established through 18 years of commercial distribution and compliance with HCFA regulations regarding restraint use.

Technological Characteristics

Materials: Woven polyester or polyester mesh; 1 1/8-inch polyester webbing straps; 1/4-inch polyurethane foam padding. Form factor: Poncho-style garment. Mechanical restraint/support principle. No energy source, software, or electronic components.

Indications for Use

Indicated for patients requiring upper body posture support and restraint while seated in a wheelchair. Slider Poncho indicated for patients requiring additional anti-sliding control. Contraindicated for use for discipline or convenience.

Regulatory Classification

Identification

A protective restraint is a device, including but not limited to a wristlet, anklet, vest, mitt, straight jacket, body/limb holder, or other type of strap, that is intended for medical purposes and that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others.

Predicate Devices

Submission Summary (Full Text)

{0} OCT - 7 1996 K963447 # TAB F: 510(k) Summary of Safety and Effectiveness Name, address, phone and fax numbers for person submitting the 510(k) notification: Arnold Silverman, President Skil-Care Corporation 167 Saw Mill River Road Yonkers, NY 10701 Phone: 1-914-963-2040 Fax: 1-914-963-2567 Contact person: Arnold Silverman Date summary was prepared: August 27, 1996 Device name: Trade name: Double Support Poncho, Slider Poncho Common name: Same Classification name: Wheelchair Accessory Predicate device: Double Support Poncho and Slider Poncho marketed by Skil-Care Corporation. Device Description: Double Support Poncho: A poncho-style, upper body device made from either woven polyester or polyester mesh. A chest strap is used to provide upper body posture support. A waist strap secures the device the device to wheelchair kickspurs. All straps are 1 1/8-inch-wide polyester webbing. The edges of the garment are finished with a bias cut binding, the color of which corresponds to a size chart included on the package insert. Device is intended for wheelchair use only. Slider Poncho: The Slider Poncho has the same configuration as the above with the addition of a crotch pad sewn to the bottom center front. The pad is lined with 1/4 inch polyurethane foam and has a 1 and 1 1/8 inch strap sewn on that is either placed directly under the patient or under the wheelchair seat and secured to the wheelchair cross brace to prevent the patient from sliding down. {1} K963447 ## Indications for use: The Double Support Poncho is used for restraint and to provide upper body posture support. The Slider Poncho provides restraint, posture support, and slider control. ## Comparative information: The device (devices) used for comparative purposes is (are) currently marketed as described in this submission. Device (devices) is (are): Double Support Poncho and Slider Poncho. These devices are currently exempt from 510(k) Premarket Notification Procedures and Good Manufacturing Practice Regulations and are legally marketed by Skil-Care Corporation as of the date of this submission. Skil-Care Corporation has been marketing and commercially distributing these devices for approximately 18 years. The difference from our currently marketed devices are that the labeling will be changed to incorporate many of the suggestions in FDA’s draft document, "Guidance on the Content of Premarket Notification [510(k)] Submissions for Protective Restraints. The use of all patient restraints in nursing homes are subject to Health Care Financing Administration’s Regulations which prohibit the use of any restraint, physical or chemical, imposed for the purpose of discipline or convenience. Further, most health care facilities are accredited. HCFA rules governing appropriate use and accreditation standards for device use and personnel training provide the control necessary to ensure that the devices are used correctly. The application of these standards along with public awareness and health care provider training have contributed significantly to ensuring that the least restrictive restraint is used, that restraints are used only when needed for proper medical treatment, and that their use is under appropriate supervision. 28
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...