VESSEL DILATOR

K963388 · Maxxim Medical · DRE · Mar 11, 1997 · Cardiovascular

Device Facts

Record IDK963388
Device NameVESSEL DILATOR
ApplicantMaxxim Medical
Product CodeDRE · Cardiovascular
Decision DateMar 11, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1310
Device ClassClass 2

Indications for Use

To enlarge the opening in a vessel in preparation of percutaneous entry of a catheter.

Device Story

Vessel dilator used over guide wire to enlarge vessel opening for percutaneous catheter entry; used independently or with sheath introducer. Device comprises plastic hub with female luer lock and thermally tapered plastic cannula. Available in 4F-15F sizes. Supplied sterile/disposable or as component in kits/trays. Used by clinicians in clinical settings to facilitate vascular access; improves procedural efficiency by preparing vessel site for catheter insertion.

Clinical Evidence

Bench testing only. Biocompatibility established via cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity, and hemocompatibility testing. Meets USP XXIII physicochemical requirements for plastics. Tensile strength testing performed.

Technological Characteristics

Plastic hub and cannula; distal end thermally tapered. Sizes 4F-15F. Female luer lock. Biocompatibility per USP XXIII. Sterilization via standard methods for disposable medical devices.

Indications for Use

Indicated for patients requiring percutaneous catheterization where vessel dilation is necessary to facilitate entry.

Regulatory Classification

Identification

A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.

Predicate Devices

Submission Summary (Full Text)

{0} K963388 MAR 11 1997 # 510(k) SUMMARY ## I. ADMINISTRATIVE Submitter: Maxxim Medical, Inc, Argon Division 1445 Flat Creek Road Athens, TX 75751 (903) 675-9321 Contact Person: Eddie Monroe Date of Preparation: August 26, 1996 ## II. DEVICE NAME Proprietary Name: None Common Name: Vessel Dilator Classification Name: Vessel Dilator for percutaneous catheterization ## III. PREDICATE DEVICE Vessel Dilator component of Introducer for PTCA Guiding Catheters; K891022; Argon Medical. ## IV. DEVICE DESCRIPTION The Vessel Dilator is a device used over a guide wire to dilate the opening in a vessel in preparation of percutaneous entry of a catheter. The device may be used separately or in conjunction with a sheath introducer. The device consists of a plastic hub attached to a plastic cannula. The distal end of the cannula is thermally tapered to assist in dilation. The hub proximal end has a female luer lock configuration. An optional hub collar may be used to attach the dilator to an introducer. The device is available in sizes 4F through 15F and is supplied as a sterile, disposable single put up in sealed pouches, in non-sterile bulk quantities, or included as a component in catheter introducer kits, trays, or packs. Biocompatibility of the device has been established by cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity and hemocompatibility tests. The device meets the requirements of the USP XXIII physicochemical tests for plastics. ## V. INTENDED USE To enlarge the opening in a vessel in preparation of percutaneous entry of a catheter. Premarket Notification. Vessel Dilator 10 {1} 11 Premarket Notification. Vessel Dilator VI. COMPARISON TO PREDICATE DEVICE The Vessel Dilator is identical in design, function and intended use, and similar in composition, to the Vessel Dilator component of the Introducer for PTCA Guiding Catheters (Argon Medical; K891022). Results of performance testing indicate that the Vessel Dilator has improved tensile strength characteristics compared to the predicate device. Accordingly, Maxxim Medical, Inc., Argon Division, concluded that the Vessel Dilator is safe and effective for its intended use and performs at least as well as the legally marketed predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...