MEDLITE ND:YAG LASER SYSTEM: NEW DYE ACCESSORY HANDPIECE

K963340 · Continuum Biomedical, Inc. · GEX · May 12, 1997 · General, Plastic Surgery

Device Facts

Record IDK963340
Device NameMEDLITE ND:YAG LASER SYSTEM: NEW DYE ACCESSORY HANDPIECE
ApplicantContinuum Biomedical, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateMay 12, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

For use in dermatology for the removal of tattoos and the treatment of vascular and pigmented lesions.

Device Story

Dye Accessory Handpiece for Medlite Nd:YAG Laser System; consists of handpiece with optical cavity/focusing optics, plastic dye module, power meter, and safety goggles. Device modifies laser output wavelengths to increase efficiency for tattoo removal and treatment of vascular/pigmented lesions. Used in clinical dermatology settings by physicians. Output allows targeted laser energy delivery to skin lesions; enhances clinical efficacy for dermatological procedures.

Clinical Evidence

No clinical data provided.

Technological Characteristics

Dye accessory handpiece; includes optical cavity, focusing optics, plastic dye module, and power meter. Modifies Nd:YAG laser output wavelengths. Class II surgical laser system (21 CFR 878.48).

Indications for Use

Indicated for patients requiring removal of tattoos or treatment of vascular and pigmented lesions in a dermatology setting.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0} 510(K) PREMARKET NOTIFICATION Medlite Nd:YAG Laser System Continuum Biomedical, Inc. August 23, 1996 MAY 12 1997 K 963340 # 510(k) Summary | Submitter: | Continuum Biomedical, Inc. 2456 Armstrong Street Livermore, CA 94550-7617 Phone: 510-606-3750 Fax: 510-371-1015 | | --- | --- | | Contact: | Robert S. Anderson, Ph. D. President | | Date Summary Prepared: | August 23, 1996 | | Device Trade Name: | Medlite™ Nd:YAG Laser System: New Dye Accessory Handpiece | | Common Name: | Dermatology laser system | | Classification Name: | Instrument, surgical, powered, laser 79-GEX 21 CFR 878.48 | | Equivalent Device: | Medlite™ Nd:YAG Laser System (K955131, K952636, K940607, K940636, K922935, K922577, K915497) | | Device Description: | The new Dye Accessory Handpiece for the Medlite™ Nd:YAG Laser System is consists of four primary parts: the handpiece which contains a short optical cavity and focusing optics, the plastic dye module, the commercial power meter and appropriate laser safety goggles. | | Intended Use: | For use in dermatology for the removal of tattoos and the treatment of vascular and pigmented lesions. | | Comparison: | The new Dye Accessory Handpiece allows the Medlite™ Nd:YAG Laser System to deliver slightly altered wavelengths for increased efficiency in treating tattoos, vascular and pigmented lesions. | | Nonclinical Performance Data: | none | | Clinical Performance Data: | none | | Conclusion: | The Medlite™ Nd:YAG Laser System is substantially equivalent to other existing surgical laser systems in commercial distribution for the treatment of tattoos, vascular and pigmented lesions. | | Additional Information: | None requested at this time. | c:\conbiol\medlite\510kcolor.doc:13
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...