K963175 · Glycar, Inc. · FTM · Sep 23, 1996 · General, Plastic Surgery
Device Facts
Record ID
K963175
Device Name
GLYCAR TISSUE REPAIR PATCH, BIO-PATCH, URO-SLING
Applicant
Glycar, Inc.
Product Code
FTM · General, Plastic Surgery
Decision Date
Sep 23, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Glycar Inc., Tissue Repair Patch family of products have the same intended uses as the predicate device, the Bio-Vascular Inc., Peri-Guard™ family of products, which are all tissue patches made from the same glutaraldehyde-treated bovine pericardium. Already approved intended uses for the Glycar Tissue Repair Patch family of glutaraldehyde-tanned bovine pericardial tissue patches: The Glycar Tissue Repair Patch (K942911) was deemed substantially equivalent to the Peri-Guard™ bovine pericardial tissue patch and received approval for U.S. marketing for the intended use of hernia and other intra-abdominal soft tissue defect or deficiency repair in December, 1994. An addition to the Glycar Tissue Repair Patch family of products, the Glycar Staple Strip (K954665) was deemed substantially equivalent to the Peri-Strips™ bovine pericardial surgical staple bolsters and received approval for U.S. marketing for the intended uses of surgical stapling of lung tissue, gastric staplings, rectal and vaginal prolapse, urethral sling, reconstruction of the pelvic floor, and hernias or defects of the diaphragm, thoracic and abdominal wall in October, 1995. The Glycar Staple Strip uses the identical material as the Glycar Tissue Repair Patch with the addition of a synthetic suture material to assist placement of the device on the surgical stapling arm.
Device Story
Glycar Tissue Repair Patch consists of glutaraldehyde-treated bovine pericardium; used for soft tissue repair, reinforcement, and reconstruction. Device provides durable patch for surgical applications including hernia repair, pelvic floor reconstruction, and staple bolsters. Implanted by surgeons in clinical settings. Material is strong, pliable, and biocompatible. Proprietary anti-inflammation treatment (XX) applied during manufacturing. Sterilized via formaldehyde and propylene oxide to SAL 10^-6. Sourced from BSE-certified cattle in South Africa. Device functions as a mechanical scaffold for tissue repair.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, including pyrogenicity testing and breaking strength comparisons to the predicate device.
Technological Characteristics
Material: Glutaraldehyde-treated bovine pericardium. Form factor: Tissue patch/strip. Sterilization: Formaldehyde and propylene oxide (SAL 10^-6). Manufacturing includes proprietary anti-inflammation treatment (XX).
Indications for Use
Indicated for repair, replacement, or reinforcement of natural soft tissue, including hernia repair (inguinal, paracolostomy, ventral, paraumbilical, scrotal, femoral, diaphragm, thoracic, abdominal wall), pericardial closures, urethral slings, pelvic floor reconstruction, rectal and vaginal prolapse repair, and use as suture/staple bolsters or pledgets.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
Tissue Guard product line, including Supple Peri-Guard product line (K923657)
{0}
510 (K) SUMMARY - GLYCAR TISSUE REPAIR PATCH
K963175
GLYCAR
# 510(K) SUMMARY
SEP 23 1996
## 1. Contact Person
Dr. Dirk Frater
4537 Belclaire Ave.
Dallas, TX 75205
Phone: (214) 526-5412
## 2. Regulatory Consultant
Bertram Hudson
Shotwell & Carr Inc.
Suite 100, 3003 LBJ Freeway
Dallas, TX 75234
## 3. Device Name
a) Trade Name
Glycar Tissue Repair Patch
Bio-Patch trademark pending
Uro-Sling trademark pending
b) Common Name
Biological Tissue Patch
## 4. F.D.A. Device Classification
Class II Device
## 5. Predicate Device
The predicate device is the Tissue Guard™ product line, including Supple Peri-Guard™ product line (K923657), and is a tissue patch made from glutaraldehyde-treated bovine pericardium. The device is produced by Bio-Vascular Inc., 2670 Patton Rd., St. Paul, MN 55113, (800) 255-4018.
For copies of Peri-Guard™ labels, see Figures 3 - 4 in the Addenda under Predicate Device Labeling.
## 6. Product Description
a) Candidate Device Composition
The Glycar Tissue Repair Patches are made of the same material as the predicate device, the Peri-Guard™
Aug.7, 1996
{1}
510 (K) SUMMARY - GLYCAR TISSUE REPAIR PATCH
GLYCAR
tissue patches, which are made from glutaraldehyde-treated bovine pericardium.
See Comparison Chart, Chart 1, in Addenda under Substantial Equivalence.
## b) Material Attributes
Glutaraldehyde-treated bovine pericardium is a strong, pliable, biocompatible material which is easy to handle surgically.
Glycar's glutaraldehyde-treated pericardium possesses tissue strength equaling that of the predicate device. It provides for a durable repair, replacement or reinforcement of natural soft tissue.
See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence and Breaking Strength under Supporting Data
## c) Pyrogenicity
Glycar glutaraldehyde-treated bovine pericardium has been found to have negative pyrogenicity.
See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence and Pyrogenicity under Supporting Data.
## d) Manufacturing Process
The manufacturing process for the *Glycar Tissue Repair Patches* involves the harvesting of bovine pericardium, fixation of the pericardium in glutaraldehyde, inspection and cleaning of the pericardium, selection of the pericardium for use in strip production, additional tanning of the bovine pericardium, and then cutting of the glutaraldehyde-treated bovine pericardium for the designated strip size, followed by sterilization in formaldehyde, followed then by treatment with Glycar's proprietary anti-inflammation treatment, XX. See in Proprietary Manufacturing Material in Addenda under Confidential Information.
Sterility is maintained by using aseptic fill under Clean Room 100 conditions, during transfer to anti-inflammation treatment and to the final storage medium, 2% propylene oxide in water.
See Outline of Manufacturing Process in Addenda under Manufacturing Information and Sterilization Information.
Aug.7, 1996
ii
{2}
510 (K) SUMMARY - GLYCAR TISSUE REPAIR PATCH
GLYCAR
e) Product Sizes
Sizes and shapes will be specific for product line reflecting different indications as appropriate for the use.
7. Intended Use of Product
a) Candidate Device Intended Use
The Glycar Inc., Tissue Repair Patch family of products have the same intended uses as the predicate device, the Bio-Vascular Inc., Peri-Guard™ family of products, which are all tissue patches made from the same glutaraldehyde-treated bovine pericardium.
Already approved intended uses for the Glycar Tissue Repair Patch family of glutaraldehyde-tanned bovine pericardial tissue patches:
The Glycar Tissue Repair Patch (K942911) was deemed substantially equivalent to the Peri-Guard™ bovine pericardial tissue patch and received approval for U.S. marketing for the intended use of hernia and other intra-abdominal soft tissue defect or deficiency repair in December, 1994.
An addition to the Glycar Tissue Repair Patch family of products, the Glycar Staple Strip (K954665) was deemed substantially equivalent to the Peri-Strips™ bovine pericardial surgical staple bolsters and received approval for U.S. marketing for the intended uses of surgical stapling of lung tissue, gastric staplings, rectal and vaginal prolapse, urethral sling, reconstruction of the pelvic floor, and hernias or defects of the diaphragm, thoracic and abdominal wall in October, 1995.
The Glycar Staple Strip uses the identical material as the Glycar Tissue Repair Patch with the addition of a synthetic suture material to assist placement of the device on the surgical stapling arm.
See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence.
b) Predicate Device Intended Use
The predicate device, the Peri-Guard™ family of glutaraldehyde-treated bovine pericardium tissue patch has been used for pericardial closures, the repair of abdominal and thoracic wall defects, inguinal, paracolostomy, ventral, paraumbilical, scrotal, femoral
Aug.7, 1996
iii
{3}
510 (K) SUMMARY - GLYCAR TISSUE REPAIR PATCH
GLYCAR
and other hernias, urethral slings, reconstruction of pelvic floor, repair of rectal and vaginal prolapse, suture and staple bolsters, pledgets and slings, and other soft tissue repair, reinforcement, and reconstruction without any significant reported patient morbidity or mortality attributed to material failures.
## 8. Sterilization
a) Method
The device is sterilized using exposure to formaldehyde and propylene oxide under environmental control in a class 100 Clean Room environment. See Outline of Sterilization Method in Addenda under Sterilization Information.
b) Sterility Assurance Level
The Glycar Tissue Repair Patches are labeled as sterile. They are sterilized to an SAL of 10⁻⁶ organisms. See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence and see Validation of Sterilization Processes under Sterilization Information.
c) Bovine Spongiform Encephalopathy
Pericardium used for the raw material for these devices is obtained from cattle from government approved abbatoirs in Transvaal, S. Africa.
The cattle of the Republic of South Africa are certified free from B.S.E. (bovine spongiform encephalopathy)
See Figure 5 in Addenda, Certificate of B.S.E.-free Status
## 9. Substantial Equivalence
a) Comparison with Predicate Device
The Glycar Tissue Repair Patches are substantially equivalent to the Peri-Guard™ family of glutaraldehyde-treated bovine pericardium in that they are:
i) manufactured from the same material, glutaraldehyde-treated bovine pericardium,
ii) have the same intended uses,
iii) are sterilized to a minimum assurance against contamination of 10⁻⁶ organisms (Sterility Assurance Level, SAL = 10⁻⁶), and
Aug.7, 1996
iv
{4}
510 (K) SUMMARY - GLYCAR TISSUE REPAIR PATCHI
GLYCAR
iv) have other performance features equal to those of the predicate device.
*See Comparison Chart, Chart 1, in Addenda under Substantial Equivalence and see Supporting Data*
## b) Technological Differences
### i) Implanted Material- Bovine Pericardium
There is no difference in the implanted material used for the *GlycarTissue Repair Patches* and the predicate device, Bio-Vascular Supple Peri-Guard™. Both devices are manufactured from gluraldehyde-treated bovine pericardium. See *Comparison Chart, Chart 1, in Addenda under Substantial Equivalence*.
### ii) Sterilization Method
There is no difference in the chemical sterilization method used for Glycar Tissue Repair Patches versus that of the predicate device.
*See Supporting Data in Addenda Under Sleeve Configuration*
### iii) Manufacturing Material
One difference in the manufacturing process which Glycar Inc. considers proprietary is the use of the manufacturing material, XX.
Glycar Inc. has demonstrated no adverse effects on safety or efficacy of the use of this manufacturing material. See *Proprietary Manufacturing Material in Addenda under Confidential Information*.
Aug.7, 1996
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.