ELECSYS CALCHECK FT3

K963163 · Boehringer Mannheim Corp. · JJX · Aug 27, 1996 · Clinical Chemistry

Device Facts

Record IDK963163
Device NameELECSYS CALCHECK FT3
ApplicantBoehringer Mannheim Corp.
Product CodeJJX · Clinical Chemistry
Decision DateAug 27, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

Elecsys CalCheck FT 3 is intended for use in the quantitative verification of the calibration curve established by the Elecsys FT 3 reagents and calibrators on Elecsys 1010 or 2010 immunoassay analyzers.

Device Story

Elecsys CalCheck FT 3 consists of three-level, lyophilized, human-based materials; used for calibration verification of FT 3 assays on Elecsys 1010/2010 immunoassay analyzers. User reconstitutes material, performs triplicate assay, and compares results against target values to verify calibration curve integrity. Used in clinical laboratory settings by laboratory personnel. Output provides verification of analytical system performance; assists in ensuring accurate patient test results.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Lyophilized, human-based material; three-level set; single analyte; intended for use on Elecsys 1010/2010 immunoassay analyzers.

Indications for Use

Indicated for quantitative verification of calibration curves for FT 3 reagents and calibrators on Elecsys 1010 or 2010 immunoassay analyzers.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Submission Summary (Full Text)

{0} K963163 AUG 27 1996 # 510(k) Summary ## Elecsys® CalCheck™ FT 3 | Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. | | --- | --- | | 1) Submitter name, address, contact | Boehringer Mannheim Corporation 9115 Hague Rd. Indianapolis, IN 46250 (317) 845-2000 Contact Person: John D. Stevens Date Prepared: August 12, 1996 | | 2) Device name | Proprietary name: Elecsys CalCheck FT 3 Common name: Calibration Verification Material Classification name: Single (specified) analyte controls (assayed + unassayed) | | 3) Predicate device | We claim substantial equivalence to the Tosoh Medic AIA-Pack HCG Calibration Verification Test Set, K924862. The intended use for both products is to verify the calibration curve of automated immunoassay analyzers | | 4) Device Description | The Elecsys CalCheck FT 3 is a three level single analyte set of lyophilized, human based materials. After reconstitution they are assayed in triplicate and the results are compared to the target values. | | 5) Intended use | Elecsys CalCheck FT 3 is intended for use in the quantitative verification of the calibration curve established by the Elecsys FT 3 reagents and calibrators on Elecsys 1010 or 2010 immunoassay analyzers. | {1} 12 # 510(k) Summary, Elecsys® CalCheck™ FT 3, Continued ## 6) Comparison to predicate device The Boehringer Mannheim Elecsys CalCheck FT 3 is substantially equivalent to other products in commercial distribution intended for similar use. Most notably it is substantially equivalent to the currently marketed Tosoh Medics AIA-Pack HCG Calibration Verification Test Set, K924862. The intended use of both the Elecsys CalCheck FT 3 and the predicate device are the same in that they are intended to be used for the calibration verification for their labeled analytes on automated immunoassay analyzers.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...