K963122 · Unitron Industries, Inc. · ESD · Sep 6, 1996 · Ear, Nose, Throat

Device Facts

Record IDK963122
Device NameUNITRON MODEL ICON KR
ApplicantUnitron Industries, Inc.
Product CodeESD · Ear, Nose, Throat
Decision DateSep 6, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3300
Device ClassClass 1
AttributesTherapeutic

Indications for Use

To amplify and transmit sound via air conduction to the ear.

Device Story

Behind-the-ear hearing aid; amplifies and transmits sound via air conduction. Features K-Amp circuit with Wide Dynamic Range compression; Class D circuits for battery efficiency; telecoil with pre-amplifier. User-adjustable controls include Low Cut Tone, High Cut Tone, Threshold Knee-point, and volume. Powered by size 13 battery. Operated by patient; adjustment dictated by individual audiogram. Benefits include improved sound perception for hearing-impaired users.

Clinical Evidence

Bench testing only; technical specifications comply with ANSI Standard S 3.22-1987.

Technological Characteristics

Behind-the-ear form factor; K-Amp circuit with Wide Dynamic Range compression; Class D output stage; telecoil with pre-amplifier; size 13 battery power; ANSI S 3.22-1987 compliant.

Indications for Use

Indicated for individuals requiring sound amplification via air conduction to the ear, adjusted based on individual audiogram.

Regulatory Classification

Identification

An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.

Submission Summary (Full Text)

{0} K963122 UNITRON INDUSTRIES INC. SEP - 6 1996 # A Summary of Safety and Effectiveness Information Name of Device: Unitron Industries, ICON KR Type of Device: Behind-the-ear hearing aid. Substantially equivalent to other behind-the-ear hearing aids Intended Use: To amplify and transmit sound via air conduction to the ear. Features: Three fitter controls - Low Cut Tone, High Cut Tone, Threshold Knee-point. K-Amp circuit with Wide Dynamic Range compression. Class D circuits for excellent battery life. Telecoil with pre-amplifier. Assembly: Assembled from standard components that are used by Unitron and other hearing aid manufacturers Technical Characteristics: Technical specifications comply with ANSI Standard S 3.22-1987 Fit: Hearing aid adjustment dictates by individual audiogram Controls: Volume control, similar to other devices. Switch for selecting microphone, telecoil and off. Power: Standard hearing aid battery - size 13. A user's guide and other information is supplied with each hearing aid. 3555 Walnut Street • P.O. Box 5010 • Post Huron, Michigan 48061-5010 • Telephone (810) 982-0166 1-800-521-5400 Fax (810) 987-2011
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...