K963065 · Shantou Institute of Ultrasonic Instuments · IYO · May 29, 1997 · Radiology
Device Facts
Record ID
K963065
Device Name
CTS-285 WITH EZU-PL21 AND EZU-PC3B TRANSDUCERS
Applicant
Shantou Institute of Ultrasonic Instuments
Product Code
IYO · Radiology
Decision Date
May 29, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
Ultrasonic pulsed echo imaging and measurement for fetal, abdominal, and adult cardiac imaging as well as adult cardiac M-mode.
Device Story
CTS-285 is a portable diagnostic ultrasound system using electronic linear and convex scanning transducers. It processes ultrasonic pulsed echo signals via a built-in digital scan converter to produce B-mode and M-mode tomographic images on a video monitor. Operated by clinicians in clinical settings, the device allows for measurements of distance, area, circumference, time interval, velocity, depth, and heart rate. Output is visualized on a 5.5-inch monitor, assisting clinicians in diagnostic assessment of fetal, abdominal, and adult cardiac structures. The system is designed for portability, enabling use outside of traditional stationary imaging suites.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Portable ultrasound system; electronic linear/convex scanning; 64-element transducers (EZU-PL21, EZU-PC3B); 3.5MHz frequency; 4-step dynamic focusing with variable aperture; 512x512x4 bit memory; 16 grey levels; PAL/NTSC video output; Class I, Type B safety per IEC 801-1.
Indications for Use
Indicated for diagnostic ultrasound imaging of the human body, specifically fetal, abdominal, and adult cardiac applications, including M-mode for adult cardiac imaging.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
K963065
SIUI
SHANTOU INSTITUTE OF ULTRASONIC INSTRUMENTS
MAY 29 1997
# Premarket Notification [510(k)] Summary
July 30, 1996 (Revised 5/6/97)
**Trade Name:** CTS-285 with EZU-PL21 and EZU-PC3B Transducers
**Common Name:** Diagnostic Ultrasound System
**Classification Name:** Ultrasonic Pulsed Echo Imaging System, 90 IYO (per 21 CFR section 892.1560)
**Manufacturer's Name:** Shantou Institute of Ultrasonic Instruments
**Address:** #2, Jinsha Road, M., Shantou Sez, 515041, China
**Corresponding Official:** Mr. Jinzhong Yao
**Title:** President
**Telephone:** (86) 754-8250150
**Fax:** (86) 754-8251499
**Predicate:** Hitachi Medical Corporation EUB-310, K862867
**Device Description:** Model CTS-285 is a compact-type linear/convex electronic scanning ultrasound system with a built-in digital scan converter (DSC). The unit allows adult heart, abdominal organ and fetal tomographic images to be observable on a video monitor. The main unit is portable and is separable from other equipment to be carried for its use at another place as well as being usable in combination with a full keyboard, 9-inch video monitor and a special photographic unit.
**Intended Use:** Ultrasonic pulsed echo imaging and measurement for fetal, abdominal, and adult cardiac imaging as well as adult cardiac M-mode.
**Technological Characteristics:** See the attached "Comparison List" of the SIUI CTS-285 and the Hitachi EUB-310.
国家·中国国际电视学院
ADD: JINSHA ROAD, M., SHANTOU, CHINA
电话:0754-8251499
{1}
COMPARISON LIST OF SIUI PRODUCT AND HITACHI PRODUCT
| Performance | | CTS-285 (portable) (SIUI) | EUB-210A (HITACHI) |
| --- | --- | --- | --- |
| main unit | scanning mode | electronic linear scanning (compatible 64 elements linear probe)
electronic convex sector (compatible convex probe of 64 elements) | electronic linear scanning (compatible 80 elements linear probe)
electronic convex sector (compatible convex probe of 80 elements) |
| | display mode | B mode, B/B mode, M mode, B/M mode simultaneously | B mode, B/B mode, M mode, B/M mode simultaneously |
| | measurement | in B mode display: distance, area and circumference
in M mode display: time interval, velocity, depth and heart rate | in B mode display: distance, area and circumference
in M mode display: time interval, velocity, depth and heart rate |
| | calculation | area, circumference, volume, heart rate, pregnant weeks and heart function | area, circumference, volume, heart rate, pregnant weeks and heart function |
| | focusing mode | 4-steps dynamic focusing with variable aperture and lens focusing | 4-steps dynamic focusing with variable aperture and lens focusing |
| | scanning width | linear scanning: 3.5MHz probe 102mm, convex sector scanning probe: sector angle 60° | linear scanning: 3.5MHz probe 104mm, 5MHz probe 61mm
convex sector scanning probe: sector angle 60° |
| | transmitting voltage | pulse height 130V | pulse height 130V |
| | transmitting pulse width | 3.5MHz pulse width 160μs | 3.5MHz pulse width 140μs
5MHz pulse width 100μs |
| | detecting depth | 3.5MHz probe maximum depth: 210mm | 3.5MHz probe maximum depth: 210mm
5MHz probe maximum depth: 140mm |
| | zoom | 3.5MHz probe: x1.0, x1.2, x1.5, x2.0 selectable as well as depth shift | 3.5MHz probe: x1.0, x1.2, x1.5, x2.0 selectable as well as depth shift
5MHz probe: x1.0, x1.5, x2.0 selectable as well as depth shift |
| | frame rate | the maximum is 40 frame/second | the maximum is 40 frame/second |
| | grey scale | 16 | 16 |
| | memory | 512x512x4 bit | 512x512x4 bit |
| | coordinate transformation function | monitor can display electronic linear scanning image or convex sector scanning image | monitor can display electronic linear scanning image or convex sector scanning image |
{2}
COMPARISON LIST OF SIUI PRODUCT AND HITACHI PRODUCT
| Performance | | CTS-285 (portable) (SIUI) | EUB-310A (HITACHI) |
| --- | --- | --- | --- |
| main unit | video output | PAL or NTSC system TV signal
(confirmed in order) | PAL or NTSC system TV signal
(confirmed in order) |
| | power supply | 100V, 110V, 117V, 200V, 220V or
234V, ±10%, 50/80Hz, about 140W | 100V, 110V, 117V, 200V, 220V or
234V, ±10%, 50/80Hz, about 250W |
| | monitor | 5.5° black and white monitor | 5.5° and 9° black and white monitor |
| | volume and weight | 280(W)×235(H)×415(L)mm
approx. 13kg | 405(w)×710(1)×1310(h)mm
approx. 40kg |
| | cursor shift | by trackball on keyboard | by joystick on keyboard |
| | electric apparatus
safety standard | conform at requirement of I class
B type apparatus of IEC 801-1
isolate resistor testing: testing
voltage 1000V L-L, L-G>10MG
leakage current: U~G<500 μA
P-G<100 μA
voltage resistance testing: L-G,
P-G 1500V 2mA, no sparking or
arcing in I minute
work normally when voltage changes
±10% | conform at requirement of I class
B type apparatus of IEC 801-1
isolate resistor testing: testing
voltage 1000V L-L, L-G>10MG
leakage current: U~G<500 μA
P-G<100 μA
voltage resistance testing: L-G,
P-G, 1500V 2mA, no sparking or
arcing in I minute
work normally when voltage changes
±10% |
| probe ultrasound
transmitting
frequency | | linear probe:
EZU-PL21 64 elements, 3.5MHz,
scanning width 102mm
Convex sector probe:
EZU-PC3B 64 elements, 40R, 3.5MHz,
scanning width 60°
; | linear probe:
EZU-PL11 80 elements, 3.5MHz
scanning width 104mm
EZU-PL12 80 elements, 5MHz
scanning width 61mm
EUU-L11S 80 elements, 3.5MHz
scanning width 84mm
convex sector probe:
EZU-PC3A 80 elements, 40R, 3.5MHz,
scanning angle 60°
EZU-PC2A 80 elements, 40R, 5MHz,
scanning angle 60°
EUP-V12A 40 elements, 40R, 5MHz,
scanning angle 40° (transvergnal
probe) |
| operation
environment | | temperature 5-40° C, relative
humidity 30-50% (no water drop) | temperature 5-40° C, relative
humidity 30-85% (no water drop) |
| storage
environment | | temperature -10-60° C, relative
humidity 30-95% (no water drop)
air pressure 700-1060mB | temperature -10-60° C, relative
humidity 30-95% (no water drop)
air pressure 700-1060mB |
| note | | | |
*U means main unit.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Shantou Institute of Ultrasonic Instruments
c/o Robert J. Morton, President
Quality and Regulatory Services
1106 Chiltern Drive
Walnut Creek, CA 94596
MAY 29 1997
Re: K963065
CTS-285 Diagnostic Ultrasound
System with Model EZU-PL 21
and EZU-PC3B Transducers
Dated: April 4, 1997
Received: April 7, 1997
Regulatory Class: II
21 CFR 892.1560/Procode: 90 IYO
21 CFR 892.1570/Procode: 90 ITX
Dear Mr. Morton:
We have reviewed your section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act. You may, therefore, market the device, subject to the general controls provisions Act (Act). The general controls provisions of the Act include requirements for registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the CTS-285 Diagnostic Ultrasound System, as described in your premarket notification:
**Transducer Model Number**
EZU-PL21 (3.5MHz)
EZU-PC3B (3.5MHz)
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the *Code of Federal Regulations*, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the *Federal Register*. *Please note*: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
{4}
Page 2 - Robert J. Morton
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, “Misbranding by reference to premarket notification” (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address “dsmo@fdadr.cdrh.fda.gov”.
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure(s)
{5}
Ultrasound Device Indications Statement Page 1 of 2
510(k) Number (if known): K963065
Device Name: Transducer 3.5 MHz EZU-PL21 for CTS-285
Fill out one form for each ultrasound system or transducer.
Indications For Use: Diagnostic ultrasound imaging XEROX (Specify) of the human body as follows:
Mode of Operation
| Clinical Application | A | B | N | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined. (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Ophthalmic | | | | | | | | | | |
| Fetal | | X | | | | | | | | |
| Abdominal | | X | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | | | |
| Intra-operative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | X | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CNSR, Office of Device Evaluation (OPE)
David L. Seyman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K963065
Prescription Use (Per 21 CFR 801.109)
{6}
Ultrasound Device Indications Statement Page 2 of 2
510(k) Number (if known): K963065
Device Name: Transducer Model 3.5MHz EZU-PC3B for CTS-285
Fill out one form for each ultrasound system or transducer.
Indications For Use: Diagnostic ultrasound imaging (Specify) of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined. (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Ophthalmic | | | | | | | | | | |
| Fetal | | X | | | | | | | | |
| Abdominal | | X | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | | | |
| Intra-operative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | X | X | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRN, Office of Device Evaluation (CDE)
David G. Segerman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K963065
Prescription Use (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.