GLOVEL GX

K963020 · Kenwell International , Ltd. · KGO · Nov 27, 1996 · General, Plastic Surgery

Device Facts

Record IDK963020
Device NameGLOVEL GX
ApplicantKenwell International , Ltd.
Product CodeKGO · General, Plastic Surgery
Decision DateNov 27, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4460
Device ClassClass 1

Indications for Use

This device is intended to be used as a single use disposable sterile surgical glove.

Device Story

GLOVEL Gx Powder-free Latex Surgical Gloves; single-use disposable sterile surgical gloves; intended for use by surgeons and clinical staff during surgical procedures to provide a protective barrier. Device manufactured from latex; sterilized via Ethylene Oxide. Output is the physical glove itself. Clinical benefit includes protection against cross-contamination between patient and clinician.

Clinical Evidence

No clinical data; bench testing only. Testing included physical and dimensional inspection (AQL 2.5, Level S-2), FDA 1000 ml water fill test (AQL 1.5, Level I), and biocompatibility testing (primary skin irritation and sensitization).

Technological Characteristics

Material: Latex. Standard: ASTM D 3577-91. Sterilization: Ethylene Oxide (ETO) per EN550/AAMI/ANSI/ISO-11145-1994 (SAL 10^-5). Parameters: 750 mg/L ETO, 6-hour exposure, 55-60°C, 50% RH.

Indications for Use

Indicated for use as a single-use disposable sterile surgical glove for surgical procedures.

Regulatory Classification

Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0} 12-12-1996 11:38AM FROM mcl CONSULTANTS 3215942977 NOV 27 1996 Exhibit #1 Page 1 of 3 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: K963020 1. Submitter's Identification: C. Alva Kemwell International Ltd. 3B, Peenya, II Phase Bangalore 560 058, India Date Summary Prepared: 10th July 1996 2. Name of the Device: GLOVEL Gx Powder-free Latex Surgical Gloves. 3. Predicate Device Information: Class I Powder-free Latex Surgical Gloves which meets the requirements of ASTM D 3577-91. The equivalent device identified in the market is SAFESKIN-SUPRA of Safeskin Corporation, USA. 4. Device Description: Classification by FDA's General and Plastic Surgery Device Panel as Class 1, 21 CFR 878.4460, Surgical Powder-free Latex Gloves, 79KGO, and meets all requirements of ASTM Standard D-3577-91. 5. Intended Use: This device is intended to be used as a single use disposable sterile surgical glove. {1} 12-12-1996 11:39AM FROM mc i CONSULTANTS 13215942977 P.23 Exhibit #1 Page 2 of 3 6. Comparison to Predicate Device: Kemwell International Ltd. Powder-free Latex Surgical Gloves is substantially equivalent in safety and effectiveness to the Gloves sold by Safeskin Corporation USA, the brand name SAFESKIN-SUPRA. 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: The standards used for Glovel Gx, Powder-free Latex Surgical Gloves production are based on ASTM-D-3577-91. All tests meet requirements of Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 2.5. The FDA 1000 ml Water Fill Test was also conducted with samplings of AQL 1.5, Inspection Level I meeting these requirements. Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions. There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels. Method of Sterilization: Ethylene Oxide Methodology Used: EN550 Equivalent to AAMI/ANSI/ISO-11145-1994 Standard. Sterility Assurance Level: 10⁻⁵ Parameters Employed: ETO Concentration: 750 mg/L {2} 12-12-1996 11:39AM FROM mdl CONSULTANTS 130159429 P.04 Exhibit #1 Page 3 of 3 Exposure Time: 6 hrs. Temperature: 55-60°C RH: 50% (min) 8. Discussion of Clinical Tests Performed: Not applicable - there is no hypoallergenic claim. 9. Conclusion: Kemwell International Ltd. Powder-free Latex Surgeon's Gloves conform fully to ASTM-D-3577-91 standards as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in our 510(k). There are no safety/efficacy issues or new claims from the "Substantial Equivalence" products cited. Based on the non-clinical tests our product has demonstrated to be as safe and effective as our predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...