KARL STORZ MODEL 264305-20 ELECTRONIC ENDOFLATOR

K962863 · KARL STORZ Endoscopy-America, Inc. · HIF · Nov 19, 1996 · Obstetrics/Gynecology

Device Facts

Record IDK962863
Device NameKARL STORZ MODEL 264305-20 ELECTRONIC ENDOFLATOR
ApplicantKARL STORZ Endoscopy-America, Inc.
Product CodeHIF · Obstetrics/Gynecology
Decision DateNov 19, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1730
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Karl Storz Model 264305 20 Electronic Endoflator is designed to facilitate the use of laparoscopes by distending the abdomen with CO₂ gas during laparoscopic surgical and diagnostic procedures.

Device Story

Microprocessor-controlled laparoscopic insufflator; delivers CO₂ gas to distend abdomen. User-adjustable pressure (0-30 mmHg); max flow rate 20 L/min. Used in surgical/diagnostic settings by physicians. Features acoustic/visual alarms for overpressure, negative pressure, and low gas supply. Assists visualization during laparoscopy by maintaining pneumoperitoneum.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Microprocessor-controlled insufflator; CO₂ gas source; 0-30 mmHg pressure range; 20 L/min max flow; acoustic/visual alarm system.

Indications for Use

Indicated for patients undergoing laparoscopic surgical and diagnostic procedures requiring abdominal distension via CO₂ insufflation.

Regulatory Classification

Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

Submission Summary (Full Text)

{0} Karl Storz Endoscopy-America, Inc. 600 Corporate Pointe Culver City, California 90230-7600 Phone 310 558 1500 Toll Free 800 421 0837 Fax 310 410 5527 K962863 NOV 19 1996 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document are accurate and complete to the best of KSEA's knowledge. **Applicant:** Karl Storz Endoscopy - America, Inc. 600 Corporate Pointe Culver City, CA 90230 (310) 558-1500 **Contact:** Betty M. Johnson Manager, Regulatory Affairs **Device Identification:** Common Name Laparoscopic Insufflator Trade Name Karl Storz Model 264305 20 Electronic Endoflator **Indication:** The Karl Storz Model 264305 20 Electronic Endoflator is designed to facilitate the use of laparoscopes by distending the abdomen with CO₂ gas during laparoscopic surgical and diagnostic procedures. **Device Description:** The Karl Storz Model 264305 20 Electronic Endoflator for laparoscopy is a microprocessor controlled insufflator system with a maximum gas flow rate of 20 liters per minute. The insufflation pressure is user adjustable between 0 and 30 mmHg. The safety features include acoustic and visual alarms for overpressure, negative pressure and low gas supply. **Substantial Equivalence:** The Karl Storz Model 264305 20 Electronic Endoflator for laparoscopy is substantially equivalent to the predicate devices since the basic features, design and intended uses are the same. The minor differences between the KSEA Endoflator and the predicate devices raise no new issues of safety and effectiveness, as these differences have no effect on the performance, function or intended use of this device. Signed: Betty M. Johnson Manager, Regulatory Affairs 000122
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...