OPUS MYOGLOBIN CONTROLS

K962841 · Behring Diagnostics, Inc. · JJX · Aug 15, 1996 · Clinical Chemistry

Device Facts

Record IDK962841
Device NameOPUS MYOGLOBIN CONTROLS
ApplicantBehring Diagnostics, Inc.
Product CodeJJX · Clinical Chemistry
Decision DateAug 15, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

The OPUS Myoglobin Controls are intended for use as quality control material to monitor the precision and accuracy of the OPUS Myoglobin assay.

Device Story

OPUS Myoglobin Controls consist of lyophilized human myoglobin in processed human serum; provided at three levels (low, mid, high). Used exclusively with OPUS Myoglobin assay on OPUS Immunoassay System. Laboratory personnel use controls to monitor assay precision and accuracy by comparing measured values against lot-specific target values. Helps ensure reliability of myoglobin measurements for clinical decision-making.

Clinical Evidence

Bench testing only. Precision evaluated on OPUS Immunoassay System. Intra-assay precision (n=20 per level): %CV 4.27% to 6.05%. Inter-assay precision (n=8 over 15 days): %CV 6.55% to 10.54%.

Technological Characteristics

Lyophilized human myoglobin in processed human serum matrix. Tri-level (low, mid, high) configuration. Designed for use with OPUS Immunoassay System. No specific materials standards or software algorithms described.

Indications for Use

Indicated for use as quality control material to monitor precision and accuracy of the OPUS Myoglobin assay in clinical laboratory settings.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Submission Summary (Full Text)

{0} AUG 15 1996 K962841 # 510(k) Summary of Safety and Effectiveness for OPUS Myoglobin Controls 1. Manufacturer Name, Address, phone number, contact name and date of preparation. Manufacturer: Behring Diagnostics, Inc. 151 University Avenue Westwood, MA 02090 617-320-3117 Contact name: Ruth C. Forstadt Date of preparation: July 17, 1996 2. Device Name/Classification: OPUS Myoglobin Controls: Quality Control material (assayed) Classification number: Class 1 (862.1660) 3. Identification of the legally marketed device to which the submitter claims equivalence. OPUS hCG Controls 4. Proposed Device Description: The OPUS Myoglobin Controls are lyophilized controls containing known levels of human myoglobin in processed human serum. The control is provided at three levels (low, mid and high). The OPUS Myoglobin Control is only for use with the OPUS Myoglobin assay and has not been evaluated for use with other assays. 000009 {1} 5. Proposed Device Intended Use: The OPUS Myoglobin Controls are intended for use as quality control material to monitor the precision and accuracy of the OPUS Myoglobin assay. 6. Medical device to which equivalence is claimed and comparison information: The OPUS Myoglobin Controls are substantially equivalent in intended use to the OPUS hCG Controls. Both products are in vitro diagnostic reagents intended for use as a quality control material to monitor specific laboratory procedures. The OPUS Myoglobin Controls, like the OPUS hCG Controls, are a tri-level serum-based matrix control for specific OPUS assays. Both controls are provided with lot specific values and are for use with the OPUS assays only. The OPUS Myoglobin Controls differ from the OPUS hCG Controls in that the OPUS Myoglobin Controls are for use with the OPUS Myoglobin assay and contain known levels of myoglobin, while the OPUS hCG Controls are for use with the OPUS hCG assay and contain known levels of hCG. Also, the OPUS Myoglobin control is provided as a lyophilized control while the OPUS hCG controls are provided in a liquid form. 7. Proposed Device Performance Characteristics: Precision of the OPUS Myoglobin Controls was evaluated on an OPUS Immunoassay System with the OPUS Myoglobin assay. Intra assay precision was evaluated by running an n=20 with each level of the OPUS control. %CV's ranged from 4.27% to 6.05% for myoglobin. The inter assay precision was evaluated by running duplicate determinations for each level of control twice over a 15 day period to total an n=8. %CV's ranged from 6.55% to 10.54% for myoglobin. 000010
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...