← Product Code [KWP](/productcode/KWP) · K962708

# TITANIUM CD HORIZON SPINAL SYSTEM (K962708)

_Sofamor Danek USA,Inc. · KWP · Feb 6, 1997 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K962708

## Device Facts

- **Applicant:** Sofamor Danek USA,Inc.
- **Product Code:** [KWP](/productcode/KWP.md)
- **Decision Date:** Feb 6, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Titanium CD HORIZON™ Spinal System, when using screws for pedicle fixation, is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. Except for situations where screws are attached to the pedicles of the lumbar and sacral spine via a posterior surgical approach in a Titanium CD HORIZON™ construct for the treatment of severe spondylolisthesis (Grade 3 and Grade 4) at the L5-S1 vertebral joint, the specific indications for the Titanium CD HORIZON™ Spinal System are the following: 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies). 2. Pseudoarthrosis 3. Stenosis 4. Spondylolisthesis 5. Spinal deformities: scoliosis, kyphosis, lordosis 6. Fracture 7. Unsuccessful previous attempts at spinal fusion 8. Tumor resection NOTA BENE: The Titanium CD HORIZON™ Spinal System is limited to non-cervical use. The titanium alloy CD HORIZON™ screws and the titanium alloy TSRH® Variable Angle Screws, when used as pedicle screws with the Titanium CD HORIZON™ Spinal System, are intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. When used for pedicle screw fixation as described above, the screws are indicated only for insertion no higher than L3 and not lower than the sacrum. Otherwise the titanium alloy CD HORIZON™ screws and the titanium alloy TSRH® Variable Angle Screws, when used with the Titanium CD HORIZON™ Spinal System, are intended for sacral/iliac attachment only. All of the titanium alloy CD HORIZON™ hooks are intended for thoracic and/or lumbar attachment only. The titanium alloy TSRH® Low Profile CROSSLINK® Offset Plate, the titanium alloy TSRH® Low Profile CROSSLINK® plates, and the titanium alloy CD HORIZON™ Low Profile MULTI-SPAN® CROSSLINK® plates are intended for posterior thoracic, lumbar, and/or sacral use only.

## Device Story

Spinal fixation system consisting of titanium alloy rods, hooks, screws, and crosslink plates; components rigidly locked into patient-specific constructs. Used by surgeons in posterior spinal procedures to stabilize the spine during fusion. Provides mechanical support for degenerative conditions, deformities, and fractures. System components are modular, allowing for tailor-made constructs. Titanium alloy construction ensures biocompatibility. Device is non-cervical; pedicle screw use is highly restricted to specific spondylolisthesis cases requiring subsequent removal. Instruments provided to facilitate surgical implantation.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Materials: Titanium alloy (ASTM F136) and commercially pure titanium (ASTM F67). Components: 6.35mm rods, hooks, screws (5.5mm-7.5mm), and crosslink plates. Modular design for rigid locking configurations. Non-cervical use. Sterilization: Sold sterile or non-sterile.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Reference Devices

- TSRH® Spinal System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FEB 04 '97 01:59PM SOFAMOR DANEK
P.23/24

Titanium CD HORIZON™ Spinal System
510(k) Summary
K962708
February, 1997
FEB - 6 1997
K962708

I. Company: Sofamor Danek USA
1800 Pyramid Place
Memphis, TN 38132
(901) 396-3133

II. Proposed Proprietary Trade Name: Titanium CD HORIZON™ Spinal System

III. The Titanium CD HORIZON™ Spinal System consists of the following implant components which can be rigidly locked into a variety of configurations, with each resulting spinal implant construct being tailor-made for the individual case.

- CD HORIZON™ rod (commercially pure titanium), 6.35mm diameter
- CD HORIZON™ knurled rod (titanium alloy), 6.35mm diameter
- CD HORIZON™ hooks (titanium alloy)
- CD HORIZON™ screws (titanium alloy): 5.5mm, 6.5mm, and 7.5mm diameters
- CD HORIZON™ Break-Off Set Screw (titanium alloy)
- CD HORIZON™ Low Profile MULTI-SPAN™ CROSSLINK® Plates (titanium alloy)

In addition, the following Titanium TSRH® Spinal system implants can be used with the Titanium CD HORIZON™ Spinal System.

- TSRH® rod (titanium alloy), 6.35mm diameter
- TSRH® Variable Angle Screws (titanium alloy): 6.5mm and 7.5mm diameters
- TSRH® Top Tightening Variable Angle T-Bolts with Locking Screw (titanium alloy): small, medium, and large
- TSRH® Low Profile CROSSLINK® Offset Plate (titanium alloy)
- TSRH® Low Profile CROSSLINK® Plates (titanium alloy)
- TSRH® Low Profile CROSSLINK® Plate Set Screws (titanium alloy)

The Titanium CD HORIZON™ Spinal System components and the others mentioned above are made out of titanium alloy described by ASTM Standard F136 or its ISO equivalent. The Titanium CD HORIZON™ Spinal System also features a rod fabricated from commercially pure titanium conforming to ASTM F67 or its ISO equivalent. Titanium CD HORIZON™ implant components must not be used with implant components fabricated from stainless steel or any other alloy in a construct. The Titanium CD HORIZON™ implants may be sold sterile or non-sterile.

Instruments are available to facilitate the use of the device.

{1}

FEB 04 '97 02:00PM SOFAMOR DANEK
P.24/24

The Titanium CD HORIZON™ Spinal System, when using screws for pedicle fixation, is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass.

Except for situations where screws are attached to the pedicles of the lumbar and sacral spine via a posterior surgical approach in a Titanium CD HORIZON™ construct for the treatment of severe spondylolisthesis (Grade 3 and Grade 4) at the L5-S1 vertebral joint, the specific indications for the Titanium CD HORIZON™ Spinal System are the following:

1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies).
2. Pseudoarthrosis
3. Stenosis
4. Spondylolisthesis
5. Spinal deformities: scoliosis, kyphosis, lordosis
6. Fracture
7. Unsuccessful previous attempts at spinal fusion
8. Tumor resection

NOTA BENE: The Titanium CD HORIZON™ Spinal System is limited to non-cervical use. The titanium alloy CD HORIZON™ screws and the titanium alloy TSRH® Variable Angle Screws, when used as pedicle screws with the Titanium CD HORIZON™ Spinal System, are intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. When used for pedicle screw fixation as described above, the screws are indicated only for insertion no higher than L3 and not lower than the sacrum. Otherwise the titanium alloy CD HORIZON™ screws and the titanium alloy TSRH® Variable Angle Screws, when used with the Titanium CD HORIZON™ Spinal System, are intended for sacral/iliac attachment only. All of the titanium alloy CD HORIZON™ hooks are intended for thoracic and/or lumbar attachment only. The titanium alloy TSRH® Low Profile CROSSLINK® Offset Plate, the titanium alloy TSRH® Low Profile CROSSLINK® plates, and the titanium alloy CD HORIZON™ Low Profile MULTI-SPAN® CROSSLINK® plates are intended for posterior thoracic, lumbar, and/or sacral use only.

V. The Titanium CD HORIZON™ Spinal System was declared substantially equivalent to other predicate or preamendments devices. Mechanical test data were provided in the application.

©1997 Sofamor Danek

---

**Source:** [https://fda-staging.innolitics.com/device/K962708](https://fda-staging.innolitics.com/device/K962708)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).