# COBE BRAT 2 AUTOLOGOUS BLOOD SALVAGE SYSTEM WITH MODIFIED PLASMA RECOVERY PROGRAM (PRP) (K962689)

_Cobe Cardiovascular, Inc. · CAC · Jan 28, 1997 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K962689

## Device Facts

- **Applicant:** Cobe Cardiovascular, Inc.
- **Product Code:** [CAC](/productcode/CAC.md)
- **Decision Date:** Jan 28, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5830
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The COBE BRAT 2 is indicated for use for recovery and/or processing of autologous blood.

## Device Story

System processes autologous blood for plasma sequestration; collects, anticoagulates, and delivers blood from patient to instrument. Modifications include expanded fill cycle volume delivery (25-200 ml/min), manual operation mode, and hardware/software for pump reversal prevention. Used in clinical settings by trained personnel. Output includes processed plasma and salvaged red blood cells for reinfusion. Benefits include efficient autologous blood recovery and reduced need for donor blood.

## Clinical Evidence

No clinical testing was performed. Bench testing included in vitro blood processing performance (plasma volume, free hemoglobin, platelet/WBC/RBC recovery, hematocrit) and human blood testing for platelet aggregation and pH following 6-hour storage.

## Technological Characteristics

Autotransfusion apparatus with modified process pump control. Hardware includes redundant sensors for pump direction detection and mechanical interlocks to prevent clockwise rotation during fill valve open state. Software controls pump speed (25-200 ml/min) and manual mode operation. Disposable set allows direct patient-to-instrument connection.

## Regulatory Identification

An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.

## Predicate Devices

- Plasma Recovery Program (PRP) for the COBE® BRAT® 2 Autologous Blood Salvage System ([K933960](/device/K933960.md))
- Electromedics Autotransfusion Machine AT 500

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K962689
JAN 28 1997

# 510(k) SUMMARY

SUBMITTER: COBE Cardiovascular, Inc.®
14401 W. 65th Way
Arvada, CO 80004

CONTACT PERSON: Mary L. Armstrong
Phone: (303) 467-6521
Fax: (303) 467-6525

DATE PREPARED: July 8, 1996

DEVICE TRADE NAME: Plasma Sequestration Option for the COBE® BRAT® 2 Autologous Blood Salvage System

COMMON NAME: Plasma sequestration option for autologous blood salvage system

CLASSIFICATION NAME: Autotransfusion Apparatus

PREDICATE DEVICES: Plasma Recovery Program (PRP) for the COBE® BRAT® 2 Autologous Blood Salvage System
Electromedics Autotransfusion Machine AT 500

## DEVICE DESCRIPTION:

The COBE BRAT 2 Autologous Blood Salvage System with Modified Plasma Recovery Program (PRP) consists of a number of changes from the original Plasma Recovery Program for the COBE BRAT® 2 approved per 510(k) number K933960. These changes include:

- Expanded volume delivery range of 25 to 200 ml/min for the fill cycle of PRP
- Addition of a manual mode of operation for PRP
- Hardware and software modifications for detection and prevention of pump reversal
- Development of a disposables set to permit direct connection to the patient for blood collection
- Other minor changes to the software to improve usability of the BRAT 2 instrument

COBE®, BRAT®, and COBE Cardiovascular, Inc.® are registered trademarks for COBE Laboratories, Inc.

255.

{1}

INDICATIONS FOR USE:

The COBE BRAT 2 is indicated for use for recovery and/or processing of autologous blood.

TECHNOLOGICAL CHARACTERISTICS:

The COBE BRAT 2 Autologous Blood Salvage System with Modified Plasma Recovery Program (PRP) is a modification of its predicate, the Plasma Recovery Program (PRP) for the COBE® BRAT® 2 Autologous Blood Salvage System.

Through software modifications, the process pump has the capability to run at a wider range of fill speeds for plasma sequestration purposes. Redundant hardware has been added for detection and prevention of pump direction reversal and hardware has been added to prevent clockwise pump rotation with the fill valve open which would result in fluid flow towards the patient if directly connected in a PRP mode of operation. Software modifications have also been made to create a manual mode of operation for plasma sequestration.

A disposable set for the COBE BRAT 2 with Modified PRP permits the collection, anticoagulation, and delivery of blood directly from the patient to the BRAT 2 instrument for plasma sequestration purposes, as is done in the predicate Electromedics AT 500 System.

NONCLINICAL TEST RESULTS:

The COBE BRAT 2 with Modified PRP was tested to assure that it met its functional specifications. In vitro blood testing was done to assure the performance of the modified system was comparable to the predicate BRAT 2 PRP system. In vitro testing consisted of plasma volume per liter of blood processed, plasma free hemoglobin generated per liter of blood processed, percent platelets harvested, percent white blood cells recovered, percent platelets recovered, percent red cells recovered, hematocrit in reinfused blood and hematocrit in collected plasma. In addition, human blood testing was performed with the direct draw set to measure platelet aggregation and pH following storage for 6 hours.

CLINICAL TEST RESULTS:

No clinical testing was performed.

CONCLUSIONS:

The COBE BRAT 2 Autologous Blood Salvage System with Modified Plasma Recovery Program (PRP), through hardware and software modifications and disposables design, is substantially equivalent to the Plasma Recovery Program (PRP) for the COBE BRAT 2 Autologous Blood Salvage System and the Electromedics Autotransfusion Machine AT 500.

---

**Source:** [https://fda-staging.innolitics.com/device/K962689](https://fda-staging.innolitics.com/device/K962689)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).