K962633 · Abx, Inc. · GKZ · Nov 4, 1996 · Hematology
Device Facts
Record ID
K962633
Device Name
VEGA HEMATOLOGY ANALYZER
Applicant
Abx, Inc.
Product Code
GKZ · Hematology
Decision Date
Nov 4, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
The VEGA™ Hematology Analyzer is a quantitative, automated in vitro diagnostic analyzer capable of performing the following twenty six (26) hematological parameters: white blood cells (leukocytes) -WBC, red blood cells (erythrocytes) -RBC, hemoglobin -HGB, hematocrit -HCT, mean corpuscular volume -MCV, mean corpuscular hemoglobin -MCH, mean corpuscular hemoglobin concentration -MCHC, red blood cell distribution width -RDW, platelets -PLT, plateletcrit -PCT, platelet distribution width -PDW, mean platelet volume -MPV, lymphocyte (number) -LYM#, lymphocyte (percent of WBC) -LYM%, monocyte (number) -MON#, monocyte (percent of WBC) -MON%, basophil (number) -BAS#, basophil (percent of WBC) -BAS%, neutrophil (number) -NEU#, neutrophil (percent of WBC) -NEU%, eosinophil (number) -EOS#, eosinophil (percent of WBC) -EOS%, Atypical Lymphocyte (number) -ALY#, Atypical Lymphocyte (percent of lymphocyte number) -ALY%, Large Immature cell (number) -LIC#, Large Immature cell (percent of WBC) -LIC%. The VEGA Hematology Analyzer can be programmed to printout any of the parameters here above independently. Only twenty (20) parameter printout will be made available for diagnostic use in the United States (eliminating PCT, PDW, ALY%, ALY#, LIC%, LIC#).
Device Story
Automated hematology analyzer; processes whole blood samples to provide 20 hematological parameters (WBC, RBC, HGB, HCT, indices, and differential counts). Uses impedance variation for cell counting (WBC, RBC, PLT); spectrophotometry (550 nm) for HGB; and combined cytochemistry (Eosinofix™), impedance, and optical transmission (tungsten halogen source) for leukocyte differentiation. Operated by laboratory personnel in clinical settings. Output provided as printed reports for healthcare providers to assess patient hematological status. Benefits include rapid, automated blood cell analysis and differential counts.
Clinical Evidence
Clinical studies conducted at three hospital sites comparing VEGA performance against TECHNICON H2, ABBOT CELL-DYN 3500, and COULTER STKS. Study evaluated precision (CV%), accuracy (R² correlation), linearity, and carryover. Reproducibility showed CV <2.0% for WBC/RBC/HCT and <5.0% for PLT. Accuracy correlations (R²) were high for primary parameters (e.g., WBC >0.99, RBC >0.98). Manual differential counts performed per NCCLS H20-A guidelines served as reference for differential accuracy.
Technological Characteristics
Automated hematology analyzer using impedance variation and optical transmission (tungsten halogen light). Employs five reagents: VEGADIL™, VEGALYSE™, VEGACLEAN™, EOSINOFIX™, and BASOLYSE™. Measures 20 parameters via electronic pulse height analysis and spectrophotometry. Standalone benchtop unit.
Indications for Use
Indicated for quantitative, automated in vitro diagnostic analysis of whole blood specimens to provide a 20-parameter hematological profile for clinical diagnostic use.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
TECHNICON H2® Hematology Analyzer
ABBOT CELL-DYN® 3500 Hematology Analyzer
COULTER STKS® Hematology Analyzer
Reference Devices
Roche COBAS Argos 5 DIFF Hematology Analyzer
Submission Summary (Full Text)
{0}
ABX, Rue du Caducée, Montpellier, France 34184.
VEGA™ Hematology Analyzer 510(k) Summary
K962633
510(k) SUMMARY
NOV - 4 1996
# VEGA™ Hematology Analyzer
In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92.
# I. Identification of the Sponsor and Objective:
ABX
Parc Euromédecine,
rue du Caducée - BP 7290,
34184 MONTPELLIER Cédex 4 - FRANCE.
510(k) Submission dated 2/7, 1996.
Contact: Mr. Brice Prouvé
Business Development Manager.
phone: (33) 6714-1516 / 59 / 72
fax:: (33) 6714-1517
# II. Device Name:
Trade/Proprietary Name - VEGA™ Hematology Analyzer
Common or Usual Name/ Classification Name - Automated cell counter [864.5200] and Automated Differential Cell Counter [864.5220]
# III. Identification of the legally marketed device to which the 510(k) & sponsor claims equivalence:
The VEGA™ Hematology Analyzer is substantially equivalent to the TECHNICON H2 ® Hematology Analyzer, the ABBOT CELL-DYN ® 3500 Hematology Analyzer and the COULTER STKS ® Hematology Analyzer.
{1}
ABX, Rue du Caducée, Montpellier, France 34184.
VEGA™ Hematology Analyzer 510(k) Summary
## IV. Description of the Device/Statement of Intended Use:
The VEGA™ Hematology Analyzer is a quantitative, automated in vitro diagnostic analyzer capable of performing the following twenty six (26) hematological parameters:
| white blood cells (leukocytes) | -WBC |
| --- | --- |
| red blood cells (erythrocytes) | -RBC |
| hemoglobin | -HGB |
| hematocrit | -HCT |
| mean corpuscular volume | -MCV |
| mean corpuscular hemoglobin | -MCH |
| mean corpuscular hemoglobin concentration | -MCHC |
| red blood cell distribution width | -RDW |
| platelets | -PLT |
| plateletcrit | -PCT |
| platelet distribution width | -PDW |
| mean platelet volume | -MPV |
| lymphocyte (number) | -LYM# |
| lymphocyte (percent of WBC) | -LYM% |
| monocyte (number) | -MON# |
| monocyte (percent of WBC) | -MON% |
| basophil (number) | -BAS# |
| basophil (percent of WBC) | -BAS% |
| neutrophil (number) | -NEU# |
| neutrophil (percent of WBC) | -NEU% |
| eosinophil (number) | -EOS# |
| eosinophil (percent of WBC) | -EOS% |
| Atypical Lymphocyte (number) | -ALY# |
| Atypical Lymphocyte (percent of lymphocyte number) | -ALY% |
| Large Immature cell (number) | -LIC# |
| Large Immature cell (percent of WBC) | -LIC% |
The VEGA Hematology Analyzer can be programmed to printout any of the parameters here above independently.
Only twenty (20) parameter printout will be made available for diagnostic use in the United States (eliminating PCT, PDW, ALY%, ALY#, LIC%, LIC#).
The VEGA Hematology Analyzer utilizes five (5) different reagents.
| - | VEGADIL™ | : | Diluant |
| --- | --- | --- | --- |
| - | VEGALYSE™ | : | Lysing Agent |
| - | VEGACLEAN™ | : | Cleaner |
| - | EOSINOFIX™ | : | Staining Agent |
| - | BASOLYSE™ | : | Lysing Agent |
{2}
ABX, Rue du Caducée, Montpellier, France 34184.
VEGA™ Hematology Analyzer 510(k) Summary
# VEGA Measurement Principles:
WBC, RBC, PLT: The counting principle is based on an impedance variation generated by the passage of cells through the calibrated micro-aperture.
HGB: A cyanmethemoglobin compound is measured by spectrophotometry through the optical part of the WBC chamber at a wavelength of 550 nanometers.
HCT: The height of the impulse generated by the passage of the cell through the micro-aperture is directly proportional to the volume of analyzed red blood cell. The HCT measurement is performed by a specialized electronic device which adds all the pulse heights.
MCH, MCHC, MCV are red blood cell indices calculated from the measured parameters.
RDW is a coefficient of variation of the RBC distribution and is expressed as a percent of the average RBC size.
MPV is a platelet index derived from the accumulation of platelet data within the platelet population distribution.
# Leukocyte Differential Counts:
The VEGA Hematology Analyzer uses the similar method counting as the Roche COBAS Argos 5 DIFF Hematology Analyzer. The VEGA Hematology Analyzer combines the use of histochemical stains, measurement of blood cell volume via optical transmission from a tungsten halogen light source to provide leukocyte differentiation. On each individual cell, three methods are applied:
1. Cytochemistry with the action of Eosinofix™ reagent.
2. Impedance to measure the volume.
3. Optical transmission measurement to obtain information on the internal structure of the cells.
# V. Summary of the technological characteristics of the new device in comparison to those of the predicate:
The intended use and measurement principles are the same for each analyzer. There are differences in operation, configuration and capabilities of each device. Algorithms and threshold discrimination of each instrument are slightly different.
{3}
ABX, Rue du Caducée, Montpellier, France 34184.
VEGA™ Hematology Analyzer 510(k) Summary
## VI. Brief discussion of the clinical and non-clinical tests relied on for a determination of substantial equivalence:
Clinical studies were performed at three hospital university sites.
The clinical study was designed to demonstrate the precision and accuracy of the VEGA Analyzer in a typical setting with operators of various levels of experience and educational backgrounds.
Each site used different reference instrument: Site 1 used a TECHNICON H2® Hematology Analyzer; Site 2 used the ABBOT CELL-DYN® 3500 Hematology Analyzer; and Site 3 used the COULTER STKS® Hematology Analyzer.
Manual differential count were performed for the samples at all three sites to reflect the widest distribution of normal and abnormal specimens. The cell counts were performed according to NCCLS guidelines H20-A, at all three hospital laboratories. A summary of performance data follows:
## Reproducibility (within run):
Reproducibility of an instrument is measured by the coefficient of variation (CV) obtained from the calculation of 20 results of the same sample.
| Parameters | %CV | Reference Values |
| --- | --- | --- |
| - WBC: | < 2.0% | 10 10³/mm³ |
| - RBC: | < 2.0% | 5 10⁶/mm³ |
| - HGB: | < 1.0% | 15.0 g/dL |
| - HCT: | < 2.0% | 45.0% |
| - MCV: | < 1.5% | |
| - MCH: | < 1.5% | |
| - MCHC: | < 1.5% | |
| - RDW: | < 3.5% | |
| - PLT: | < 5.0% | 300 10³/mm³ |
| - MPV: | < 5.0% | |
| - LYM%: | < 5.0% | 30% |
| - LYM #: | < 5.0% | 2.1 10³/mm³ |
| - MON%: | < 15.0% | 5% |
| - MON#: | < 15.0% | 0.5 10³/mm³ |
| - NEU%: | < 3.0% | 60% |
| - NEU#: | < 3.0% | 4.2 10³/mm³ |
| - EOS%: | < 25.0% | 3% |
| - EOS#: | < 25.0% | 0.3 10³/mm³ |
| - BASO%: | < 40.0% | 1% |
| - BASO#: | < 40.0% | 0.1 10³/mm³ |
{4}
ABX, Rue du Caducée, Montpellier, France 34184.
VEGA™ Hematology Analyzer 510(k) Summary
## Linearity:
Linearity can be considered as the high and low limits for the instrument use for each parameter.
| Parameter | Range | Limits |
| --- | --- | --- |
| WBC (10³/mm³) | 0.1 - 85.0 (10³/mm³) | 0.2 x10³/mm³ or 3% of the expected value |
| RBC (10⁶/mm³) | 0.5 - 8.1 (10⁶/mm³) | 0.3 x10⁶/mm³ or 2% of the expected value |
| PLT (10³/mm³) | 10 - 1000 (10³/mm³) | 10 x10³/mm³ or 5% of the expected value |
| HGB (g/dL) | 2.0 - 25.0 (g/dL) | 0.2 g/dL or 2% of the expected value |
| HCT(%) | 10.0 - 70.0 (%) | 3% of the expected value |
## Carryover:
| | WBC | RBC | HGB | HCT | PLT |
| --- | --- | --- | --- | --- | --- |
| Mean Level | 57.5 | 8.5 | 24.0 | 72.0 | 944 |
| % Carry-over actual | 0.00 | 0.00 | 0.00 | 0.00 | 0.04 |
| CV % Claim | < 2.0% | <0.5% | <0.5% | <2.0% | <2.0% |
## Normal Ranges:
These values were taken from the "Hematology Clinical and Laboratory Practice" Copyright 1993 by Mosby - Year Book, Inc., Editor in Chief: Rodger L. Bick, MD.
| PARAMETERS | MALE | FEMALE |
| --- | --- | --- |
| WBC (10³/mm³) | 4.5 - 11 | 4.5 - 11 |
| Lymphocytes (%) | 25 - 33 | 25 - 33 |
| Monocytes (%) | 3 - 7 | 3 - 7 |
| Neutrophils (%) | 54 - 62 | 54 - 62 |
| Eosinophils (%) | 1 - 3 | 1 - 3 |
| Basophils (%) | 0 - 0.75 | 0 - 0.75 |
| RBC (10⁶/mm³) | 4.6 - 6.2 | 4.2 - 5.2 |
| HGB (g/dl) | 14 - 18 | 12 - 16 |
| HCT (%) | 40 - 54 | 37 - 47 |
| MCV (fL) | 80 - 96 | 80 - 96 |
| MCH (pg) | 26 - 34 | 26 - 34 |
| MCHC (g/dL) | 32 - 36 | 32 - 36 |
| RDW (%) | 12 - 16 | 12 - 14 |
| PLT(10³/mm³) | 150 - 350 | 150 - 350 |
| MPV (fL) | 7 - 9 | 8 - 10 |
All parameters here above can be adapted to the lab. requirement.
{5}
ABX, Rue du Caducée, Montpellier, France 34184.
VEGA™ Hematology Analyzer 510(k) Summary
**Accuracy:**
| | SITE 1 | | SITE 2 | | SITE 3 | |
| --- | --- | --- | --- | --- | --- | --- |
| parameter | N | R² | N | R² | N | R² |
| WBC (10³/mm³) | 190 | 0.9945 | 238 | 0.9965 | 216 | 0.9975 |
| RBC (10⁶/mm³) | 190 | 0.9873 | 238 | 0.9857 | 216 | 0.9933 |
| HGB (g/dL) | 190 | 0.9908 | 238 | 0.9947 | 216 | 0.9970 |
| HCT (%) | 190 | 0.9619 | 238 | 0.9820 | 216 | 0.9903 |
| MCV (μm³) | 190 | 0.8089 | 238 | 0.9346 | 216 | 0.9593 |
| MCH (pg) | 190 | 0.9402 | 238 | 0.9113 | 216 | 0.9646 |
| MCHC (g/dL) | 190 | 0.1484 | 238 | 0.2008 | 216 | 0.5266 |
| RDW (%) | 190 | 0.5457 | 238 | 0.8003 | 216 | 0.7450 |
| PLT (10³/mm³) | 190 | 0.9822 | 238 | 0.9827 | 216 | 0.9854 |
| MPV (μm³) | 190 | 0.2286 | 238 | 0.7149 | 216 | 0.8379 |
| %LYMPH | 190 | 0.8081 | 238 | 0.9669 | 216 | 0.9919 |
| %MON | 190 | 0.3030 | 238 | 0.7280 | 216 | 0.6996 |
| %NEUT | 190 | 0.7975 | 238 | 0.9179 | 216 | 0.9829 |
| %EOS | 190 | 0.8853 | 238 | 0.6618 | 216 | 0.9558 |
| %BASO | 190 | 0.3546 | 238 | 0.3789 | 216 | 0.2391 |
| Reference Analyzers | TECHNICON H2 | | CELL DYN 3500 | | COULTER STKS | |
**VII. Conclusions:**
Based on these data, the VEGA™ Hematology Analyzer is substantially equivalent to the TECHNICON H2® Hematology Analyzer, the ABBOT CELL-DYN® 3500 Hematology Analyzer and the COULTER STKS® Hematology Analyzer. for the intended use of performing a twenty parameters hematological profile of whole blood specimens.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.