DERMAPLEX HYDROGEL WOUND DRESSING

K962513 · Maxxim Medical · MGQ · Sep 12, 1996 · SU

Device Facts

Record IDK962513
Device NameDERMAPLEX HYDROGEL WOUND DRESSING
ApplicantMaxxim Medical
Product CodeMGQ · SU
Decision DateSep 12, 1996
DecisionSN
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

For the local management of pressure ulcers, stasis ulcers, 1st and 2nd degree burns, and full and partial thickness wounds.

Device Story

Dermaplex Hydrogel Wound Dressing; solid hydrogel dressing with polyester liner; supplied sterile in sealed aluminum/plastic laminate pouches; available in 3"x3", 5"x5", and 6"x10" sizes. Applied by clinician to wound site for local management of ulcers and burns. Provides moist wound environment. Biocompatibility established via rabbit dermal irritation, guinea pig sensitization, and USP Cytotoxicity (Agarose Diffusion) testing.

Clinical Evidence

Bench testing only. Biocompatibility established via primary dermal irritation (rabbits), sensitization (guinea pigs), and USP Cytotoxicity (Agarose Diffusion) testing.

Technological Characteristics

Solid hydrogel dressing; polyester liner; supplied in aluminum/plastic laminate pouches; sterile. Biocompatibility per USP Cytotoxicity (Agarose Diffusion), rabbit dermal irritation, and guinea pig sensitization tests.

Indications for Use

Indicated for local management of pressure ulcers, stasis ulcers, 1st and 2nd degree burns, and full and partial thickness wounds in patients requiring wound care.

Predicate Devices

Submission Summary (Full Text)

{0} K962513 pg. 1 of 2 # 510 (k) SUMMARY ## I. ADMINISTRATIVE Submitter: Maxxim Medical, Inc. 104 Industrial Boulevard Sugar Land, Texas 77478 713-240-2442 Contact Person: Ernest J. Henley Date of Preparation: July 8, 1996 ## II. DEVICE NAME Proprietary Name: Dermaplex Hydrogel Wound Dressing Common Name: Wound Dressing Classification Name: Hydrogel wound and burn dressing ## III. PREDICATE DEVICE Curagel Hydrogel Wound Dressing; Kendall Healthcare Products Co.; K950412 ## IV. DEVICE DESCRIPTION The Dermaplex Hydrogel Wound Dressing is a solid hydrogel dressing covered with a polyester liner which is removed prior to application. The product is available in three sizes (3" × 3", 5" × 5", and 6" × 10") and is supplied sterile in sealed aluminum/plastic laminate pouches. Biocompatibility of the device has been established by a primary dermal irritation test in rabbits and a sensitization test in guinea pigs. The product meets the requirements of the USP Cytotoxicity Test (Agarose Diffusion). ## V. INTENDED USE For the local management of pressure ulcers, stasis ulcers, 1st and 2nd degree burns, and full and partial thickness wounds. Premarket Notification: K962513 {1} K962513 Pg. 2 of 2 # 510 (k) SUMMARY ## VI. COMPARISON TO PREDICATE DEVICE The Dermaplex Hydrogel Wound Dressing is similar in composition, function and intended use to the Curagel Hydrogel Wound Dressing (Kendall Healthcare Products Co.) and other legally marketed wound dressing products consisting of solid hydrogel. Accordingly, Maxxim Medical, Inc. concluded that the Dermaplex Hydrogel Wound Dressing is as safe and effective for its intended use and performs at least as well as legally marketed predicate devices, such as the Curagel Hydrogel Wound Dressing. Premarket Notification K962513
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...