UREA NITROGEN (BUN) LIQUI-UV

K962418 · Stanbio Laboratory · CDQ · Dec 26, 1996 · Clinical Chemistry

Device Facts

Record IDK962418
Device NameUREA NITROGEN (BUN) LIQUI-UV
ApplicantStanbio Laboratory
Product CodeCDQ · Clinical Chemistry
Decision DateDec 26, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1770
Device ClassClass 2

Indications for Use

The device is used for the kinetic quantitative determination of Urea Nitrogen (BUN) in serum.

Device Story

In vitro diagnostic test kit; measures Urea Nitrogen (BUN) in serum. Principle: kinetic measurement of NADH to NAD conversion via urease and glutamic dehydrogenase (GLDH) enzymes; decrease in absorbance monitored. Components: two reagents (BUN Buffer R1, BUN Enzyme R2) and 30mg/dL standard. Used in clinical laboratory settings by technicians/pathologists. Output: quantitative BUN concentration. Clinical utility: assessment of renal function and nitrogen balance.

Clinical Evidence

Bench testing only. Method comparison against predicate device yielded correlation coefficient of 0.999 and regression equation y = 1.012x - 0.47. Additional validation included precision, linearity, sensitivity, stability, and interference studies; all results acceptable.

Technological Characteristics

In vitro diagnostic reagent kit. Enzymatic kinetic assay (Urease/GLDH). Measures absorbance decrease at 340nm. Components: Buffer (R1), Enzyme (R2), Standard. Manual or automated analyzer application.

Indications for Use

Indicated for the kinetic quantitative determination of Urea Nitrogen (BUN) in serum for clinical diagnostic purposes.

Regulatory Classification

Identification

A urea nitrogen test system is a device intended to measure urea nitrogen (an end-product of nitrogen metabolism) in whole blood, serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.

Predicate Devices

Submission Summary (Full Text)

{0} DEC 26 1996 K9C2418 # 510(k) SUMMARY ## Submitter's Name Kirk Johnson Tel. (210) 222-2108 Stanbio Laboratory, Inc. Fax (210) 227-6367 2930 East Houston Street San Antonio, TX 78202 Prepared By Kirk Johnson June 18, 1996 ## Product Name Trade name-Urea Nitrogen (BUN) Liqui-UV® Common name-Urea Nitrogen (BUN) Classification name-Urease and Glutamic Dehydrogenase, Urea Nitrogen Class II, 75CDQ ## Substantial Equivalence of Device This test is substantially equivalent to: Product Name: BUN, UV Urease/GLDH (Kinetic) Manufacturer: Boehringer Mannheim ## Description of Device The device test kit is comprised of two reagents, BUN Buffer (R1) and BUN Enzyme (R2), and a standard, 30mg/dL. A working reagent is prepared by combining five parts BUN Buffer (R1) to one part BUN Enzyme (R2). ## Intended Use of Device The device is used for the kinetic quantitative determination of Urea Nitrogen (BUN) in serum. ## Comparison of Devices Both Urea Nitrogen (BUN) methods measure kinetically the decrease in absorbance due to consumption and conversion of NADH to NAD. ## Performance Data Substantial equivalency was demonstrated by method comparison. Correlation was performed between the two test kits with a correlation coefficient of 0.999 and a regression equation of $y = 1.012x - 0.47$. In addition, precision, linearity, sensitivity, stability and interference studies were performed on Stanbio Laboratory, Inc., version of Urea Nitrogen (BUN) Liqui-UV®, Catalog # 2020. Results of these tests were found to be acceptable.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...