← Product Code [DZE](/productcode/DZE) · K962394

# IMPLANT INTEGRATION SYSTEMS DRILLS AND TAPS (K962394)

_Implant Integration Systems, Inc. · DZE · Oct 11, 1996 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K962394

## Device Facts

- **Applicant:** Implant Integration Systems, Inc.
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Oct 11, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The Implant Integration Systems Surgical Drills are intended to be used in the placement of the “Dual Integrator” Dental Implant (K944225).

## Device Story

Surgical drills used for site preparation during dental implant placement; specifically for Dual Integrator Dental Implant (K944225). Operated by dental surgeons in clinical settings. Drills facilitate bone preparation to receive implant. Benefit: enables precise surgical site creation for implant integration.

## Clinical Evidence

No clinical data.

## Technological Characteristics

Titanium nitride coated stainless steel surgical drills. Form factor: dental drill bits for bone preparation.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- Implant Innovations, Inc. Surgical Drills ([K891615](/device/K891615.md))
- Impla-Med Corp. Titanium/Stainless Steel Dental Implant Drills ([K902158](/device/K902158.md))
- Steri-oss Drills and Taps ([K915733](/device/K915733.md))

## Reference Devices

- Dual Integrator Dental Implant ([K944225](/device/K944225.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K962394

Implant Integration Systems, Inc.

6161 Clark Road, Suite 8 • Paradise, CA 95969 USA • (916) 872-1020 • Fax: (916) 877-7555

OCT 11 1996

Summary of Safety and Effectiveness Information

The Implant Integration Systems Surgical Drills are intended to be used in the placement of the “Dual Integrator” Dental Implant (K944225).

This product is similar in design, composition (titanium nitride coated stainless steel) and function to the Implant Innovations, Inc. Surgical Drills (K891615), the Impla-Med Corp. Titanium/Stainless Steel Dental Implant Drills (K902158) and the Steri-oss Drills and Taps (K915733).

![img-0.jpeg](img-0.jpeg)

6-19-96

)

---

**Source:** [https://fda-staging.innolitics.com/device/K962394](https://fda-staging.innolitics.com/device/K962394)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).