K962317 · Biomagnetic Technologies, Inc. · OLY · May 7, 1997 · Neurology
Device Facts
Record ID
K962317
Device Name
MAGNES 2500 WH BIOMAGNETOMETER
Applicant
Biomagnetic Technologies, Inc.
Product Code
OLY · Neurology
Decision Date
May 7, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The Magnes 2500 WH is intended for use in diagnostic procedures that require the measurement and display of extracranial magnetic fields and information about the electrical activity in the brain as inferred from those fields.
Device Story
Magnes 2500 WH biomagnetometer detects/measures extracranial magnetic fields produced by human brain. System uses array of 148 superconducting magnetic field pickup coils (magnetometers) and SQUIDs cooled by liquid helium via solid thermal conduction. Ancillary equipment includes magnetically shielded room, non-magnetic patient gantry/table, sensory stimulus systems (somatosensory, auditory, visual, motor), and head position measurement system. Operated by clinicians in medical facility. Computer workstation processes signals; displays results as waveforms, contour maps, or equivalent single current dipoles. Provides data for diagnosis or therapy planning; benefits patient by non-invasively mapping brain electrical activity.
Clinical Evidence
Bench testing only. Comparative studies using head phantoms evaluated dipole localization (location, strength, orientation), sensitivity under varying helium levels, and head position measurement accuracy. Results showed no significant differences between Magnes 2500 WH and Magnes II predicate.
Technological Characteristics
148-channel magnetometer array; SQUID sensors; liquid helium cooling via solid thermal conduction; magnetically shielded room (nickel-rich alloy/aluminum); non-magnetic gantry/table; sensory stimulus hardware; computer-based data acquisition/analysis; head position measurement via low-level transmitter coils.
Indications for Use
Indicated for patients whose physician requires information regarding brain-produced magnetic fields and source locations to assist in diagnosis or therapy planning.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}
K962317
MAY - 7 1997
# 510(k) Summary
This 510(k) Summary is being submitted pursuant to the requirement of 21 CF 807.92(c).
1. Submitted By: Biomagnetic Technologies, Inc.
9727 Pacific Heights Blvd
San Diego, CA 92121
Telephone: (619)453-6300
Fax: (619)453-4913
Contact: Eugene C Hirschkoff, Ph.D.
Director, Clinical Applications
Date prepared: April 23, 1996
2. Device trade name: Magnetic Encephalograph
Proprietary name: Magnes 2500 WH biomagnetometer
3. Substantial equivalence is claimed to the Magnes II biomagnetometer, manufactured and marketed by Biomagnetic Technologies, Inc. (Reference 510(k) K941553)
4. Description of Device: The Magnes 2500 WH biomagnetometer (hereafter "Magnes 2500 WH") comprises a magnetic sensor for detecting and measuring the magnetic fields produced by the human brain, along with the auxiliary equipment required to perform the measurements in a conventional medical facility environment and to display the results of the measurements to physicians in a variety of ways.
The sensor utilizes an array of superconducting magnetic field pickup coils arranged in such a manner as to sense the magnetic fields over the portion of the skull enclosing the brain. For each such coil, a superconducting quantum interference device (SQUID) is used to detect the current induced in that coil by the brain magnetic field and produce a voltage proportional to the magnetic flux change. Conventional electronic and computer circuitry is used to amplify, filter, digitize, store and display the result of the measurement. The sensor includes an insulated reservoir of liquid helium as a refrigerant for cooling the superconducting components - pickup coils, SQUIDs, and interconnecting leads - to temperatures below their superconducting transition temperature. Heat is conducted from these superconducting elements along thermally conductive pathways into the helium reservoir.
Provided as part of the Magnes 2500 WH biomagnetometer are the following ancillary items:
- Magnetically shielded room, comprised primarily of nickel-rich alloy and aluminum sheeting, to provide shielding from environmental sources of magnetic or rf noise
- Manually operated non-magnetic gantry to place the sensor over the head of the patient in either a seated or supine position
{1}
Non-magnetic patient table with hydraulic elevation, to support the patient securely in either seated or supine position
Non-magnetic patient monitoring and communication devices, including video monitor, intercom, and head motion detector
Head shape and head position measurement system, to provide head shape and location relative to the sensor for data modeling and display
Computer workstation, operator console, and software to control system operation, data acquisition and storage, and data analysis and display
Sensory stimulus systems, to provide stimulation of the patient's somatosensory, auditory, visual, and motor systems for magnetic measurement of evoked response
Results of the measurement of magnetic fields produced by the brain are available in tabular form and may be visualized as waveforms, contour maps of lines of equal magnetic field values, and as equivalent single current dipoles.
5. Intended Use: The Magnes 2500 WH is intended for use in diagnostic procedures that require the measurement and display of extracranial magnetic fields and information about the electrical activity in the brain as inferred from those fields.
6. Technological Characteristics: The Magnes 2500 WH operates with the same principles of operation as the predicate device, the Magnes II biomagnetometer. Magnetic field changes induce currents in superconducting pickup coils, those currents are detected and converted to voltages by superconducting quantum interference devices or SQUIDs, and those voltages are amplified, filtered, digitized, stored, analyzed and displayed for use by a physician. The primary technological difference between the Magnes II and the Magnes 2500 WH is that the latter features 148 pickup coils distributed over the entire surface of the skull overlying cortical tissue while the former features two arrays of 37 pickup coils, each covering a circular area approximately $15\mathrm{cm}$ in diameter.
Because the cryogenic container housing the pickup coils is by necessity rigid and head sizes and shapes vary from patient to patient, the distance from some of the pickup coils to the skull of a patient may be greater than with the Magnes II. To compensate for this, the pickup coils are constructed as magnetometers rather than axial gradiometers as in the Magnes II. Magnetometer coils offer greater sensitivity to brain sources of magnetic fields than gradiometer coils but at the expense of greater sensitivity to extrinsic magnetic noise. A secondary set of pickup coils, spatially separated from the primary pickup coils, are used to detect magnetic noise and subtract it from the primary signals, creating the functional equivalent of an axial gradiometer.
The Magnes II refrigerates one set of the superconducting components of the sensor by immersing them directly in liquid helium, the other by solid thermal conduction. In the Magnes 2500 WH, all of the superconducting components are refrigerated by solid thermal conductors which thermally connect each such component to the liquid helium reservoir.
In the Magnes II, the position of the patient's head was measured using a third-party position indicator system which operated by essentially measuring the mutual inductance between small receiver coils mounted on the patient's head or in a wand and a low level rf transmitter coil mounted on one sensor housing. In the Magnes 2500 WH, the same principle of operation is
{2}
employed, but small very low level transmitter coils are mounted on the patient's head and the field produced by those coils is measured by the biomagnetometer pickup coils themselves.
7. Non-clinical tests: The substantial equivalence between the Magnes II and Magnes 2500 WH was verified by comparative non-clinical tests of each of the technological characteristics which are different, as described in 6) above.
To demonstrate the substantial equivalence of the magnetometer pickup coils as used in the Magnes 2500 WH to the gradiometer coils as used in the Magnes II, two bench tests were run. In the first, a dipolar source in a head phantom was activated and measured with each of the sensors, both in the normal ambient environment of the factory and in the presence of artificially produced magnetic noise. The spectral content of the waveform as measured with each sensor were then compared for both conditions. The waveforms showed no significant difference from each other, in either operating condition.
As a second test of this characteristic, the dipole in the head phantom was localized using the Magnes II and Magnes 2500 WH sensors in a number of trials. There was no significant difference for any of the dipole parameters between the two sensors - physical location, strength, or orientation.
To demonstrate the substantial equivalence of the method of refrigeration, the magnetic field produced by a fixed source was detected by representative pickup coils in the Magnes 2500 WH when the helium level in the reservoir was full and when it was nearly empty. A similar test was conducted with the upper sensor of a Magnes II, in which the direct immersion method of refrigeration is used. Sensitivity of the various channels was compared for these two operating conditions and no material difference measured.
Finally, to demonstrate the substantial equivalence of the methods of measuring head position, repetitive measurements of the location of reference points on a head phantom were conducted with the positioning system in the Magnes II and with the new approach taken in the Magnes 2500 WH. No significant difference between the relative locations of the reference points was found.
8. Conclusion: The comparative non-clinical tests produced no significant differences between those elements of the Magnes II and the correspondent elements of the Magnes 2500 WH for which there has been some change in technological approach. Since all other elements of the two systems are technologically the same, we conclude that the Magnes 2500 WH is substantially equivalent to the predicate Magnes system.
Please direct any questions concerning this 510(k) Summary to the contact person noted above.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
10903 New Hampshire Avenue
Document Control Room -WO66-G609
Silver Spring, MD 20993-0002
Eugene C. Hirschkoff, Ph.D.
Director of Clinical Applications
Biomagnetic-Technologies
9727 Pacific Heights Boulevard
San Diego, California 92121-3719
Re: K962317
Trade/Device Name: Magnes 2500 WH Biomagnetometer
Regulation Number: 21 CFR 882.1400
Regulation Name: Electroencephalograph
Regulatory Class: II
Product Code: OLY
Dated (Date on orig SE ltr): March 5, 1997
Received (Date on orig SE ltr): March 6, 1997
Dear Mr. Hirschkoff:
This letter corrects our substantially equivalent letter of May 7, 1997.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
APR - 9 2012
{4}
Page 2 – Mr. Eugene C. Hirschkoff
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health’s (CDRH’s) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

Malvina B. Eydelman, M.D.
Director
Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
510(k) Number (if known): K962317
Place Name: Magnes 2500 WH Biomagnetometer
Indications For Use:
Use of the Magnes 2500 WH is indicated for the patient whose physician believes that information about the magnetic fields produced by that patient's brain and information about the location of the sources of those magnetic fields could contribute to diagnosis or therapy planning.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Of)
Division of Cardiovascular Respiratory and Neurological Devices
510(k) Number K962317
Prescription Use ☑
Per CFR 801.109
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.