ENDOSONICS AUTOMATIC PULL BACK DEVICE

K962293 · Endosonics Corp. · ITX · Sep 16, 1996 · Radiology

Device Facts

Record IDK962293
Device NameENDOSONICS AUTOMATIC PULL BACK DEVICE
ApplicantEndosonics Corp.
Product CodeITX · Radiology
Decision DateSep 16, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1570
Device ClassClass 2

Indications for Use

The Endosonics Automatic PBD is intended for use as a pull back accessory to the Endosonics Intravascular Ultrasound Imaging catheters. The Automatic Pull Back Device withdraws the imaging catheter from the vessel through the guide catheter.

Device Story

Battery-powered accessory for intravascular ultrasound (IVUS) imaging; used in catheterization laboratory by clinical staff. Device provides uniform, slow withdrawal of imaging catheter from vessel through guide catheter. Linear travel distance ~12 cm; selectable pull-back speeds of 0.5 mm/sec or 1 mm/sec. No direct/indirect patient contact. Facilitates consistent imaging data acquisition by ensuring uniform catheter movement, aiding physician assessment of vessel morphology.

Clinical Evidence

Bench testing only; no clinical data provided. Device is an accessory with no patient contact.

Technological Characteristics

Battery-powered electromechanical device. Linear travel distance ~12 cm. Selectable pull-back rates: 0.5 mm/sec or 1 mm/sec. Constructed of standard medical-grade materials. No patient contact. Standalone operation.

Indications for Use

Indicated for use as a pull back accessory to Endosonics Intravascular Ultrasound Imaging catheters to withdraw the imaging catheter from the vessel through the guide catheter during intravascular ultrasound assessment in the catheterization laboratory.

Regulatory Classification

Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

Predicate Devices

Submission Summary (Full Text)

{0} Endosonics Corporation Automatic Pull Back Device 510(k) Premarket Notification June 13, 1996 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS ## ENDOSONICS Automatic Pull Back Device **K962293** SEP 16 1996 As an accessory, the Automatic Pull Back Device (PBD) does not affect the safe and effective use of the Endosonics Intravascular Ultrasonic Imaging System or the Endosonics Imaging catheters. ## DETERMINATION OF SUBSTANTIAL EQUIVALENCE The Automatic PBD is substantially equivalent in intended use and technologic characteristics to cardiovascular accessory devices: the CVIS Catheter Pull-Back Device and the Intertherapy Interpret Catheter Linear Translator. These predicate devices are legally commercially available. ## COMPANY AND CONTACT PERSON Endosonics Corporation 3078-B Prospect Park Drive Rancho Cordova, CA 95670 Adam Savakus Vice President, Clinical and Regulatory Affairs 916-638-8008 ## DEVICE NAME Automatic Pull Back Device ## PREDICATE DEVICE(S) 1. CVIS Catheter Pull-Back Device 2. Intertherapy Interpret Catheter Linear Translator -- Page 33 -- {1} Endosonics Corporation Automatic Pull Back Device 510(k) Premarket Notification June 13, 1996 ## DESCRIPTION OF DEVICE The Automatic PBD is a battery powered device that is used in the catheterization laboratory during intravascular ultrasound assessment. The Automatic PBD has a nominal linear travel distance of approximately 12 cm and provides a uniform, slow pull back rate of the catheter’s imaging element. The speed is selectable between 1 mm/sec or 0.5 mm/sec. The Auto PBD has no direct or indirect patient contact. The device is constructed of materials common to the medical industry for both patient and non-patient contact devices and equipment. ## STATEMENT OF INTENDED USE OF THE DEVICE The Endosonics Automatic PBD is intended for use as a pull back accessory to the Endosonics Intravascular Ultrasound Imaging catheters. The Automatic Pull Back Device withdraws the imaging catheter from the vessel through the guide catheter. ## INTENDED USE OF PREDICATE DEVICE 1. The Auto PBD is intended to be used as a pull-back accessory to CVIS intravascular ultrasound catheters. The Auto PBD pulls back the imaging element inside the catheter while the outer catheter body is stationary in the vessel. 2. The CVIS Catheter Pull-Back Device is intended to be used for uniform, slow velocity pull-back of intravascular ultrasound catheters. 2. The Intertherapy Interpret Catheter Linear Translator is intended to be used for uniform, slow velocity pull-back of intravascular ultrasound catheters. ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS BETWEEN DEVICE AND PREDICATE DEVICE The Automatic PBD and the predicate devices have substantially equivalent technological characteristics (materials, design, specifications, and mode of operation). These accessory devices have no patient contact. The devices are designed so catheters will attach to a movable carriage. The Automatic PBD incorporates pull-back rates comparable to predicate devices. -- Page 34 --
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...