← Product Code [JDI](/productcode/JDI) · K962267

# PERFECTA PS REVISION STEM (K962267)

_Wrightmedicaltechnologyinc · JDI · Sep 10, 1996 · Orthopedic · SN_

**Canonical URL:** https://fda-staging.innolitics.com/device/K962267

## Device Facts

- **Applicant:** Wrightmedicaltechnologyinc
- **Product Code:** [JDI](/productcode/JDI.md)
- **Decision Date:** Sep 10, 1996
- **Decision:** SN
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3350
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The PERFECTA® Revision Hip System is intended to be used in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

## Device Story

Monolithic femoral component for total hip arthroplasty; titanium alloy (ASTM F-136) construction with plasma spray coating; available in straight or bowed configurations; features calcar platform and coronal slot (select sizes); press-fit or cemented implantation; used by orthopedic surgeons in clinical/OR settings; provides structural support for hip joint reconstruction; benefits patients via pain relief and improved joint function.

## Clinical Evidence

Bench testing only. Abrasion testing demonstrates plasma spray coating comparability to sintered bead coating. Fatigue testing confirms device withstands simulated in vivo loading.

## Technological Characteristics

Material: Titanium alloy (ASTM F-136). Coating: Titanium plasma spray. Design: Monolithic femoral stem, straight or bowed, with calcar platform and coronal slot. Fixation: Press-fit or cemented. No software or electronic components.

## Regulatory Identification

A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- PERFECTA® Hip System ([K962267](/device/K962267.md))
- PERFECTA® Revision Hip System ([K962267](/device/K962267.md))
- Integral Total Hip System (Biomet)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SENT BY: 9-8-96 :11:05AM :WRIGHT MEDICAL TECH. 301 594 2358;# 3/ 3
K962267
WRIGHT
MEDICAL TECHNOLOGY, INC.
5677 AIRLINE ROAD
ARLINGTON, TN 38002
901-867-9971
SEP 10 1996

Contact Person: Cristie Manuel
Date: September 8, 1996

# 510(k) Summary

Trade/Proprietary Name: PERFECTA® PS Revision Hip Stem
510(k) Number: K962267
Common Name: Plasma Spray Coated Femoral Hip Stem
Classification: Class II
Predicate Devices: PERFECTA® Hip System and PERFECTA® Revision Hip System manufactured by Wright Medical Technology, Inc. and Integral Total Hip System manufactured by Biomet

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

## Description/Intended Use

The PERFECTA® Revision Hip System is a monolithic design femoral component. The device is manufactured from titanium alloy (ASTM F-136) with a titanium plasma spray coating. The stem is available in two configurations: straight and bowed. All configurations have a calcar platform, from 2.0cm to 3.75cm. All sizes except the 12mm x 190mm have a coronal slot. The stem is designed to be press-fit or can be used with bone cement.

The PERFECTA® Revision Hip System is intended to be used in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

## Substantial Equivalence

The PERFECTA® PS Revision Hip Stem is substantially equivalent to the PERFECTA® Hip System and the PERFECTA® Revision Hip System manufactured by Wright Medical Technology, Inc. and the Integral Total Hip System manufactured by Biomet.

## Testing

- Submitted abrasion testing demonstrates that the plasma spray porous coating is comparable to sintered bead coating.
- Submitted fatigue testing demonstrates that the PERFECTA® PS Revision Hip Stem was able to withstand test loading designed to simulate in vivo loading.

---

**Source:** [https://fda-staging.innolitics.com/device/K962267](https://fda-staging.innolitics.com/device/K962267)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).