← Product Code [HWC](/productcode/HWC) · K962236

# SCARF THREAD-HEAD SCREW (K962236)

_Landos, Inc. · HWC · Sep 20, 1996 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K962236

## Device Facts

- **Applicant:** Landos, Inc.
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** Sep 20, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The SCARF THREAD-HEAD™ SCREW is intended to be implanted for fixation of bone fractures or bone reconstructions.

## Device Story

Cannulated bone fixation screw; titanium alloy construction; lengths 10-32 mm. Used by orthopedic surgeons for internal fixation of foot osteotomies. Provides mechanical stability to bone fragments during healing. Implanted via surgical procedure; cannulated design allows placement over guidewire. Benefits patient by maintaining bone alignment post-osteotomy.

## Clinical Evidence

Bench testing only. Rupture torque testing performed per French Standard NF-F-90-414; device demonstrated compliance with torsion resistance requirements.

## Technological Characteristics

Cannulated bone screw; titanium alloy; lengths 10-32 mm (2 mm increments); hexagonal socket head; threaded design.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Medinov-AMP Scarf Screw ([K931155](/device/K931155.md))
- Herbert-Whipple Bone Screw (Zimmer) ([K792022](/device/K792022.md))
- M3 Screw System (Osteomed) ([K924018](/device/K924018.md), 924138)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Landos Inc. K962236
A SUBSIDIARY OF GROUP LANDANGER CAMUS
SEP 20 1996

# 3. SUMMARY OF SAFETY AND EFFECTIVENESS

## MANUFACTURER IDENTIFICATION:
Medinov - AMP
Z.I. Bapaume
27-31 rue Lucien Langénieux
42335 Roanne Cedex
FRANCE

## SPONSOR IDENTIFICATION
LANDOS INC.
Mr. Jean-Paul Burtin
100 Lindenwood Drive
Malvern, PA 19355

## ESTABLISHMENT REGISTRATION NUMBER:
2530066

## OFFICIAL CONTACT PERSON:
Norman F. Estrin, Ph.D., RAC
President
Estrin Consulting Group, Inc.
9109 Copenhaver Drive
Potomac, MD 20854
Tel.: (301) 279-2899
Fax: (301) 294-0126

## DATE OF PREPARATION OF THIS SUMMARY:
May 22, 1996

## TRADE NAME:
SCARF THREAD-HEAD™ SCREW

## COMMON NAME:
Bone Fixation Screw

## CLASSIFICATION NAME AND REFERENCE:
Screw, Fixation, Bone
(21 CFR, Section 888.3040)
Class II

## PROPOSED REGULATORY CLASS:
87HWC

## DEVICE PRODUCT CODE:
87HWC

## PANEL CODE:
87OR

100 Lindenwood Drive • Malvern, PA 19355
Telephone: 610-644-4670 • Fax: 610-644-4429
00001C

{1}

000017

DESCRIPTION OF DEVICE:

The SCARF THREAD-HEAD™ SCREW is a cannulated thread-head screw made of a Titanium alloy. Screws come in lengths of 10-32 mm in 2 mm increments.

INTENDED USE:

The SCARF THREAD-HEAD™ SCREW is intended to be implanted for fixation of bone fractures or bone reconstructions.

INDICATIONS FOR USE:

The SCARF THREAD-HEAD™ SCREW is indicated for fixing the elective osteotomies of the mid-foot bones and the metatarsal and phalanges of the foot only.

PREDICATE DEVICE:

The SCARF THREAD-HEAD™ SCREW is substantially equivalent to the Medinov-AMP Scarf Screw (K931155), the Herbert-Whipple Bone Screw (Zimmer) (K792022) and the M3 Screw System (Osteomed) (K924018, 924138).

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:

Both the SCARF THREAD-HEAD™ SCREW, the Scarf Screw and the Herbert-Whipple Bone Screw have the same intended use and all are indicated for fixing small fractures or osteotomies. All are made from Titanium alloys. The SCARF THREAD-HEAD™ SCREW and Scarf Screws are threaded on nearly all their length with the same thread pitch. Both have a thread head and a hexagonal socket. The Scarf screw is not cannulated; the Herbert-Whipple screw is cannulated and is topped with a hexagonal socket.

SUMMARY OF STUDIES:

Rupture torque of the SCARF THREAD-HEAD™ SCREW was compared with requirements of the French Standard N° NF-F-90-414 and found to have a resistance to torsion in compliance with the selected Standard.

---

**Source:** [https://fda-staging.innolitics.com/device/K962236](https://fda-staging.innolitics.com/device/K962236)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).