The GEN-PROBE AMPLIFIED Chlamydia Trachomatis Assay (AMP CT) Kit is a target-amplified nucleic acid probe test used for the in vitro diagnostic qualitative detection of rRNA from Chlamydia trachomatis in endocervical and male urethral swab specimens and in female and male urine specimens.
Device Story
The AMP CT assay is an in vitro diagnostic test for detecting Chlamydia trachomatis rRNA in clinical specimens. It utilizes Transcription Mediated Amplification (TMA) to amplify target rRNA and the Hybridization Protection Assay (HPA) for detection. In HPA, an acridinium ester-labeled DNA probe hybridizes to the target amplicon; a selection reagent then hydrolyzes unhybridized probes, leaving only hybridized probes to emit chemiluminescent light upon the addition of detection reagents. The device is used in clinical laboratory settings by trained personnel. The output is a light emission measurement (Relative Light Units, RLU) compared against a 50,000 RLU cut-off to determine a positive or negative result. This provides clinicians with a rapid, sensitive diagnostic tool to identify C. trachomatis infections, facilitating prompt patient management and treatment, and helping to control disease transmission.
Clinical Evidence
Multicenter clinical study of 1,897 swab and 2,139 urine specimens from 2,295 patients. Compared to tissue culture and DFA. Overall sensitivity 93.6% (349/373) and specificity 98.0% (3,591/3,663). Performance varied by specimen type and population (symptomatic/asymptomatic). Analytical performance included limit of detection (1 IFU/assay) and specificity testing against 115 culture isolates (bacteria, fungi, viruses). Interference studies evaluated blood, mucus, lubricants, and common urine constituents.
Technological Characteristics
In vitro nucleic acid amplification (TMA) and target amplicon hybridization with acridinium ester-labeled DNA probe (HPA). Qualitative detection of C. trachomatis rRNA. Cut-off: 50,000 RLU. Detects as few as 20 copies of rRNA (1 IFU).
Indications for Use
Indicated for the qualitative detection of Chlamydia trachomatis rRNA in endocervical and male urethral swab specimens and female and male urine specimens in symptomatic and asymptomatic patients.
Regulatory Classification
Identification
Chlamydia serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to chlamydia in serum. Additionally, some of these reagents consist of chlamydia antisera conjugated with a fluorescent dye used to identify chlamydia directly from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Chlamydia and provides epidemiological information on these diseases. Chlamydia are the causative agents of psittacosis (a form of pneumonia), lymphogranuloma venereum (a venereal disease), and trachoma (a chronic disease of the eye and eyelid).
Predicate Devices
Tissue culture media and components
Submission Summary (Full Text)
{0}
NOV 27 100r
GEN-PROBE
K962217
# 510(k) SUMMARY
## SUMMARY OF SAFETY AND EFFECTIVENESS
### GEN-PROBE AMPLIFIED Chlamydia Trachomatis Assay Kit
#### General Information
Submitter’s Name: **Company Name/Contact**
Gen-Probe Incorporated
9880 Campus Point Drive
San Diego, California 92121-1589
(619) 546-8000
(619) 546-7979 (FAX)
Company Contact:
Gerald H. Schell, M.S., RAC
Director, Regulatory Affairs
(619) 535-2862
(619) 546-7979(FAX)
Date Prepared: June 9, 1996
#### Device Name
Trade Name:
GEN-PROBE® AMPLIFIED Chlamydia Trachomatis Assay Kit
Common or Usual Name:
rRNA Target-amplified nucleic acid probe test for the *in vitro* diagnostic detection of *Chlamydia trachomatis*
Classification Names:
*Chlamydia*, DNA reagents Serological Reagents (Microbiology Classification Device List)
000042
Gen-Probe Incorporated • 9880 Campus Point Drive • San Diego, California 92121-1589
(619) 546-8000 • FAX: (619) 452-5848
{1}
Classification Codes:
Class I
Panel: Microbiology
Number: CFR 866.3120
Name: Chlamydia Serological Reagents
Reagents used to identify Chlamydia directly from clinical specimens and/or cultured isolates derived from clinical specimens.
## Substantially Equivalent Device
Substantially Equivalent Device: Tissue culture media and components, synthetic cells and tissue culture.
## Device Description
## Background on the Disease and Principle of the Test
## The Disease
C. trachomatis infections are a major cause of sexually transmitted diseases (STD) in the United States. More than four million new cases of chlamydial infections occur annually [CDC, 1993]. In humans, C. trachomatis species are responsible for infections of the cervix, urethra, and upper genital tract in women; infections of the urethra and epididymitis in men; and conjunctivitis and pneumonia in newborns [Schacter, J. and M. Grossman, 1981]. The Chlamydia trachomatis species is comprised of 15 serovars. Human isolates of C. trachomatis are grouped into trachoma (serovars A-K) and LGV biovars (serovars LGV I, LGV II, and LGV III). The serovars D through K are the major cause of genital chlamydial infection in men and women [Schacter, J. 1978].
## Historical and Conventional Methods
Tissue culture methods remain the gold standard for diagnosing infection with these organisms, and they are the reference methods to which new nonculture assays are compared. Tissue culture methods include isolation in cell culture and demonstration of typical intracytoplasmic inclusions by appropriate immunofluorescence, iodine, or Giemsa staining procedures. Culture has several merits. For C. trachomatis, culture is considered 100% specific because the organism is directly observed. As well, specimen adequacy can be assessed by microscopically examining a sample of the culture transport medium. Culture is the sole method recommended in forensic cases. Despite its merits, culture has several disadvantages. C. trachomatis culture is expensive, time and labor intensive, and very dependent on the training and skill level of the technicians involved.
000043
{2}
Specimen transport issues and the requirement for organism viability are the main disadvantages of C. trachomatis culture.
Non-culture diagnostic tests for C. trachomatis make more aggressive screening and prevention strategies possible because of their ease of use, shorter time-to-result, and absence of viability requirements. Nonculture methods for C. trachomatis detection include direct fluorescent antibody staining (DFA), enzyme immunoassay (EIA), nonamplified nucleic acid hybridization (probe assays), polymerase chain reaction (PCR), ligase chain reaction (LCR), and transcription mediated amplification (TMA), which is the most recently developed target-amplified nucleic acid test for C. trachomatis.
Target nucleic acid-based amplification assays are the newest non-culture methods available [Wolcott, M.J., 1992]. Amplification technology allows for detection of lower numbers of organisms as compared to other nonculture methods due to the higher sensitivity inherent in the technology. This allows for detection of infection in individuals harboring much lower levels of organisms than are detected by nonamplified nonculture methods. Because of their high sensitivity, all target-amplified nucleic acid assays for C. trachomatis are capable of detecting very low levels of nucleic acid, including that from nonviable cells. However, because C. trachomatis is a pathogen, the presence of any detectable nucleic acid indicates a biological condition that requires treatment. Amplified assays cannot be used, however, to monitor therapeutic progress because they detect nonviable organisms.
## Patient Care and Public Health Implications
Screening for the presence of chlamydial infections is the cornerstone for prevention strategies. A large number of infections may be asymptomatic or have symptoms that are not specific; this makes reduction of the prevalence of chlamydial infections difficult. Accurate and prompt diagnoses of these infections is important to ensure appropriate patient management, to prevent disease complications and their associated medical costs, and to control transmission to uninfected partners.
## Intended Use
The GEN-PROBE AMPLIFIED Chlamydia Trachomatis Assay (AMP CT) Kit is a target-amplified nucleic acid probe test used for the in vitro diagnostic qualitative detection of rRNA from Chlamydia trachomatis in endocervical and male urethral swab specimens and in female and male urine specimens.
000044
{3}
Summary of Technological Characteristics
The AMP CT assay kit incorporates the technologies of in vitro nucleic acid amplification and target amplicon hybridization with an acridinium ester-labeled DNA probe to specifically detect C. trachomatis ribosomal ribonucleic acid (rRNA) in clinical specimens. Gen-Probe's proprietary technologies are combined in this product to allow qualitative detection of C. trachomatis rRNA in quantities as minute as those contained in a single organism. The presence of C. trachomatis rRNA provides evidence of the organisms's presence in a clinical specimen in the same way as does the observation of C. trachomatis on a coverslip from tissue culture.
## Principles of the AMP CT Technology
The AMP CT assay utilizes Transcription Mediated Amplification (TMA) and the Hybridization Protection Assay (HPA) to qualitatively detect C. trachomatis (rRNA). TMA is an RNA transcription-dependent amplification technology, in which RNA strands serve as templates for the synthesis of DNA intermediates. These DNA intermediates are then used for the transcription of multiple copies of RNA amplicon. The RNA amplicon can then serve as templates for further synthesis of DNA intermediates, which in turn are used for further transcription of copies of RNA amplicon. The AMP CT assay uses TMA to amplify C. trachomatis rRNA in clinical specimens to easily detectable levels. HPA then is used to detect the amplicon.
## Mechanism of the Hybridization Protection Assay (HPA)
The key aspect of HPA is the steric protection of the acridinium ester chemiluminescent label, linked to the Detection Probe, from a hydrolysis reaction that destroys the chemiluminescence of the label. This protection occurs only when the Detection Probe hybridizes with the complementary amplicon sequence. Thus, when a selection reagent is added to a solution containing a mixture of hybridized and unhybridized Detection Probes, only the hybridized probes will maintain their chemiluminescent properties. Following this "differential hydrolysis" step, Detection Reagents are added to the solution and the acridinium ester molecules that are attached to the hybridized Detection Probes emit photons. Thus, photons are emitted only if the target nucleic acid is present, and this light emissions constitutes a positive result [Arnold, L.J., 1989 and Nelson, N.C., 1992].
## Potential Adverse Effects of the Device on Health
When this in vitro device is used as indicated, there are no known adverse effects on health. Failure of this device would produce either a false positive result or a false negative result. A false positive result occurs when the test result is above the test cut-off even though C. trachomatis rRNA is not present. A false negative result occurs when the
000045
{4}
test result is below the test cut-off even though rRNA specific to C. trachomatis is present.
A false positive result could lead to misdiagnosing a patient’s medical status, resulting in the administration of unnecessary therapy. A false negative result could delay the correct diagnosis of C. trachomatis infection. Misdiagnosis of a patient who has a C. trachomatis infection could result in a delay of antibiotic therapy. In addition, the sexual partners of an infected patient might not be notified.
False positive or false negative results can be caused by incorrect specimen collection and storage, sample misidentification, presence of low numbers of C. trachomatis in the specimen, specimen inhibition, interfering substances, use of the test by unqualified personnel, procedural errors and deviations, insufficient mixing, carryover contamination, transcription errors, or inappropriate test result reporting.
## Summary of Non-Clinical (Analytical Laboratory) Performance Data
## Determination and Rationale for Establishing the Cut-Off
A total of 130 C. trachomatis positive and 982 negative clinical samples were tested using the AMP CT assay to establish the assay cut-off. A total of 431 swabs (45 positive and 386 negative) and 681 urine samples (85 positive and 596 negative) were tested. Samples were considered positive based on culture and DFA. The data, analyzed by Receiver Operating Characteristic (ROC) and Decision Level (DL) curves, indicated a cut-off value between 10,000 RLU and 50,000 RLU was appropriate. A cut-off of ≥50,000 RLU was selected for both sample types to minimize false positive and false negative results. The selection of this cut-off value was validated through the clinical trial.
## Limits of Detection (Analytical Sensitivity)
To establish the limits of detection in terms of rRNA copies, a purified stock of C. trachomatis rRNA was serially diluted and assayed in pure system (transport medium) and in swab and urine specimens. The data indicated that the AMP CT assay was capable of reliably detecting as few as 20 copies (0.05 fg) of rRNA. Twenty (20) copies of rRNA is equivalent to 1/100 of the rRNA in a C. trachomatis organism.
To establish the limits of detection in terms of cultured C. trachomatis, the 15 serovars were grown in McCoy cell monolayers and harvested to yield stocks. The stocks were diluted in a buffered salt solution to yield low numbers of infective units. Equal volume aliquots of each dilution were used to infect 2 fresh McCoy cell monolayers and to inoculate Gen-Probe transport medium. Culture IFU counts were made by DFA-staining the monolayers. Dilutions of the last culture-positive dilution were also made in order to determine the lower limit of detection for the AMP CT assay. The AMP CT assay was
000046
{5}
able to detect fractional equivalents of IFU for all serovars. However, because 1 IFU is the lowest clinically relevant detection point, the AMP CT limit of detection was set at 1 IFU per assay in the AMP CT the labeling.
## Analytical Specificity
A total of 115 culture isolates were evaluated using the AMP CT assay. These isolates included 72 organisms that may be isolated from the urogenital tract and 43 additional organisms that represent a phylogenetic cross-section of organisms. The tested organisms included bacteria, fungi, yeast, and parasites. In addition, the following viruses were tested: (1) Herpes Simplex I, (2) Herpes Simplex II, (3) Cytomegalovirus, and (4) Human Papilloma Virus Type 16. Culture isolates of *C. trachomatis* (15 serovars), *C. psittaci* and *C. pneumoniae* were tested. The bacterial and fungal organisms were tested at concentrations of $2.5 \times 10^{5}$ cells/assay; 5 organisms were tested using purified rRNA (2.4 to $23.8\ \mathrm{ng}/\mathrm{assay}$) because cell lysates were not available. Cytomegalovirus was tested using infected culture cells at a concentration of $4.8 \times 10^{5}$ cells/assay. ATCC stock preparations of the viruses Herpes Simplex I and Herpes Simplex II were tested at the supplied concentrations of $2.5 \times 10^{4}\ \mathrm{TCID}_{50}$ and $6.0 \times 10^{4}\ \mathrm{TCID}_{50}$, respectively. Purified DNA ($2.0 \times 10^{6}$ copies/assay) was used to test Human Papilloma Virus Type 16. Only the *C. trachomatis* samples produced positive results in the AMP CT assay. Urine specimens from 89 *Chlamydia* culture negative patients reporting symptoms of urinary tract infections (UTI) were also tested. No cross-reactions were observed in the AMP CT assay.
## Interference Studies
To determine whether non-target organisms would interfere with the AMP CT assay, *C. trachomatis* rRNA (0.5-500 fg/assay) was tested in the presence and absence of *E. coli*, *N. gonorrhoeae*, *G. vaginalis*, which are organisms that can be found in urogenital specimens. Specimens from patients with UTI symptoms were also tested since they likely contained a wide range of organisms naturally encountered in clinical specimens. All positive AMP CT reactions remained positive even in the presence of 240,000 cells/assay of the non-target organisms and in the presence of the naturally occurring UTI organisms.
In laboratory studies, the presence of up to $12\%$ (v/v) blood in swab specimens and up to $0.2\%$ (v/v) blood in urine specimens did not interfere with the AMP CT assay. In the clinical studies, even grossly bloody swab and urine specimens did not yield assay interference. Mucus, gynecological lubricants, and spermicides in swab specimens did not interfere with the assay. A hemorrhoidal anesthetic and a body oil also did not interfere when tested at elevated concentrations of $10\%$ (w/v); loss of positive signal recovery occurred for both agents when a vast excess ($50\%$ w/v) was tested. Protein, glucose, ketones, bilirubin, nitrite, urobilinogen, vitamins, minerals, and over-the-counter pain relievers did not result in high background or prevent recovery of signal in urine specimens when present at normal or elevated levels. The assay tolerated a wide range
000047
{6}
of urine pH (i.e., 4 to 9). Leucocytes at $7 \times 10^{6} / \mathrm{mL}$ ($1.2 \times 10^{6}$/assay) [highest level tested] and excessive levels of cellular debris in urine specimens did not interfere.
Talcum powder and feminine hygiene spray interfered with the assay by elevating the background signals and reducing positive signal. These agents should not be used prior to specimen collection. The effects of tampon use and douching were not assessed.
## Precision
A panel of 10 clinical swab and urine specimens, consisting of 2 negatives, 4 low positives (avg. $\leq 10$ IFU in culture), and 4 high positives (avg. $\geq 50$ IFU in culture), were tested. All samples were natural positives or negatives with the exception of 1 low and 1 high positive, these samples were prepared by adding purified C. trachomatis IFU to culture- and amplification-negative specimens. The panel was tested in a random order in triplicate twice a day for 3 days. Positive and negative amplification controls were included in each run. All samples yielded the expected results for each replicate in each run; no false positive or false negative results occurred.
## Reproducibility/Proficiency
As part of the clinical studies, a proficiency panel consisting of 3 positive swab, 3 positive urine, 2 negative swab, and 2 negative urine samples was tested in triplicate by 2 or 3 operators at each of 5 sites on 3 consecutive days. The operators were trained on the assay just prior to starting proficiency runs. Of the 12 operators, 11 operators generated $100\%$ correct panel results. One (1) operator at 1 site had 1 false result on Day 3 testing; all results were correct after the operator repeated the entire 3 day panel.
## Clinical Study Design Summary
A multicenter clinical study was conducted to assess the performance characteristics of the GEN-PROBE AMPLIFIED Chlamydia Trachomatis (AMP CT) Assay Kit in identifying C. trachomatis in endocervical and male urethral specimens, and in female and male urine specimens, as compared to the Predicate Device (i.e., tissue culture media and components using standard laboratory culture methods for detection of C. trachomatis). Resolution of AMP CT apparent false positive results included reculture, direct immunofluorescence staining (DFA) of the culture transport media, or DFA of urine (urine discrepants only).
The AMP CT assay was evaluated using 1,897 swab and 2,139 urine specimens from 2,295 patients (both symptomatic and asymptomatic) at 5 geographically distributed sites. This total included 1,510 females and 785 males. Of the females evaluated, 489 (327 symptomatic and 162 asymptomatic) represented high prevalence populations, and 1,021 (245 symptomatic and 776 asymptomatic) represented low prevalence populations. Males evaluated included 530 symptomatic patients and 255 asymptomatic patients.
000048
{7}
Prevalence of C. trachomatis by culture ranged from 4.0% to 13.6% with an overall prevalence of 6.8%.
Patients enrolled in the study represented all patients evaluated for C. trachomatis using tissue culture media and components at the participating clinical facilities during the study period, providing the inclusive study criteria were satisfied.
Symptomatic patients were defined as patients reporting symptoms consistent with a possible C. trachomatis infection at the time of the clinic visit. Asymptomatic patients were defined as: (i) patients not reporting symptoms at the time of the clinic visit, but known to be a sexual partner or contact of a person with a confirmed or suspected C. trachomatis infection or (ii) patients undergoing screening evaluations for possible C. trachomatis infection. Any clinical signs of an infection found by the clinician at the time of examination were recorded, but did not alter the classification of the patient as asymptomatic.
The AMP CT assay protocol in the Package Insert was followed. Positive controls and negative controls were included in each run. AMP CT specimen results that were within the range of 15,000 to 200,000 RLU were retested in duplicate at the study site to evaluate the need for an assay equivocal range (retest zone).
The clinical performance of the AMP CT assay was characterized by sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV). Sensitivity and specificity were evaluated based on the pre-selected 50,000 RLU cut-off.
## Clinical Study Results Summary
Receiver Operating Characteristic (ROC) and Decision Level (DL) curves of the clinical data supported the validity of the pre-selected cut-off of 50,000 RLU using the clinical data. The results of the equivocal zone testing indicated that a equivocal (retest) zone was not warranted for the AMP CT assay.
The sensitivity and specificity of the AMP CT assay were determined by comparing the assay results to C. trachomatis tissue culture results. Specimens that were AMP CT positive/culture negative were recultured, tested by DFA of the culture transport medium, and assayed by urine DFA (urine discrepants only).
000049
{8}
AMP CT Assay Performance Summary Compared to Culture and DFA Testing
| | AMP CT: | Pos | Pos | Pos | Pos | Neg | Neg | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Culture: | Pos | Neg | Neg | Neg | Pos | | | Neg |
| | | DFA: | N/A | Pos | Neg | Neg | N/A | | | N/A |
| | | Urine DFA: | N/A | N/A | Pos | Neg | N/A | N/A | Sensitivity (95% CI) | Specificity (95% CI) |
| Sample Type by Population | | N | | | | | | | 98.1% (52/53) (93.5-100) | 98.2% (501/510) (97.0-99.5) |
| Female Endocervical | Symptomatic | 563 | 35 | 17 | N/A | 9 | 1 | 501 | 100% (53/53) (99.0-100) | 98.8% (870/881) (98.0-99.5) |
| | Asymptomatic | 934 | 44 | 9 | N/A | 11 | 0 | 870 | 100% (53/53) (99.0-100) | 98.6% (1,371/1,391) (97.9-99.2) |
| | TOTAL | 1,497 | 79 | 26 | N/A | 20 | 1 | 1,371 | 99.1% (105/106) (96.7-100) | 98.6% (1,371/1,391) (97.9-99.2) |
| Female Urine | Symptomatic | 529 | 30 | 14 | 5 | 6 | 6 | 468 | 89.1% (49/55) (79.9-98.2) | 98.7% (468/474) (97.6-99.8) |
| | Asymptomatic | 844 | 32 | 6 | 1 | 7 | 9 | 789 | 81.3% (39/48) (69.2-93.3) | 99.1% (789/796) (98.4-99.8) |
| | TOTAL | 1,373 | 62 | 20 | 6 | 13 | 15 | 1,257 | 85.4% (88/103) (78.1-92.7) | 99.0% (1,257/1,270) (98.4-99.6) |
| Male Urethral | Symptomatic | 237 | 30 | 14 | N/A | 12 | 1 | 180 | 97.8% (44/45) (92.4-100) | 93.8% (180/192) (90.1-97.4) |
| | Asymptomatic | 163 | 7 | 4 | N/A | 5 | 0 | 147 | 100% (11/11) (95.5-100) | 96.7% (147/152) (93.5-99.9) |
| | TOTAL | 400 | 37 | 18 | N/A | 17 | 1 | 327 | 98.2% (55/56) (93.9-100) | 95.1% (327/344) (92.6-97.5) |
| Male Urine | Symptomatic | 516 | 52 | 24 | 6 | 17 | 5 | 412 | 94.3% (82/87) (88.8-99.7) | 96.0% (412/429) (94.1-98.0) |
| | Asymptomatic | 250 | 12 | 4 | 3 | 5 | 2 | 224 | 90.5% (19/21) (75.5-100) | 97.8% (224/229) (95.7-99.9) |
| | TOTAL | 766 | 64 | 28 | 9 | 22 | 7 | 636 | 93.5% (101/108) (88.4-98.6) | 96.7% (636/658) (95.2-98.1) |
| Total Specimens | | 4,036 | 242 | 92 | 15 | 72 | 24 | 3,591 | 93.6% (349/373) | 98.0% (3,591/3,663) |
{9}
Hypothetical positive and negative predicative values (PPV and NPV) for different prevalence rates were calculated using the AMP CT overall clinical sensitivity and specificity of 93.6% and 98.0%, respectively.
AMP CT Assay Hypothetical Predictive Values at Different Prevalence Rates
| Prevalence Rate (%) | Sensitivity (%) | Specificity (%) | Positive Predictive Value (%) | Negative Predictive Value (%) |
| --- | --- | --- | --- | --- |
| 5 | 93.6 | 98.0 | 71.5 | 99.7 |
| 10 | 93.6 | 98.0 | 84.1 | 99.3 |
| 15 | 93.6 | 98.0 | 89.4 | 98.9 |
| 20 | 93.6 | 98.0 | 92.2 | 98.4 |
## Conclusions from Non-Clinical and Clinical Data
The nonclinical and clinical performance data generated for the GEN-PROBE AMPLIFIED Chlamydia Trachomatis Assay supports the statement of SUBSTANTIAL EQUIVALENCE of the GEN-PROBE AMPLIFIED Chlamydia Trachomatis Assay Kit to tissue culture media and components for accurately detecting C. trachomatis in clinical specimens.
The results of this clinical study demonstrate reasonable evidence that when the GEN-PROBE AMPLIFIED Chlamydia Trachomatis Assay kits are labeled as proposed, the GEN-PROBE AMPLIFIED Chlamydia Trachomatis Assay is safe and effective for its stated Intended Use.
The GEN-PROBE AMPLIFIED Chlamydia Trachomatis Assay, as a diagnostic tool, provides information that measurably contributes to a diagnosis of C. trachomatis infection by being substantially equivalent to tissue culture media in terms of Safety and Effectiveness.
## Contraindications and Cautions
There are no contraindications or cautions.
## References
Arnold, L.J., P.W. Hammond, W.A. Wiese, and N.C. Nelson. 1989. Assay Formats Involving Acridinium-Ester-Labeled DNA Probes. Clin. Chem. 35:1588-1594.
Centers for Disease Control and Prevention. 1993. Recommendations for the Prevention and Management of Chlamydia trachomatis infections. Morbid and Mortal. Weekly Rep. 42 (RR-12):1-39.
000051
{10}
Nelson, N.C. 1992. Detection of Acridinium Esters by Chemiluminescence. In L.J. Kricka (ed.), Nonisotopic DNA Probe Techniques. Academic Press, San Diego, CA.
Schachter, J., and M. Grossman. 1981. Chlamydial infections. Ann. Rev. Med. 32:45-61.
Schachter, J. 1978. Medical progress: Chlamydial infections (first of three parts). NEJM 298:428-435.
Wolcott, M.J. 1992. Advances in Nucleic Acid-Based Detection Methods. Clin. Micro. Rev. 5:370-386.
000052
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.