← Product Code [DBI](/productcode/DBI) · K962198

# QUANTEX C4 (K962198)

_Instrumentation Laboratory CO · DBI · Sep 26, 1996 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K962198

## Device Facts

- **Applicant:** Instrumentation Laboratory CO
- **Product Code:** [DBI](/productcode/DBI.md)
- **Decision Date:** Sep 26, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5240
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

This product permits the quantitative in vitro diagnostic determination of Complement 4 in serum and plasma on the ILab Clinical Chemistry System by turbidimetric immunoassay method. C4 is one of a group of serum proteins which destroy infectious agents. Measurement of this protein aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of C4.

## Device Story

Turbidimetric immunoassay for quantitative C4 determination in serum/plasma; performed on ILab Clinical Chemistry System. Input: patient serum/plasma samples. Principle: antigen-antibody reaction causing turbidity proportional to C4 concentration. Output: quantitative C4 levels. Used in clinical laboratory settings by trained laboratory personnel. Results assist clinicians in diagnosing immunologic disorders and C4 deficiencies.

## Clinical Evidence

Method comparison study; 38 serum samples; compared quantex C4 on ILab system vs. IL Test™ C4 on Monarch system. Correlation coefficient (r) = 0.9791.

## Technological Characteristics

Turbidimetric immunoassay; reagents include anti-human C4 and buffer. Designed for use on ILab Clinical Chemistry System. Quantitative measurement of serum proteins.

## Regulatory Identification

A complement components immunological test system is a device that consists of the reagents used to measure by immunochemical techniques complement components C1q , C1r , C1s , C2 , C3 , C4 , C5 , C6 , C7 , C8 , and C9 , in serum, other body fluids, and tissues. Complement is a group of serum proteins which destroy infectious agents. Measurements of these proteins aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components.

## Predicate Devices

- IL Test™ C4 ([K920010](/device/K920010.md)/B)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# SECTION 3

## quantex C4 - 510(k) SUMMARY

### (Summary of Safety and Effectiveness)

**Submitted by:**

Betty Lane
Director, Regulatory Affairs
Instrumentation Laboratory Company
113 Hartwell Avenue
Lexington, MA 02173
phone (617) 861-4182
fax (617) 861-4464

*K962198*

**SEP 26 1996**

**Contact Person:**

Betty Lane
phone (617) 861-4182

**Summary Prepared:**

May 29, 1996

**Name of the device:**

quantex C4:
8 x 6 mL anti-human C4 P/N 3000-22137
2 x 100 mL Buffer P/N 3000-22130

**Classification name(s):**

866.5240 Complement components immunological test systems Class II

**Identification of predicate device(s):**

IL Test™ C4 #K920010/B

**Description of the device/intended use(s):**

This product permits the quantitative *in vitro* diagnostic determination of Complement 4 in serum and plasma on the ILab Clinical Chemistry System by turbidimetric immunoassay method. C4 is one of a group of serum proteins which destroy infectious agents. Measurement of this protein aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of C4.

**Statement of how the Technological Characteristics of the Device compare to the Predicate device:**

Biokit’s quantex C4 is substantially equivalent in performance, intended use, safety and effectiveness to the predicate, IL Test™ C4 -- 510(k) #K920010/B.

**Summary of Performance Data:**

A comparative performance study was performed with IL Test™ C4 on a Monarch Clinical Chemistry System and Biokit’s quantex C4 on an ILab Clinical Chemistry System. Method comparison studies evaluated C4 levels of thirty-eight serum samples on the respective instruments. The correlation $(r)$ in these studies was 0.9791.

Section 3
quantex C4 510(k)
Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K962198](https://fda-staging.innolitics.com/device/K962198)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
