K962194 · Medicinelodge, Inc. · MBI · Jul 30, 1996 · Orthopedic
Device Facts
Record ID
K962194
Device Name
MLI SOFT TISSUE SCREW & WASHER SYSTEM
Applicant
Medicinelodge, Inc.
Product Code
MBI · Orthopedic
Decision Date
Jul 30, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MedicineLodge, Inc. soft tissue anchoring bone screws and washers are indicated for use in soft tissue graft fixations to the femur or tibia, or both, during cruciate ligament, tendon or other soft tissue graft (e.g. semitendinosus) repair procedures.
Device Story
System consists of orthopedic bone screws and washers used for soft tissue graft fixation to bone; screws are self-tapping, manually inserted using 3.5mm hex driver; washers feature distal spikes to anchor graft; used by surgeons in orthopedic procedures; device provides mechanical fixation of graft to femur or tibia; benefits include secure graft attachment during ligament/tendon repair.
Clinical Evidence
Bench testing only. Static tensile testing performed on four screw/washer configurations using a bass wood model to simulate clinical application. Primary endpoint was yield load; results showed MLI constructs performed as well or better than predicate systems. Failure modes were consistent with established literature (Matthews et.al, Kurosaka et.al).
Technological Characteristics
Materials: Ti6Al4V ELI (ASTM F-316). Screw design: ISO 5835 compliant, self-tapping, 4.5mm cortical and 6.5mm cancellous diameters. Washer configurations: round, triangular, and oval with distal spikes. Insertion: manual 3.5mm hex driver. Sterilization: sterile and non-sterile options.
Indications for Use
Indicated for soft tissue graft fixation to femur or tibia during cruciate ligament, tendon, or other soft tissue repair procedures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Linvatec Concept Soft Tissue Anchoring Implants
Arthrotek Ligament Screw System
Mitek Screws and Washers
Reference Devices
Matthews et.al (literature)
Kurosaka et.al (literature)
Submission Summary (Full Text)
{0}
K962194
JUL 30 1996
# Premarket Notification
510(k) Summary of Safety and Effectiveness Information
## DATE OF SUMMARY PREPARATION
June 5, 1996
## MANUFACTURER
MedicineLodge, Incorporated
152 South 600 West
Logan, Utah 84321
Telephone: (801) 753-7675
Fax: (801) 753-7698
## MANUFACTURER CONTACT PERSON
Wade Fallin
Vice President of R&D
MedicineLodge, Inc.
152 South 600 West
Logan, Utah 84321
Telephone: (801) 753-7675
Fax: (801) 753-7698
## PROPRIETARY NAME
MedicineLodge, Inc. (MLI) Soft Tissue Screw and Washer System
## COMMON NAMES
Orthopedic Bone Screw
Orthopedic Washer
## CLASSIFICATION NAME
Non-degradable Soft Tissue Fastener
## CLASSIFICATION REFERENCE
Currently, no FDA classification exists for Non-degradable Soft Tissue Fasteners
## DEVICE PRODUCT CODE
MBI
34
{1}
35
# CLASSIFICATION PANEL
Orthopedic and Rehabilitation Devices
# PROPOSED REGULATORY CLASS
In accordance with FDA classification of fixation bone screws and orthopedic washers as Class II medical devices, this device system is proposed for placement in Class II.
# REASON FOR PREMARKET NOTIFICATION
The MLI Soft Tissue Screw and Washer System is a new medical device proposed for placement in Class II.
# SPECIAL CONTROLS
At this time, Food and Drug Administration generated Performance Standards applicable to MedicineLodge, Inc. Soft Tissue Screws and Washers are not in force. However, MedicineLodge, Inc. has conducted mechanical testing on four screw/washer configurations to demonstrate the substantial equivalence of the device. Additionally, MedicineLodge, Inc. utilizes materials and vendor certifications, in-house standard operating procedures (SOP's), and ASTM standards, as appropriate.
# INDICATIONS FOR USE
MedicineLodge, Inc. soft tissue anchoring bone screws and washers are indicated for use in soft tissue graft fixations to the femur or tibia, or both, during cruciate ligament, tendon or other soft tissue graft (e.g. semitendinosus) repair procedures.
# SUBSTANTIALLY EQUIVALENT PREDICATE DEVICES
MedicineLodge, Inc. believes that several commercially available devices are substantially equivalent to the MLI Soft Tissue Screw and Washer System. Within the proposed class, the following devices are used as predicate devices for comparison:
1) Linvatec Concept Soft Tissue Anchoring Implants
Linvatec, Inc.; Largo, FL
2) Arthrotek Ligament Screw System
Arthrotek, Inc., Ontario, CA
3) Mitek Screws and Washers
Mitek Surgical Products, Inc.; Norwood, MA
MedicineLodge, Inc. believes that the information presented below demonstrates the substantial equivalence of the MLI Soft Tissue Screw and Washer System when compared to the above selected predicate devices.
{2}
36
# SUBSTANTIAL EQUIVALENCE COMPARISON
To begin, the materials, method of manufacture, method of sterilization, and device packaging for all MLI Soft Tissue Screw and Washer System components are equivalent to those used for the selected predicate devices.
MLI Soft Tissue Screws and Washers are indicated for use in soft tissue graft fixations to the femur or tibia, or both, during cruciate ligament, tendon or other soft tissue graft (e.g. semitendinosus) repair procedures. The previously listed predicate device systems are similarly indicated for procedures involving soft tissue reconstruction.
The MLI Soft Tissue Screws and Washers are designed to be used as an assembly unit. The MLI Soft Tissue Anchoring Screws are provided in standard sizes, as tabulated below:
| Screw Diameter | Screw Length |
| --- | --- |
| 4.5 mm (Cortical) | 30 mm - 60 mm, in 2 mm increments |
| 6.5 mm (Cancellous) | 20 mm - 55 mm, in 5 mm increments |
MLI Screws are equivalent in design to currently marketed standard bone screws, with all threads and defining screw features conforming to ISO 5835, Metal Bone Screws with Hexagonal Drive Connection, Spherical Under-Surface of Head, Assymmetrical Thread Dimensions. The screws are self-tapping and are designed to be manually inserted using a standard 3.5 mm hex driver. Engineering Drawings are provided in Exhibit I for reference.
The MLI Washers are designed to be universally compatible with both the 4.5 mm and 6.5 mm MLI Screws and are provided in the three following design configurations to allow the surgeon a variety of washers to choose from:
| Washer Shape | Washer Diameter |
| --- | --- |
| Round | 14 mm, 17 mm, 20 mm |
| Triangular | 17.5 mm, 20.5 mm |
| Oval | 19 mm, 22 mm |
The washers possess spikes spaced equidistant from each other on the distal side of the washer to penetrate the host bone and thus anchor the graft in place. Engineering drawings of the washers are also provided in Exhibit I for reference.
Although the washer design configurations listed above may vary slightly from standard washers used in soft tissue repair, testing performed by MedicineLodge, Inc. shows that these differences do not compromise the performance of the washers relative to standard round washers in their intended clinical use. This testing is discussed below.
## DEVICE TESTING
MedicineLodge, Inc. has conducted device testing on a variety of screw/washer configurations to assess the performance of the MLI Soft Tissue Screw and Washer
{3}
System. These results were then used to compare the performance of the MLI Screws and Washers to a commercially marketed soft tissue screw and washer system.
The following tests were conducted using the indicated screw/washer combinations:
(i) Test #1: Fixation strength of the MLI 4.5mm X 52mm Cortical Screw and MLI-Small Round Washer (13 mm) in Static Bone Patellar Tendon Tensile Testing in a Bass Wood Model.
(ii) Test #2: Fixation strength of the MLI 4.5mm X 52mm Cortical Screw and MLI-Large Round Washer (18 mm) in Static Bone Patellar Tendon Tensile Testing in a Bass Wood Model.
(iii) Test #3: Fixation strength of the MLI 4.5mm X 52mm Cortical Screw and MLI-Triangular Washer (11.5 mm) in Static Bone Patellar Tendon Tensile Testing in a Bass Wood Model.
(iv) Test #4: Fixation strength of the MLI 4.5mm X 52mm Cortical Screw and MLI-Football Washer (11.5 mm x 8.5 mm) in Static Bone Patellar Tendon Tensile Testing in a Bass Wood Model.
The mode of failure for all MLI Screw and Washer configurations was similar for all screw/washer constructs tested, wherein the ligament pulled out from under the washer. The modes of failure are also consistent with results obtained by Matthews et.al and Kurosaka et.al where the site of failure was at the point of ligament anchorage to the implant. It is important to note that yield of the graft construct results in permanent deformation of the ligament, which can in turn result in a loose or lax graft. For this reason, MedicineLodge, Inc. believes that the yield load data best measures the observed mode of failure.
When the yield load results for the various MLI Soft Tissue Screw and Washer System screw/washer constructs were compared to the currently marketed screw/washer construct, the MLI screw/washer constructs all performed as well or better than the predicate screw/washer construct.
It is important to state that the washer sizes that will comprise the MLI Soft Tissue Screw and Washer System will be larger than those tested. The thus increases the distal surface area of the washers, which in turn should result in increased fixation strength of the graft construct in true clinical application of the product. In particular, for the Football washer (tested size: 11.5mm; marketed sizes: 19mm and 22mm), this size increase should provide additional assurance that the device will perform equivalent to the predicate screw/washer construct.
It is our belief that these results confirm the equivalence of the MLI Soft Tissue Screw and Washer System relative to the tested predicate device in testing devised to simulate clinical application of the product.
## CONCLUSION
Based on the design concept, indications for use, use of standard materials, and feature comparisons to the selected predicate devices, MedicineLodge, Inc. believes that sufficient evidence exists to conclude that MLI Soft Tissue Screw and Washer System is
{4}
substantially equivalent to existing legally marketed soft tissue anchor screw and washer systems.
Table 1 on the following page summarizes the relevant feature comparisons between the MLI Soft Tissue Screw and Washer System and the selected predicate devices.
38
{5}
39
TABLE 1
Feature comparisons between the MedicineLodge, Inc. Soft Tissue Anchoring Screw and Washer System and a selected predicate device system (differences are indicated in **bold** print).
| | MLI Soft Tissue Screw and Washer System | Mitek Screws and Washers | Linvatec Concept Soft Tissue Anchoring Set | Arthrotek Ligament Screw and Washer System |
| --- | --- | --- | --- | --- |
| Indications | MedicineLodge, Inc. soft tissue screws and washers are indicated for use in soft tissue graft fixations to the femur or tibia, or both, during cruciate ligament, tendon or other soft tissue graft (e.g. semitendinosus) repair procedures. | Mitek screws and washers are used to attach ligaments, tendons and other soft tissue to bone. | Concept soft tissue screws and washers are used to attach ligaments, tendons and other soft tissue to bone. | Arthrotek Ligament Screw and Washer system is designed to produce an efficient surgical procedure for iliotibial tenodesis in anterior cruciate reconstruction. |
| Materials | Ti6Al4V ELI (per ASTM F-316) | Ti6Al4V ELI (per ASTM F-316) | Ti6Al4V ELI (per ASTM F-316) | Ti6Al4V ELI (per ASTM F-316) |
| Cortical Screws | Length: 30 mm to 60 mm, in 2 mm increments
Outer Diameter: 4.5 mm | Length: 26 mm to 76 mm, in 2 mm increments
Outer Diameter: 4.5 mm | Length: 30 mm to 45 mm, in 5 mm increments; 50 mm to 70 mm with smooth proximal shank, in 10 mm increments.
Outer Diameter: 3.5 mm | Length: 30 mm to 80 mm, in 2 mm increments
Outer Diameter: 4.5 mm |
| Cancellous Screws | Length: 20 mm to 55 mm, in 5 mm increments
Outer Diameter: 6.5 mm | Length: 20 mm to 50 mm, in 5 mm increments
Outer Diameter: 6.5 mm | Not Listed | Length: 25 mm to 55 mm, in 5 mm increments
Outer Diameter: 6.5 mm |
| ISO Conformance of Screw Design | Yes | Yes | Yes | Yes |
| Screw Head Profile | Low profile | Low Profile | Low Profile | Low Profile |
| Surface Finish | Satin Texture | Satin Texture | Satin Texture | Satin Texture |
| Tapped/Self-tapping | Self-tapping | Self-tapping | Self-tapping | Tapped hole |
| Method of Insertion | 3.5mm Hex Driver | 2.5mm Hex Driver | 3.5mm Hex Driver | 2.5mm Hex Driver |
| Washers | Round: 14mm, 17mm, 20mm
Triangular (Delta): 17.5mm, 20.5mm
Oval (Football): 19mm, 22mm | Round: 14mm, 17mm, 20mm | Round (with post): 14mm, 17mm, 20mm
Round (Flat Washers): 14mm, 17mm | Round: 14mm, 18mm |
| Suture Holes in Washer | No | No | Yes | Yes |
| Imbrications on Distal Surface of Washer | Yes | Yes | No | No |
| Sterilization Status | Sterile & Non-sterile | Sterile & Non-sterile | Sterile & Non-sterile | Sterile & Non-sterile |
| Manufacturer | MedicineLodge, Inc. | Mitek Surgical Products, Inc. | Linvatec, Inc. | Arthrotek, Inc. |
| Currently Marketed? (Presumed 510(k) Approved) | N/A | Yes | Yes | Yes |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.