← Product Code [GTZ](/productcode/GTZ) · K962146

# EXCEL ONESTEP GROUP A STREP ANTIGEN MODULE TEST KIT (K962146)

_Excel Scientific, Inc. · GTZ · Oct 7, 1996 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K962146

## Device Facts

- **Applicant:** Excel Scientific, Inc.
- **Product Code:** [GTZ](/productcode/GTZ.md)
- **Decision Date:** Oct 7, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3740
- **Device Class:** Class 1
- **Review Panel:** Microbiology

## Indications for Use

Excel OneStep Group A Strep Antigen Module Test is Excel Scientific, Inc.'s name for the professional use of solid phase migratory chromatography immunoassay in vitro diagnostic test kit for the qualitative detection of group A strep antigen from throat swabs.

## Device Story

In vitro diagnostic test kit; utilizes solid phase migratory chromatography immunoassay. Input: throat swab specimen. Process: chemical extraction of carbohydrate antigen using Reagent A and B; capillary migration through porous membrane; antibody-antigen-antibody/colloidal gold double antibody sandwich formation. Output: visual color lines in read window (two lines = positive; one control line = negative). Used in clinical settings by healthcare professionals. Benefit: rapid (6-7 minutes) qualitative detection of group A strep antigen to aid in diagnosis.

## Clinical Evidence

Correlation study performed comparing Excel OneStep Group A Strep Antigen assay against standard culture method (sheep blood trypticase soy agar plate) followed by latex agglutination confirmation.

## Technological Characteristics

Solid phase migratory chromatography immunoassay; colloidal gold signaling molecules; polyclonal antibodies; capillary action; manual test module; includes drying agent in sealed foil pouch.

## Regulatory Identification

Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identify Streptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Streptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.

## Reference Devices

- Diagnostic Product Corporation (DPC) latex agglutination strep grouping kit

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

EXCEL "Commitment to Excellence"

K960146

OCT - 7 1996

# Summary of Safety and Effectiveness

## Excel OneStep Group A Strep Antigen Module Test

Excel OneStep Group A Strep Antigen Module Test is Excel Scientific, Inc.'s name for the professional use of solid phase migratory chromatography immunoassay *in vitro* diagnostic test kit for the qualitative detection of group A strep antigen from throat swabs. During correlation studies, prior to using the Excel OneStep Group A Strep Antigen assay, each throat swab was used to inoculate a sheep blood (trypticase) soy agar plate for culture. Group A Streptococci were confirmed by a latex agglutination assay (a commercially available strep grouping kit from Diagnostic Product Corporation, Los Angeles, CA). The exact difference between the two methodologies are as follows:

1. The Excel OneStep Group A Strep Antigen Module Test utilizes the chemical extraction of a carbohydrate antigen from group A streptococcus from the throat swabs followed by the utilization of migratory color immunoassay technology for the qualitative detection of group A streptococcus. The test module has two polyclonal antibodies, one immobilized on the porous membrane, while the other antibody on an absorbent near the sample application is conjugated with colloidal gold particles as signaling molecules. A swab specimen taken from a patient's throat is treated with a fresh mixture of Reagent A and Reagent B to extract out the group A streptococcus antigen. The extract is then transferred to the reaction module with the aid of a transfer pipette and allowed to migrate throughout the absorbent and porous membrane in the reaction module by capillary action until it reaches the opposite end of the read window (in approximately 6~7 minutes). The antibody-antigen-antibody/colloidal gold double antibody sandwich assay is formed if strep A antigen is present in the throat swab. If two colored lines develop in the read window, one in the test zone and the other in the control zone, the presence of strep A antigen is indicative. The absence of this antigen will result in only a single color line located in the control zone in the read window. The appearance of a single colored line for negative specimens also gives an added measure of quality control by demonstrating antibody recognition, assuring that the procedure was performed correctly and that the reagents are chemically active. A drying agent is enclosed with the reaction module in a sealed foil pouch to ensure the stability of the reactive reagents.

2. In the culturing procedure, isolation of streptococcal group A organisms includes the use of culture plates followed by a confirmation assay (such as DPC's latex agglutination strep grouping kit) for strep A. A sheep blood (trypticase) soy agar plate (containing 5% or 10% sheep blood) is used for the cultivation and isolation of microorganisms from the throat swabs and for visualizing hemolytic reactions.

---

**Source:** [https://fda-staging.innolitics.com/device/K962146](https://fda-staging.innolitics.com/device/K962146)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
