SONOLINE VERSA 555 DIAGNOSTIC ULTRASOUND SYSTEM

K962142 · Siemens Medical Solutions USA, Inc. · IYO · May 29, 1997 · Radiology

Device Facts

Record IDK962142
Device NameSONOLINE VERSA 555 DIAGNOSTIC ULTRASOUND SYSTEM
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeIYO · Radiology
Decision DateMay 29, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1560
Device ClassClass 2
AttributesPediatric

Indications for Use

The SONOLINE Versa 555 ultrasound imaging system is intended for the following applications: General Radiology, Abdominal, Intraoperative (abdominal and neurosurgical), Small Parts, Transcranial, OB/GYN, Pelvic, Neonatal Cephalic, Urology, Vascular, Peripheral Vascular, and Transesophageal Echocardiography (TEE) applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

Device Story

Mobile, software-controlled diagnostic ultrasound system; acquires ultrasound data via transducers; displays B-Mode, M-Mode, Color Flow, Pulsed (PW) Doppler, and Continuous (CW) Doppler on CRT. Operated by clinicians in clinical settings. System processes echo pulses to generate visual images and anatomical measurements; provides analysis packages for clinical diagnosis. Includes on-screen display of thermal (IT) and mechanical (MI) indices for bioeffect monitoring. Supports various phased, linear, and curved array transducers. Output assists healthcare providers in diagnosing disease states and monitoring normal conditions like pregnancy.

Clinical Evidence

No clinical data. Substantial equivalence is based on technological comparison, adherence to safety standards (UL 2601, CSA 22.2, IEC 601-1), and identical operating principles to the predicate device.

Technological Characteristics

Mobile ultrasound system; software-controlled electronics; CRT display. Supports phased, linear, and curved array transducers. Complies with UL 2601, CSA 22.2 No. 601-1, and IEC 601-1-1/2 safety standards. Displays real-time thermal (IT) and mechanical (MI) indices per AIUM/NEMA 1992. Connectivity via transducer ports.

Indications for Use

Indicated for diagnostic ultrasound imaging and Doppler analysis of the human body, including cephalic (adult/neonatal) and cardiac (adult/pediatric) applications. Used for general radiology, abdominal, intraoperative, small parts, transcranial, OB/GYN, pelvic, urology, vascular, peripheral vascular, and TEE applications.

Regulatory Classification

Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0} MAY 29 1997 K962142 # 510(K) SUMMARY ## SONOLINE Versa 555 Diagnostic Ultrasound System This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with the Safe Medical Device Act of 1990 revisions to 21 CFR, Part 807.92, Content and Format of a 510(k) Summary. 1. **Submitted By:** Siemens Medical Systems, Inc., Ultrasound Group 22010 SE 51st Street Issaquah, WA 98027-7002 **Contact Person:** Steve Hesler Manager of Regulatory Affairs (206) 557-1629 **Date Prepared:** May 30, 1996 2. **Proprietary Name:** SONOLINE Versa 555 Diagnostic Ultrasound System **Common/ Usual Name:** Diagnostic Ultrasound System with Accessories **Classification Name:** Ultrasonic Pulsed Doppler Imaging System (Product Code 90 YIN, 21 CFR 892.1550) 3. **Predicate Device:** Siemens SONOLINE Versa Diagnostic Ultrasound System (K945773), 8/7/95. Siemens SONOLINE Elegra Diagnostic Ultrasound System (K945072), 11/21/95. 4. **Device Description:** The SONOLINE Versa 555 is a general purpose, mobile, software-controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical values related to potential bioeffect mechanisms. Its function is to acquire ultrasound data and display it in B-Mode, M-Mode, Color Flow Modes, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, or in a combination of modes, on a CRT display. Page 1 16 {1} Siemens Medical Systems, Inc. Ultrasound Group SONOLINE Versa 555 Ultrasound System Modification to Existing Cleared System The SONOLINE ® Versa 555 has been designed to meet the following product safety standards: - UL 2601, Safety Requirements for Medical Equipment - CSA 22.2 No. 601-1, Safety Requirements for Medical Equipment - Standard for Real Time Display of Thermal and Mechanical Indices on Diagnostic Ultrasound Equipment, AIUM/NEMA, 1992. - 93/94/EEC Medical Devices Directive - EN60601 = (IEC 601-1-1 + IEC 601-1-2), Safety and EMC, Requirements for Medical Equipment ## 5. Intended Uses: The SONOLINE Versa 555 ultrasound imaging system is intended for the following applications: General Radiology, Abdominal, Intraoperative (abdominal and neurosurgical), Small Parts, Transcranial, OB/GYN, Pelvic, Neonatal Cephalic, Urology, Vascular, Peripheral Vascular, and Transesophageal Echocardiography (TEE) applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes. ## 6. Technological Comparison to Predicate Device: The SONOLINE Versa 555 is similar to the SONOLINE Versa in that both incorporate software controlled electronics to transmit ultrasonic pulses, via a transducer, into a patient, then receive return (echo) pulses and to convert those pulses into a visual display, to be provided information for use in the diagnosis of various disease states as well as normal conditions such as pregnancy and physical development. The operating principles of both systems are identical. However, the SONOLINE Versa 555 supports phased array transducers. Both systems incorporate an on-screen display of Mechanical (MI) and Thermal (IT) Indices, in compliance with the Standard for Real Time Display of Thermal and Mechanical Indices on Diagnostic Ultrasound Equipment, AIUM/NEMA, 1992. End of 510(k) Summary Page 2 165 {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 29 1997 Steve Hesler Manager Regulatory Affairs Siemens Medical Systems 22010 S.E. 51st St. Issaquah, WA 98027-7002 Re: K962142 Sonoline Versa 555 Dated: April 10, 1997 Received: April 11, 1997 Regulatory class: II 21 CFR 892.1550/Procode: 90 IYO Dear Mr. Hesler: We have reviewed your section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act. You may, therefore, market the device, subject to the general controls provisions Act (Act). The general controls provisions of the Act include requirements for registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the Sonoline Versa 555, as described in your premarket notification: | Transducer Model Number | | --- | | 2.5 MHZ 2.5P20 Phased Linear Array | | 3.5 MHZ 3.5P14 Phased Linear Array | | 3.5 MHZ 3.5C20 Curved Array | | 3.5 MHZ 3.5C40 Curved Array | | 3.5 MHZ 3.5C60 Curved Array | | 3.5 MHZ 3.5C80 Curved Array | | 5.0 MHZ TEE | | 5.0 MHZ 5.0P10 Phased Linear Array | | 5.0 MHZ 5.0L45 Linear Array | | 5.0 MHZ 5.0C50 Curved Array | {3} Page 2 - Mr. Steve Hesler 5.0 MHZ 5.0L90 Linear Array 6.5 MHZ 6.5F13 Curved Array 6.5 MHZ 6.5EV13 Curved Array 7.5 MHZ 7.5C30 Curved Array 7.5 MHZ 7.5L40 Linear Array 7.5 MHZ 7.5L501 Linear Array 7.5 MHZ 7.5L50Q Linear Array 7.5 MHZ 7.5L70 Linear Array 10.0 MHZ 10.0L25 Linear Array If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the *Code of Federal Regulations*, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the *Federal Register*. *Please note*: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for *in vitro* fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for *in vitro* fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the provision of the investigational device exemptions (IDE) regulations, 21 CFR, Part 812. Therefore, your product labeling must be consistent with FDA’s position on this use. This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center’s February 17, 1993 “Revised 510(k) Diagnostic Ultrasound Guidance for 1993.” If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded. The special report should reference the manufacturer’s 510(k) number. It should be clearly and prominently marked “ADD-TO-FILE” and should be submitted in duplicate to: Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850 {4} Page 3 - Mr. Steve Hesler This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, “Misbranding by reference to premarket notification” (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address “dsmo@fdadr.cdrh.fda.gov”. If you have any questions regarding the content of this letter, please contact Maureen Butler at (301) 594-1212. Sincerely yours, ![img-0.jpeg](img-0.jpeg) Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure(s) {5} # Ultrasound Device Indications Statement 510 (k) Number (if known) K962142 Device Name 2.5P20 transducer for use with the SONOLINE Versa 555 Indications For Use Diagnostic ultrasound imaging and Doppler analysis of the human body as follows: Mode of Operation | Clinical Application | B | M | PWD | CWD | C (Color Maps) | Combined (B,C,D) | Other (Specify) | | --- | --- | --- | --- | --- | --- | --- | --- | | Ophthalmic | | | | | | | | | Fetal | | | | | | | | | Abdominal | | | | | | | | | Intraoperative Abdominal Neurosurgical | | | | | | | | | Pediatric | | | | | | | | | Small Organ | | | | | | | | | Cephalic Adult Neonatal | ☑ | ☑ | ☑ | ☑ | ☑ | ☑ | | | Cardiac Adult Pediatric | ☑ | ☑ | ☑ | ☑ | ☑ | ☑ | | | Trans-esophageal | | | | | | | | | Trans-rectal | | | | | | | | | Trans-vaginal | | | | | | | | | Intra-luminal | | | | | | | | | Trans-urethral | | | | | | | | | Peripheral vessel | | | | | | | | | Laparoscopic | | | | | | | | (Other Indications or Modes) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use: Per 21 CFR 801.109) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 5.0(k) Number K962142 page 2 of 20
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...