← Product Code [GEX](/productcode/GEX) · K961935

# SHARPLAN LASERS, INC. SILKTOUCH CO2 FLASH SCANNER (K961935)

_Sharplan Lasers, Inc. · GEX · Nov 1, 1996 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K961935

## Device Facts

- **Applicant:** Sharplan Lasers, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Nov 1, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

It is indicated for soft tissue incision of recipient sites for laser assisted hair transplantation.

## Device Story

Motorized laser beam deflection system; attaches to CO2 laser articulating arm. Redirects focused laser beam via moving mirrors within specified pattern scan area. Limits tissue exposure time while maintaining power density for vaporization at low average power. Includes controls for scan area and exposure time. Used with CO2 handpiece. When off, laser functions normally for manual incision/vaporization. Operated by surgeons in clinical settings. Benefits: precise tissue vaporization for hair transplant recipient sites.

## Clinical Evidence

Histological information submitted to support safety and efficacy for soft tissue incision.

## Technological Characteristics

Motorized laser beam deflection system; moving mirror assembly; attaches to CO2 laser articulating arm; manual controls for scan area and exposure time.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Norcross Scientific SurgiScan CO2 Scanner
- A-Z Surgical Super Punch
- Robbins Acrotorque Hand Engine

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

NOV 1 1996
K 96 1935

# 510(k) Summary of Safety and Effectiveness
## Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner Accessory

In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.

The safety and effectiveness of the Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner is based on a determination of the substantial equivalence as well as the safety and effectiveness of its predicate devices which include the Norcross Scientific SurgiScan CO2 Scanner, the A-Z Surgical Super Punch, and the Robbins Hand Engine.

1. Sharplan Lasers, Inc.
1 Pearl Court
Allendale, NJ 07401
George J. Hattub, Director of Regulatory Affairs
May 6, 1996
1-800-394-2000, ext. 531

2. Model: Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner Accessory

3. Predicate Devices: Norcross Scientific SurgiScan CO2 Scanner, the A-Z Surgical Super Punch, the Robbins Acrotorque Hand Engine.

4. Description: The Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner Accessory is a motorized, laser beam deflection system that is attached to a CO2 laser articulating arm. The device redirects a focused laser beam through moving mirrors, within a specified pattern scan area. Similar to the common surgeon’s technique of manually scanning a focused beam over a treatment area, the constantly moving beam of the SilkTouch Flash Scanner limits the time of laser exposure on tissue while providing sufficient power density for tissue vaporization at low average power levels. The SilkTouch Flash Scanner has controls for varying the scanned area setting the exposure time. This device is used with a CO2 handpiece. When the unit is turned off, the connected laser operates to its specifications and allows incision or manual scanning for vaporization purposes.

5. The Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner Accessory is substantially equivalent to the following devices: the Norcross Scientific SurgiScan CO2 Scanner, the A-Z Surgical Super Punch, and the Robbins Acrotorque Hand Engine. It is indicated for soft tissue incision of recipient sites for laser assisted hair transplantation.

{1}

# 510(k) Summary of Safety and Effectiveness (continued)
Sharplan Lasers, Inc. SilkTouch CO2 Flash Scanner Accessory

6. From a clinical performance perspective, the aforementioned devices while are not identical, perform the same function for the same indication. Regarding the technological similarities; while two of the predicate devices are of the same photonic technology with minor differences, the non-laser devices relies upon a mechanical technology to achieve its clinical effect. Based upon the histological information submitted, Sharplan Lasers, Inc. believes that the expanded indications for this device raises no safety or efficacy concerns.

**Advisory:** This information was prepared for the sole purpose of compliance with the Safe Medical Devices Act of 1990. It does not imply that the procedures described herein can be performed with the equipment described without substantial risk of personal injury or death to patients due to operator error or in procedures requiring a high degree of skill.

---

**Source:** [https://fda-staging.innolitics.com/device/K961935](https://fda-staging.innolitics.com/device/K961935)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).