PISTON SYRINGE

K961877 · Univec, Inc. · FMI · Aug 7, 1996 · General Hospital

Device Facts

Record IDK961877
Device NamePISTON SYRINGE
ApplicantUnivec, Inc.
Product CodeFMI · General Hospital
Decision DateAug 7, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The UNIVEC Nonreusable Locking Syringe with Coupling Assembly can be used by health care professionals, diabetics and other persons for injection of insulin, tuberculin, immunization serum, and other injectable medications.

Device Story

Single-use piston syringe with integrated coupling assembly; assembly inserted between plunger and rubber stopper; mechanism decouples plunger from stopper after single dosage injection; disables syringe to prevent reuse; operated by healthcare professionals, diabetics, or patients; manual administration of injectable medications; prevents needle stick injuries and reuse; provides safe delivery of insulin, tuberculin, and immunization serum.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Piston syringe and hypodermic needle; polypropylene construction; manual energy source; single-use locking mechanism via coupling assembly; non-aspirating design.

Indications for Use

Indicated for health care professionals, diabetics, and other individuals requiring subcutaneous or intramuscular injection of insulin, tuberculin, immunization serum, or other injectable medications.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Submission Summary (Full Text)

{0} K961877 # 1.0 510(k) SUMMARY AUG - 7 1996 (1) Submitter’s Information UNIVEC Inc. 999 Franklin Avenue Garden City, New York 11530 (516) 294-1000 (tel) (516) 739-3343 (fax) (2) Device Name Proprietary Name: UNIVEC Nonreusable Locking Syringe Common or Usual Names: Piston; Hypodermic Needle Classification Names: Piston Syringe 21 C.F.R. 880.5680 Class II Hypodermic single lumen needle 21 C.F.R. 880.5570 Class II (3) Predicate Device: K 943679, UNIVEC Nonreusable Locking Syringe with non-aspirating plunger (4) Description of Device: The UNIVEC Nonreusable Syringe with Coupling Assembly provides a single use function. This assembly, inserted between the syringe plunger and the syringe rubber stopper, will decouple the plunger from the stopper (and disable the syringe) after injecting one dosage of serum. The assembly materials will be fabricated from the same polypropylene as the syringe. (5) Statement of Intended Use The UNIVEC Nonreusable Locking Syringe with Coupling Assembly can be used by health care professionals, diabetics and other persons for injection of insulin, tuberculin, immunization serum, and other injectable medications. (6) Statement of Technological Characteristics {1} 4 The UNIVEC Nonreusable Locking Syringe with Coupling Assembly has no significant change in design, except for the stainless steel locking clip of the predicate. Otherwise, it has the same material, chemical composition, energy source, or other technological characteristics compared to the predicate device listed in paragraph (3).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...