K961876 · Hitachi Medical Systems America, Inc. · LNH · Jul 25, 1996 · Radiology
Device Facts
Record ID
K961876
Device Name
AIRIS
Applicant
Hitachi Medical Systems America, Inc.
Product Code
LNH · Radiology
Decision Date
Jul 25, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination, surgery planning, or therapy monitoring. In order to ensure the availability for evaluating the accuracy of MRI results, intended patients are those subjects for whom it is anticipated that adequate, independent confirmation of any lesion apparently detected with the MR system can be obtained. With the exception of normal volunteers, eligible patients must have suspected or documented neoplastic, degenerative, infectious, or developmental disease process strongly suspected or established by physical exam, history, or conventional histologic, biochemical, bacteriological, or imaging techniques, or have surgical or aspiration biopsy pending that will be used to establish a diagnosis. The intended patients are primarily drawn from a pool of those subjects undergoing diagnostic evaluation by physicians who are skilled in diagnosis and treatment of the disease process(es) under consideration.
Device Story
Revised RF Transmit Coil for AIRIS MR system; improves RF field uniformity at imaging volume edges to accommodate larger patients. Device functions by applying RF pulses to excite hydrogen protons within a static magnetic field; spatial localization achieved via magnetic field gradients (slice, frequency, phase encoding). System detects radiated RF energy; computer performs Fourier transforms to reconstruct 2D/3D images based on proton density, T1, and T2 relaxation times. Operated by trained physicians/technicians in clinical settings. Output provides anatomical images for diagnosis, surgery planning, and therapy monitoring. Benefits include non-invasive visualization of internal structures without ionizing radiation.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
RF Transmit Coil for MR imaging; utilizes static magnetic field (up to 1.5T) and RF pulses. System includes gradient coils, RF amplifiers, and computer for image reconstruction. Connectivity includes standard MR imaging interfaces. Software performs Fourier transforms for image processing.
Indications for Use
Indicated for patients with suspected or documented neoplastic, degenerative, infectious, or developmental disease processes in the head, body, or extremities, requiring diagnostic evaluation, surgery planning, or therapy monitoring. Contraindicated for patients where independent confirmation of detected lesions cannot be obtained.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
AIRIS
Submission Summary (Full Text)
{0}
Page 12
K961876
JUL 25 1996
Attachment 1
510(k) Summary
of Safety and Effectiveness
{1}
# 1.0 SUBMITTER INFORMATION:
1.1 Submitter: Hitachi Medical Systems America
1963 Case Parkway
Twinsburg, OH 44087
PH: 216 425 1313
FX: 216 425 1410
1.2 Contact: James Jochen Rogers
1.3 Date: May 13, 1996
# 2.0 DEVICE NAME
2.1 Classification Panel: Radiology
2.2 Classification Number: 892.1000 Magnetic Resonance Diagnostic Device
2.3 Product Number: 90LNH
2.4 Product Nomenclature: System, Nuclear Magnetic Resonance Imaging
2.5 Trade/Proprietary Name: Revised RF Transmit Coil
2.6 Predicate Device: AIRIS
# 3.0 DEVICE DESCRIPTION:
3.1 FUNCTION
The AIRIS RF Transmit Coil offers improved RF Field Uniformity at the edge of the Imaging Volume to accommodate larger patients.
3.2 SCIENTIFIC CONCEPTS
MR images are obtained by placing the patient or area of interest within a powerful, highly uniform, static magnetic field. A portion of the protons (hydrogen nuclei) within the patient align with this main magnetic field, similar to small bar magnets. These protons precess, or rotate about the axis of the main magnetic field, much like a spinning child's top. The frequency of rotation is directly proportional to the strength of the main magnetic field, and is given by the Larmor equation, $\omega = 2\pi \gamma B_0$, where $\omega$ is the frequency of rotation, $\gamma$ is the gyromagnetic ratio, and $B_0$ is the strength of the main magnetic field. In the case of Hydrogen protons, the precessional frequency is $4.26\mathrm{MHz}$ for a magnetic field strength of 0.1Tesla. This precessional frequency of rotation is the resonant frequency of the protons. A resistive, permanent, or cryostat magnet provides the strong, homogeneous static magnetic field.
{2}
The Hydrogen nucleus aligns itself in one of two ways, such that it precesses either parallel or anti-parallel to the applied magnetic field. These two alignments have slightly different energy levels, and nuclei are said to be in one of two discrete energy states. The parallel energy state possesses slightly lower energy than the anti-parallel state. Correspondingly, there will be a greater number of nuclei in the lower-energy state than in the higher-energy state.
Energy may be added to the system in the form of radio frequency (RF) waves at the resonant frequency of the precessing nuclei. Some of the nuclei will absorb sufficient energy to flip over from the lower-energy parallel level to the higher-energy anti-parallel level. The more energy added to the system, the greater number of nuclei that will flip over to the higher energy state. If the RF energy source is switched off, the nuclei will gradually lose energy to the surrounding environment, by radiating the energy in the form of RF waves, and flip back to the lower-energy level. A powerful (>5kW) RF Amplifier and RF transmit coils are utilized to introduce RF energy to the system.
The RF signal received immediately after the RF pulse is switched off is proportional to the number of protons resonating in the sample, and is referred to as proton density. The decay of RF signal as the protons return to equilibrium is exponential in nature, and is known as spin-lattice relaxation; its time constant of decay is designated T1. During the RF pulse, the protons precess in synchronization (in phase) with each other. After the pulse is switched off, the protons' magnetic moments interact with each other, and with local variations in the magnetic field. These interactions lead to a gradual, exponential dephasing. This dephasing is known as spin-spin relaxation, with a time constant designated T2. These three quantities, proton density, T1, and T2, vary with tissue, and may be used to form a diagnostic image, provided that they may be spatially, and if necessary, temporally, located.
A variety of sensitive antennae are available to detect this radiated RF energy, however, the 'raw' signal emitted from the patient in response to an RF pulse gives no indication of the position within the patient of the excited nuclei. In order to produce an image, it is necessary to encode positional information into the returned RF signal. This is done by utilizing small magnetic field gradients which alter the local magnetic fields, which, in turn, alter the precessional frequency of the emitting nuclei. Three orthogonal gradients are utilized for slice selection, frequency encoding, and phase encoding (x, y, z). Spatial localization is encoded into the echo by varying the RF excitation and by appropriately applying magnetic field gradients in x, y, and z directions, and changing the direction and strength of these gradients.
If a magnetic field gradient is applied from head to foot of a patient, (HF), then protons in different transverse planes of the patient will experience different local magnetic fields, and will precess at different frequencies. If an RF pulse
Page 14
{3}
of a single frequency is utilized to excite nuclei, only those nuclei in the plane corresponding to the RF pulse's frequency will be excited. Therefore, the return signal will come only from that transverse plane or slice. The signal is thus localized to a single plane. This process is slice selection, and utilizes the slice gradient.
A second magnetic gradient may be applied across the patient, from left to right, (LR), during signal acquisition, so that different sections of the selected plane emit signals of differing frequencies. This process is frequency encoding, and utilizes the read gradient. The read gradient localizes signals in one dimension across the selected plane.
Finally, a third gradient may be applied from anterior to posterior across the patient, (AP), just before the read gradient is applied. This has the effect of altering the phase of precession of the nuclei in the selected slice, by an amount dependent upon their position along the AP axis. This process is phase encoding, and utilizes the phase gradient. The phase gradient localizes signals in the second dimension across the selected plane.
This excitation and follow-up read must be repeated several times, each time with a different phase encode gradient, such that the phase of any point along the AP axis varies with a frequency dependence upon its position. The returned signal is now a complex waveform containing many frequencies and phases. The signal amplitude at each frequency can be calculated using a Fourier transform. A second Fourier transform extracts signal amplitude at each phase. Amplitudes can be assigned to gray scale levels, resulting in the construction of an image. In its simplest form, this amplitude plot shows the proton density in a single plane through the patient.
Images depicting the spatial distribution of NMR characteristics of the nuclei under consideration can be constructed by using image processing techniques similar to those used in CT. A computer is responsible for the generation and timing of the RF and gradient waveforms. High-current analog amplifiers and gradient coils convert the gradient waveforms to slice, frequency and phase gradients across the patient. The computer also the 2D Fourier transforms required for image reconstruction, and provides the operator with a variety of image manipulation and processing techniques.
For magnetic fields up to 1.5T, the RF frequencies commonly used range up to 65MHz. The RF fields have pulse powers from several watts to greater than 10 kilowatts, and repeat at rates from once every few seconds to greater than fifty per second. The time-varying magnetic gradient fields have a typical duration of sub-millisecond to several milliseconds.
Page 15
{4}
Page 16
# 3.3 PHYSICAL AND PERFORMANCE CHARACTERISTICS
MR is currently of great interest because it is capable of producing high quality anatomical images without the associated risks of ionizing radiation. In addition, the biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In x-ray imaging, differences in x-ray attenuation, largely based on differences in electro density are responsible for the contrast observed in x-ray images. In MR imaging, differences in proton density, blood flow, and relaxation times T1 and T2 all may contribute to image contrast. In addition, by varying the duration and spacing of the RF pulses, images may be produced in which the contrast is primarily dependent on T1 relaxation, T2 relaxation, proton density, or a combination of all three.
# 4.0 DEVICE INTENDED USE:
The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination, surgery planning, or therapy monitoring.
In order to ensure the availability for evaluating the accuracy of MRI results, intended patients are those subjects for whom it is anticipated that adequate, independent confirmation of any lesion apparently detected with the MR system can be obtained. With the exception of normal volunteers, eligible patients must have suspected or documented neoplastic, degenerative, infectious, or developmental disease process strongly suspected or established by physical exam, history, or conventional histologic, biochemical, bacteriological, or imaging techniques, or have surgical or aspiration biopsy pending that will be used to establish a diagnosis. The intended patients are primarily drawn from a pool of those subjects undergoing diagnostic evaluation by physicians who are skilled in diagnosis and treatment of the disease process(es) under consideration.
- Anatomical Region: Head, Body, Spine, Extremities
- Nucleus excited: Proton
- Diagnostic uses: 2D T1- / T2-weighted imaging
T1, T2, proton density measurements
MR Angiography
image processing
{5}
- Imaging capabilities:
2D, 3D Spin Echo (SE)
2D Fast Spin Echo (FSE)
2D Inversion Recovery (IR)
2D Fast Inversion Recovery (FIR)
2D, 3D Gradient Echo (GE)
2D, 3D Gradient Echo with Rephasing (GR)
2D, 3D Steady state acquisition with rewinded GE (SARGE)
2D Dual Slice acquisition (DS)
MR Angiography (2D TOF, 3D TOF, MTC, SSP, half echo, high resolution/high definition)
## 5.0 DEVICE TECHNOLOGICAL CHARACTERISTICS:
Identical to the Predicate Device.
Page 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.