← Product Code [DTN](/productcode/DTN) · K961836

# MAXIMA FORTE HARDSHELL VENOUS RESERVOIR (K961836)

_Medtronic Vascular · DTN · Nov 1, 1996 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K961836

## Device Facts

- **Applicant:** Medtronic Vascular
- **Product Code:** [DTN](/productcode/DTN.md)
- **Decision Date:** Nov 1, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4400
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The MAXIMA FORTÉ™ Hardshell Venous Reservoir is intended for use in procedures requiring the storage and filtration of blood in the extracorporeal circuit during cardiopulmonary bypass. The MAXIMA FORTÉ™ Hardshell Venous Reservoir is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for volume replacement.

## Device Story

Single-use, sterile, nonpyrogenic hardshell venous reservoir; collects venous blood and cardiotomy blood during cardiopulmonary bypass. Venous blood enters via inlet port; flows through central tube to bottom of filter/defoamer assembly; passes through polyurethane defoamers. Cardiotomy blood enters via six suction ports; flows through center chamber of filter/defoamer assembly; passes through 20-micron polyester filter and defoamers. Blood streams mix in main chamber; filtered/defoamed blood exits through bottom outlet port. Used in OR by perfusionists/clinicians. Features rotatable turret for inlet positioning, pressure relief valve, and sampling manifold. Enables autotransfusion of shed blood post-surgery. Benefits patient by providing filtered, defoamed blood volume replacement.

## Clinical Evidence

Bench testing only. No clinical data presented. In-vitro testing included breakthrough volume, blood trauma, static and dynamic hold-up volume, pressurization, filtration efficiency, and cleanliness assessments to demonstrate substantial equivalence to predicate devices.

## Technological Characteristics

Polycarbonate housing; integrated filter/defoamer assembly. Filter: 20-micron nominal polyester depth filter. Defoamer: Polyurethane foam. Capacity: 4,000 ml. Flow rates: Venous 1-7 lpm, Cardiotomy 1-5 lpm. Features: rotatable turret, 3-gang sampling manifold, pressure relief valve. Sterile, single-use.

## Regulatory Identification

A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.

## Special Controls

*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Medtronic MAXIMA® Hardshell Venous Reservoir ([K933496](/device/K933496.md))
- Medtronic MAXIMA® Filtered Hardshell Venous Reservoir ([K891230](/device/K891230.md) and [K932254](/device/K932254.md))
- Terumo Medical Corporation CAPIOX SX Hardshell Venous Reservoir ([K922799](/device/K922799.md))
- Avecor Cardiovascular AFFINITY CVR Cardiotomy/Venous Hardshell Reservoir ([K936003](/device/K936003.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

NOV - 1 1996

510(K) SUMMARY
(Summary of Information Respecting Safety and Effectiveness) K961836

## COMPANY AND CONTACT PERSON

Medtronic, Inc.
Cardiopulmonary Division
4633 E. La Palma Avenue
Anaheim, CA 92807
714-779-3700 (phone)
714-779-7964 (fax)

Debra J. Kridner, Manager
Regulatory Affairs

## DEVICE NAME

MAXIMA FORTE™ Hardshell Venous Reservoir
(FT-HSVR)

## CLASSIFICATION NAME

Reservoir, Blood, Cardiopulmonary Bypass

## NAME OF PREDICATE OR LEGALLY MARKETED DEVICE

- Medtronic MAXIMA® Hardshell Venous Reservoir (K933496)
- Medtronic MAXIMA® Filtered Hardshell Venous Reservoir (K891230 and K932254)
- Terumo Medical Corporation CAPIOX SX Hardshell Venous Reservoir (K922799)
- Avecor Cardiovascular AFFINITY CVR Cardiotomy/Venous Hardshell Reservoir (K936003)

510(k) Notification (K961836)
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTE™ Hardshell Venous Reservoir
Addendum I
September 5, 1996
Confidential
Appendix V
Revised - Page 2

{1}

DESCRIPTION OF DEVICE

The Medtronic MAXIMA FORTÉ™ Hardshell Venous Reservoir is a single use, disposable, sterile and nonpyrogenic fluid path device. Venous blood is collected and defoamed while cardiotomy blood is collected, defoamed and filtered prior to mixing with the venous blood.

The Medtronic MAXIMA FORTÉ™ Hardshell Venous Reservoir consists of a polycarbonate housing which incorporates a filter/defoamer assembly. The filter/defoamer assembly, which consists of polyurethane defoamers, a 20 micron (nominal) polyester filter and a styrene frame, is designed to defoam and filter cardiotomy blood and defoam venous blood.

The maximum capacity of the MAXIMA FORTÉ™ Hardshell Venous Reservoir is approximately 4,000 ml. The reservoir housing tapers toward the blood outlet port. The hardshell venous reservoir operates at venous blood flow rates from 1 to 7 liters per minute and cardiotomy blood flow rates from 1 to 5 liters per minute. Following intraoperative use, these reservoirs are used for the collection and autotransfusion of shed blood.

## Venous Blood Flow

Venous blood enters the top of the reservoir through the venous inlet port. The venous blood flows downward through a central tube to the bottom of the filter/defoamer frame where it passes through the venous defoamer and outer defoamer into the main chamber of the reservoir.

## Cardiotomy Blood Flow

Cardiotomy blood enters through any of the six (6) inlet cardiotomy ports. The intrathoracic suctioned cardiotomy blood flows downward through the center chamber of the cardiotomy filter/defoamer assembly, and passes through the defoamers and filter into the main chamber of the reservoir. The center chamber is separate from and surrounds the central tube through which the venous blood flows.

The venous and cardiotomy blood are mixed outside the filter/defoamer assembly, in the main chamber of the reservoir and in the low-volume, tapered area at the bottom of the reservoir. The filtered, defoamed and mixed blood then exits through the outlet port at the bottom of the reservoir.

The lid of the MAXIMA FORTÉ™ Hardshell Venous Reservoir contains various access ports, which include:

- one (1) 1/2" venous inlet port with temperature probe and luer port
- six (6) cardiotomy suction inlet ports,
- three (3) - 1/4"
- two (2) - 3/8"
- one (1) - combination 1/4-3/8"
- two (2) filtered luer ports,
- two (2) unfiltered luer ports,
- one (1) recirculation inlet port, and
- one (1) vent port

510(k) Notification
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTE™ Hardshell Venous Reservoir
April 29, 1996
Confidential
Appendix V
Page 3

{2}

The venous inlet port and cardiotomy inlet ports are mounted on a rotatable turret that allows the clinician optimal positioning during set-up and use. In addition, for ease of use the venous inlet 1/2" port may be converted to a 3/8" port by using a molded adaptor. The reservoir lid also has a;

- three-way stopcock sampling manifold, one-way check valve, and venous blue)/arterial (red) coiled sampling lines,
- a 3/8" recirculation line, and
- a built-in two-way pressure relief valve that is designed to relieve either high positive or low negative pressures.

A 3/8" blood outlet port is located at the bottom of the device.

## STATEMENT OF INTENDED USE

The MAXIMA FORTÉ™ Hardshell Venous Reservoir is intended for use in procedures requiring the storage and filtration of blood in the extracorporeal circuit during cardiopulmonary bypass. The MAXIMA FORTÉ™ Hardshell Venous Reservoir is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for volume replacement.

## STATEMENT OF INTENDED USE OF PREDICATE/MARKETED DEVICES

Medtronic MAXIMA® Hardshell Venous Reservoir (MHR-T) - in procedures requiring the storage and filtration of blood in the extracorporeal circuit during cardiopulmonary bypass.

Medtronic MAXIMA® Filtered Hardshell Venous Reservoir (1315) - in procedures requiring the storage and filtration of blood in the extracorporeal circuit during cardiopulmonary bypass and for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for volume replacement.

Terumo Medical Corporation CAPIOX SX Hardshell Venous Reservoir - to be used during open heart surgical procedures requiring cardiopulmonary bypass for periods up to 6 hours.

Avecor Cardiovascular AFFINITY CVR Cardiotomy/Venous Hardshell Reservoir - To be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

## STATEMENT OF TECHNOLOGICAL CHARACTERISTICS COMPARISON

A table comparing the intended use and technological characteristics of the Medtronic Cardiopulmonary MAXIMA FORTÉ™ Hardshell Venous Reservoirs with the four noted substantially equivalent devices is provided in Attachment 1.

510(k) Notification
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTÉ™ Hardshell Venous Reservoir
April 29, 1996
Confidential
Appendix V
Page 4

{3}

A table comparing the intended use and technological characteristics of the Medtronic Cardiopulmonary MAXIMA FORTÉ™ Hardshell Venous Reservoirs with the four noted substantially equivalent devices is provided in Attachment I.

## DETERMINATION OF SUBSTANTIAL EQUIVALENCE

The Medtronic Cardiopulmonary MAXIMA FORTÉ™ Hardshell Venous Reservoirs are substantially equivalent to other hardshell venous reservoirs currently in commercial distribution. These predicate/marketed devices include:

- Medtronic MAXIMA® Hardshell Venous Reservoir (K933496)
- Medtronic MAXIMA® Filtered Hardshell Venous Reservoir (K891230 and K932254)
- Terumo Medical Corporation CAPIOX SX Hardshell Venous Reservoir (K922799)
- Avecor Cardiovascular AFFINITY CVR Cardiotomy/Venous Hardshell Reservoir (K936003)

The Medtronic Cardiopulmonary MAXIMA FORTÉ™ Hardshell Venous Reservoirs have an intended use which is substantially equivalent to other hardshell venous reservoirs currently in commercial distribution. These predicate/marketed devices include:

- Medtronic MAXIMA® Hardshell Venous Reservoir (K933496) - indicated for only intraoperative use.
- Medtronic MAXIMA® Filtered Hardshell Venous Reservoir (K891230 and K932254) indicated for intraoperative and postoperative use.
- Terumo Medical Corporation CAPIOX SX Hardshell Venous Reservoir (K922799) - indicated for only intraoperative use.
- Avecor Cardiovascular AFFINITY CVR Cardiotomy/Venous Hardshell Reservoir (K936003) - indicated for only intraoperative use.

The Medtronic Cardiopulmonary MAXIMA FORTÉ™ Hardshell Venous Reservoirs have technological characteristics which are substantially equivalent to other hardshell venous reservoirs currently in commercial distribution. These predicate/marketed devices include:

- Medtronic MAXIMA® Hardshell Venous Reservoir (K933496)
- Medtronic MAXIMA® Filtered Hardshell Venous Reservoir (K891230 and K932254)
- Terumo Medical Corporation CAPIOX SX Hardshell Venous Reservoir (K922799)
- Avecor Cardiovascular AFFINITY CVR Cardiotomy/Venous Hardshell Reservoir (K936003)

The design, construction, materials and nominal specifications of the Medtronic Cardiopulmonary MAXIMA FORTÉ™ Hardshell Venous Reservoirs are either identical or substantially equivalent to other hardshell venous reservoirs currently in commercial distribution. These predicate/marketed devices include:

510(k) Notification (K961836)
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTE™ Hardshell Venous Reservoir
Addendum I
September 5, 1996
Confidential
Appendix V
Revised - Page 5

{4}

- Medtronic MAXIMA® Hardshell Venous Reservoir (K K933496)
- Medtronic MAXIMA® Filtered Hardshell Venous Reservoir (K891230 and K932254)
- Terumo Medical Corporation CAPIOX SX Hardshell Venous Reservoir (K922799)
- Avecor Cardiovascular AFFINITY CVR Cardiotomy/Venous Hardshell Reservoir (K936003)

In addition, the in-vitro testing demonstrated that when compared to a predicate device the MAXIMA FORTE™ Hardshell Venous Reservoirs do not significantly affect safety and effectiveness and are substantially equivalent to other commercially distributed hardshell venous reservoirs. The testing included analysis of:

- Breakthrough Volume Testing
- Blood Trauma Testing
- Static Hold-Up Volume Testing
- Pressurization Testing
- Filtration Efficiency Testing
- Dynamic Hold-Up Volume Testing
- Cleanliness

510(k) Notification (K961836)
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTE™ Hardshell Venous Reservoir
Addendum I
September 5, 1996
Confidential
Appendix V
Revised - Page 6

{5}

# Attachment I

# DEVICE COMPARISONS -- GENERAL CHARACTERISTICS AND NOMINAL SPECIFICATIONS

|   | Medtronic, Inc.
MAXIMA FORTE™
Hardshell Venous Reservoir | Medtronic, Inc.
MAXIMA®
Hardshell Venous Reservoirs | Terumo Medical Corporation
CAPIOX SX
Hardshell Venous Reservoir | Avecor Cardiovascular
AFFINITY CVR Cardiotomy
Venous Hardshell Reservoir  |
| --- | --- | --- | --- | --- |
|  510(k) Number: | This submission | K891230 and K932254
K933296 | K922799 | K936003  |
|  Catalog Number(s): | FT-HSVR | 1315 (K891230 and K932254)
MHR-T (K933496) | CX*SXXX | CVR  |
|  Intended Use: | In procedures requiring the storage and filtration of blood in the extracorporeal circuit during cardiopulmonary bypass. For use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for volume replacement. | 1315 and MHR-T
In procedures requiring the storage and filtration of blood in the extracorporeal circuit during cardiopulmonary bypass.
1315
For use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for volume replacement. | To be used during open heart surgical procedures requiring cardiopulmonary bypass for periods up to 6 hours. | To be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.  |
|  Performance Characteristics:  |   |   |   |   |
|  Duration of Use (Maximum) | 6 hours | 6 hours | 6 hours | 6 hours  |
|  Blood Flow Rate (Maximum) |  |  |  |   |
|  Venous (lpm) | 7 | 7 | 4 | 7  |
|  Cardiotomy (lpm) | 5 | 6 | 4 | 6  |
|  Filtration |  |  |  |   |
|  Cardiotomy Filter | 20 micron depth filter | 20 micron depth filter | 20 micron depth filter | 30 micron depth filter  |

510(k) Notification (K961836)

Medtronic, Inc

Medtronic Cardiopulmonary

MAXIMA FORTE™ Hardshell Venous Reservoir

Addendum I

September 5, 1996

Confidential

Appendix V

Revised - Page 7

{6}

Attachment I
DEVICE COMPARISONS -- GENERAL CHARACTERISTICS AND NOMINAL SPECIFICATIONS

|   | Medtronic, Inc.
MAXIMA FORTE™
Hardshell Venous Reservoir | Medtronic, Inc.
MAXIMA®
Hardshell Venous Reservoirs | Terumo Medical Corporation
CAPIOX SX
Hardshell Venous Reservoir | Avecor Cardiovascular
AFFINITY CVR Cardiotomy
Venous Hardshell Reservoir  |
| --- | --- | --- | --- | --- |
|  Technological Characteristics:  |   |   |   |   |
|  Filter/Defoamer Assembly  |   |   |   |   |
|  Filter  |   |   |   |   |
|  Material | Polyester | Polyester | Polyester | Polyester  |
|  Type | Depth | Depth | Depth | Depth  |
|  Defoamer  |   |   |   |   |
|  Material | Polyurethane Foam | Polyurethane Foam | Polyurethane Foam | Polyurethane Foam  |
|  Reservoir  |   |   |   |   |
|  Design and Construction  |   |   |   |   |
|  System | Open Circuit | Open Circuit | Open Circuit | Open Circuit  |
|  Volume Capacity (ml) | 4000 | 3800 | 3000 | 4000  |
|  Minimum Operating Level (ml) | 500 | 300 | 100 | 500  |
|  Filter/Defoamer | Integrated within reservoir | Integrated within reservoir | Integrated within reservoir | Integrated within reservoir  |
|  Materials | Polycarbonate | Polycarbonate | Polycarbonate | Not stated  |
|  Ports/Inlets  |   |   |   |   |
|  Venous Blood Inlet | Yes | Yes | Yes | Yes  |
|  Blood Outlet | Yes | Yes | Yes | Yes  |
|  Cardiotomy Suction Inlet/Ports | Yes | Yes | Yes | Yes  |
|  Prime Ports | Yes | Yes | Yes | Yes  |
|  Luer and Vent Ports | Yes | Yes | Yes | Yes  |
|  Rotatable Turret/Venous Inlet | Yes | No | Yes | Yes  |
|  Sample Manifold - 3 Gang | Yes | No | Yes | Yes  |

510(k) Notification
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTE™ Hardshell Venous Reservoir
April 29, 1996
Confidential
Appendix V
Page 8

{7}

# Attachment I

# DEVICE COMPARISONS -- GENERAL CHARACTERISTICS AND NOMINAL SPECIFICATIONS

|   | Medtronic, Inc.
MAXIMA FORTE™
Hardshell Venous Reservoir | Medtronic, Inc.
MAXIMA®
Hardshell Venous Reservoirs | Terumo Medical Corporation
CAPIOX SX
Hardshell Venous Reservoir | Avecor Cardiovascular
AFFINITY CVR Cardiotomy
Venous Hardshell Reservoir  |
| --- | --- | --- | --- | --- |
|  Technological Characteristics:  |   |   |   |   |
|  Oxygenator/Hardshell Venous
Reservoir May Be Coupled | Yes | Yes | Yes | Yes  |
|  Hardshell Venous Reservoir
Mode of Operation | To collect and defoam venous blood
while cardiotomy blood is collected,
defoamed and filtered prior to mixing
with the venous blood. The defoamed/
filtered blood is then returned to the
patient. | To collect and defoam venous blood
while cardiotomy blood is collected,
defoamed and filtered prior to mixing
with the venous blood. The defoamed/
filtered blood is then returned to the
patient. | To collect and defoam venous blood
while cardiotomy blood is collected,
defoamed and filtered prior to mixing
with the venous blood. The defoamed/
filtered blood is then returned to the
patient. | To collect and defoam venous blood
while cardiotomy blood is collected,
defoamed and filtered prior to mixing
with the venous blood. The defoamed/
filtered blood is then returned to the
patient.  |

510(k) Notification
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTE™ Hardshell Venous Reservoir
April 29, 1996
Confidential
Appendix V
Page 9

---

**Source:** [https://fda-staging.innolitics.com/device/K961836](https://fda-staging.innolitics.com/device/K961836)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
