The B SMART Sensory Response Timer is indicated for use in evaluating neurological status during surgery under regional anesthesia by monitoring the capability of the patient to detect and to respond to a timed series of mechanical and auditory stimuli.
Device Story
B SMART is a patient-activated sensory response timer used during carotid endarterectomy under regional anesthesia. The system consists of a control unit and a finger-mounted attachment. The control unit generates periodic auditory tones and triggers a vibratory tactile stimulus in the finger attachment at operator-selected intervals (30, 60, or 120 seconds). The patient responds to these stimuli by pressing a cycle reset button on the finger attachment, which terminates the stimulus. If the patient fails to respond, the continued audio tone alerts the anesthesiologist to potential cerebral ischemia or neurological dysfunction. The device allows for multiple-site monitoring by connecting up to four finger attachments to one control unit. It provides a standardized, objective method for monitoring patient consciousness compared to manual methods like handgrips or squeeze toys, helping clinicians decide when to intervene, such as by inserting a temporary shunt.
Clinical Evidence
Clinical study of 47 patients undergoing carotid endarterectomy under regional anesthesia. The B SMART system was used successfully in 45 patients (95.7%). Two patients (4.3%) were excluded due to pre-existing motor impairment. Monitoring confirmed proper sensory/motor function in 44 patients (93.6%) and successfully detected neurological dysfunction in 3 patients (6.4%), allowing for successful treatment via temporary shunt insertion.
Technological Characteristics
System consists of a control unit and finger attachment connected by electrical cable. Powered by two 1.5V AA alkaline batteries (3V DC total). Features include manual slide switch for stimulus intervals (30, 60, 120s), audio tone generator, and vibratory tactile stimulator. Finger attachment uses an impedance-pad-type reset button. Device is portable, battery-operated, and designed for clinical use. Electromagnetic compatibility testing confirmed compliance with radiated emissions and immunity standards.
Indications for Use
Indicated for patients undergoing surgery under regional anesthesia to evaluate neurological status by monitoring the patient's ability to detect and respond to mechanical and auditory stimuli. Contraindicated in patients with motor impairment preventing response to stimuli.
Regulatory Classification
Identification
A biofeedback device is an instrument that provides a visual or auditory signal corresponding to the status of one or more of a patient's physiological parameters (e.g., brain alpha wave activity, muscle activity, skin temperature, etc.) so that the patient can control voluntarily these physiological parameters.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter when it is a prescription battery powered device that is indicated for relaxation training and muscle reeducation and prescription use, subject to § 882.9.
Predicate Devices
Nicolet Biomedical Instruments Corp. Expert Sleep/Wake Analysis System (K873535)
Submission Summary (Full Text)
{0}
K961817
# 510(k) SUMMARY
JUL 23 1997
## 1. COMPANY INFORMATION
Name (Manufacturer) : Hayashi Denki Co., Ltd.
Name (U.S. Agent) : Koven Technology, Inc.
Address (Manufacturer) : 7-11 Arima, 2-Chome Miyamae-ku Kawasaki 216, Japan
Address (U.S. Agent) : 300 Brookes Drive, Suite 105 St. Louis, MO 83042
Telephone (Manufacturer) : 011-81-44-877-4361
Telephone (U.S. Agent) : (314) 731-0008
Contact Person : Mr. Paul G. Koven, President Koven Technology, Inc.
300 Brookes Drive, Suite 105 St. Louis, MO 83042
(314) 731-0008
Date Prepared : May 8, 1996
Revised October 10, 1996
Revised June 11, 1997
## 2. DEVICE NOMENCLATURE.
Trade Name : B SMART™ (Buchbinder Sensory Motor Activated Response Timer)
Common Name : Sensory Response Timer
Classification Name : Evoked Response Mechanical Stimulator [84 GZP, 882.1880]
## 3. PREDICATE DEVICE.
Nicolet Biomedical Instruments Corp. : Expert Sleep/Wake Analysis System
510(k) Number : K873535
SE Decision : October 26, 1987
## 4. DEVICE DESCRIPTION.
The B SMART Sensory Response Timer system consists of two components connected by an electrical cable. These components are a Control Unit and a Finger Attachment. The complete system is shown in Figure 1.
{1}

Figure 1. B SMART System (Control Unit and Finger Attachment)
The Control Unit contains a power supply, a cycle timer, an audio tone generator and speaker, and necessary controls, indicators, and connectors. There are three controls: the power on/off switch, the stimulus cycle selector switch which is used to set the interval between successive activations of the stimulator, and an audio tone volume control. The indicators are power-on and battery status (an LED indicator). The connector is a 3.5mm mini-tip receptacle for the interconnect cable. A block diagram of the Control Unit is shown in Figure 2 below:

Figure 2. Block Diagram of Control Unit
{2}
The Finger Attachment contains a vibratory tactile stimulus mechanism and an impedance-pad-type cycle reset button housed in a small case which is attached to the patient's finger using a Velcro band. When the patient feels the vibratory stimulus (or hears the speaker tone), he or she presses the cycle reset button to terminate the stimuli. Cessation of the audio tone informs the attending physician or nurse that the patient is awake and able to respond. If the audio tone does not cease, the anesthesiologist is alerted that intervention may be needed due to impairment of neuromuscular responsiveness.
For special situations, additional Finger Attachments (up to a total of four) may be connected to a single Control Unit for multiple-site monitoring.
The Stimulus Cycle Selector Switch is a manual slide switch that controls the interval between automatic activations of the patient stimulus test cycle. Any of three intervals may be selected by the operator. The available interval settings are 30, 60, and 120 seconds.
The system is powered by two AA-size 1.5-volt alkaline batteries. Total voltage is 3 volts DC. Upon power turn-on, an LED indicator on the Control Unit flashes on and off for 10 seconds to indicate that battery charge status is adequate for proper operation. If the light fails to come on or no audio tone is heard, the exhausted batteries should be replaced with two AA-size alkaline batteries.
## 5. INTENDED USES.
The B SMART Sensory Response Timer is indicated for use in evaluating neurological status during surgery under regional anesthesia by monitoring the capability of the patient to detect and to respond to a timed series of mechanical and auditory stimuli.
## 6. COMPARISON WITH PREDICATE DEVICE.
The subject device, the B SMART system is a patient-activated stimulus/response system intended for use in monitoring neurological status during carotid endarterectomy and related cerebrovascular surgical procedures performed using local or regional anesthesia. The literature states that adequate monitoring is vital during carotid endarterectomy whether performed under general or regional anesthesia to insure that adequate cerebral perfusion is maintained. In addition to ECG and blood pressure monitoring, many different methods of neurological monitoring are in use to detect evidence of cerebral ischemia early enough so that appropriate counter-measures can be instituted.
{3}
The choice of monitoring methods is partly determined by the type of anesthesia selected. Many of the institutions where local or regional anesthesia is used during carotid surgery employ very simple monitoring techniques. For example, the anesthesiologist will ask the patient to respond to a handgrip or verbal questions at frequent intervals. Some hospitals use squeeze toys, and others have used fluid-filled plastic bags connected to a pressure transducer.
Although the exchange of hand grips between patient and anesthesiologist, squeeze toys, and fluid-filled bags all have the same intended use as the B Smart system, they are not cleared medical products. It is therefore necessary to compare the B SMART system with more sophisticated equipment capable of being used for the same purpose.
Electroencephalography (EEG) is probably the most widely used form of physiologic monitoring during cerebrovascular surgery. It's use is virtually standard in those patients operated on under general anesthesia. Many surgeons who prefer local or regional anesthesia employ EEG monitoring also. Although conventional EEG is often used, many hospitals employ EEG instrumentation equipped with a computer processing accessory as an aid in interpreting results, which can be difficult if personnel with specific EEG training are not available to assist the anesthesiologist.
The predicate device, the Nicolet Expert Sleep/Wake Analysis System, is a computer processing accessory for use in analyzing EEG and other physiologic signals for the evaluation of waking and sleeping states. Although the program is optimized for use in sleep disorder laboratories, the EEG analysis programming incorporated in the Nicolet system is very similar to the several versions of "processed EEG" widely used in carotid endarterectomy monitoring. When used in the awake patient during surgery under local or regional anesthesia, processed EEG systems monitor the neurological status of the patient and detect lapses into unconsciousness signifying the onset of cerebral ischemia. Therefore, an approved computerized EEG processing accessory, the Nicolet Expert Sleep/Wake Analysis System, was selected as the predicate device for the B SMART system because it is capable of performing the same function in the same type of patient in the same surgical environment as the subject device.
7. PERFORMANCE DATA.
Buchbinder and Melick studied the performance of a prototype version of the B SMART system over a period of three years in 47 patients in whom carotid endarterectomy was performed under regional anesthesia. Results were reported in a scientific exhibit presented at the 81st Clinical Congress of the American College of Surgeons, New Orleans, LA, October 22-27, 1995.
{4}
The B SMART system was employed successfully in 45 of the 47 patients (95.7%). Two patients (4.3%) could not be monitored satisfactorily due to motor impairment caused by partial paralysis from prior cerebral vascular accidents. Monitoring demonstrated proper sensory and motor function during endarterectomy in 44 patients (93.6%). Neurological dysfunction was detected in three patients (6.4%) and was treated successfully by immediate insertion of a temporary shunt.
The B SMART system permits quick and convenient verification of proper operation. Before placement on the patient, the operator can mount the Finger Attachment on his or her own finger and check that the mechanical stimulus is activated and that the cycle reset button turns off the audio tone and that the next stimulus/response cycle occurs at the selected time. A watch may be used to verify performance of the cycle selector switch on the Control Unit.
Electromagnetic compatibility testing was conducted by an independent laboratory. Test results demonstrated that the auditory and vibratory signals produced by the B SMART system conform with specifications for radiated emissions and thus will not cause electromagnetic interference (EMI) with other equipment used in the same environment. The B SMART system was also found to meet EMI immunity standards is that operating integrity is maintained in the presence of electromagnetic signals generated by other emission sources.
8. CONCLUSION.
The intended use of the B SMART Sensory Response Timer is identical to one of the potential diagnostic uses--monitoring consciousness during carotid endarterectomy under regional anesthesia--for which the predicate device may be employed. The subject device raises no new questions concerning safety or effectiveness. All pertinent product data and all required certifications have been filed with the Food and Drug Administration. Koven Technology, Inc. concludes that the subject device is substantially equivalent to devices of the same classification previously legally introduced into interstate commerce in the United States.
{5}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Mr. Paul G. Koven
President and
Official Correspondent
Koven Technology, Inc.
300 Brookes Drive, Suite 105
St. Louis, Missouri 63042
JUL 23 1997
Re: K961817
Trade Name: B SMART Buchbinder Sensory Motor Activated Response Timer
Regulatory Class: Unclassified
Product Code: 84LEL
Dated: June 18, 1997
Received: June 19, 1997
Dear Mr. Koven:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}
Page 2 - Mr. Paul G. Koven
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{7}
Page 1 of 1
510(k) Number (if known): K961817
Device Name: B SMART Buchbinder Sensory Motor Activated Response Timer
Indications For Use:
The B SMART Sensory Response Timer is indicated for use in evaluating neurological status during surgery under regional anesthesia by monitoring the capability of the patient to detect and to respond to a timed series of mechanical and auditory stimuli.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Thomas J. Cullahan
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K961817
Prescription Use ☑
(Per 21 CFR 801.109)

(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.