BIOCIRCUITS IOS SERUM PREGNANCY TEST CARTRIDGES/IMMUNOASSAY CONTROLS
K961792 · Biocircuits Corp. · DHA · Aug 21, 1996 · Clinical Chemistry
Device Facts
Record ID
K961792
Device Name
BIOCIRCUITS IOS SERUM PREGNANCY TEST CARTRIDGES/IMMUNOASSAY CONTROLS
Applicant
Biocircuits Corp.
Product Code
DHA · Clinical Chemistry
Decision Date
Aug 21, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1155
Device Class
Class 2
Indications for Use
The IOS™ Serum Pregnancy Test Cartridge is to be used for the qualitative determination of human chorionic gonadotropin (hCG) in serum for the early detection of pregnancy. It is intended to be used with the IOS™ instrument in clinical laboratories, physicians' office laboratories, and other alternate sites of use close to the point of patient care. The IOS™ Immunoassay Controls Kit is to be used to assist in monitoring accuracy and precision in the IOS™ immunoassay test cartridges.
Device Story
Device is a two-site simultaneous sandwich immunoassay for serum hCG detection. Input: patient serum sample added to IOS™ cartridge. Operation: sample rehydrates dried reagents; hCG binds to alkaline phosphatase-labeled monoclonal antibody; complex binds to immobilized polyclonal capture antibody on plastic cartridge. Instrument performs automated wash steps and adds buffer to rehydrate substrate. Output: fluorescent signal generated by substrate-enzyme reaction, measured via front-surface fluorometry. Instrument compares signal rate to barcode-stored cut-off value to provide qualitative positive/negative result. Used in clinical labs and physician offices by technologists or office staff. Benefits: rapid, automated, objective interpretation compared to manual visual assays; reduces operator error via automated reagent delivery and standardized interpretation.
Clinical Evidence
Bench testing established 25 mIU/mL analytical sensitivity and confirmed no interference from common substances (LH, FSH, TSH, hemoglobin, bilirubin, cholesterol, triglycerides). Accuracy study compared IOS to ICON II using 142 positive and 149 negative samples, showing 100% and 98% agreement respectively. Clinical study in three physician office labs (n=172) showed 98.7% agreement for positive samples and 100% for negative samples compared to predicate.
Technological Characteristics
Fluorometric enzyme immunoassay; sandwich format. Components: plastic cartridge, immobilized polyclonal capture antibody, alkaline phosphatase-labeled monoclonal antibody, 4-MUP substrate. Automated fluidics for reagent rehydration and washing. Front-surface fluorometry for signal detection. Barcode-based calibration and lot-specific control range management. Standalone instrument platform.
Indications for Use
Indicated for qualitative detection of hCG in serum for early pregnancy diagnosis. Applicable to clinical settings including physician offices and labs. No specific age or gender contraindications listed, though intended for use in patients where pregnancy status is clinically relevant.
Regulatory Classification
Identification
A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.
Predicate Devices
Hybritech ICON II hCG ImmunoConcentration Assay
Submission Summary (Full Text)
{0}
K961792
AUG 21 1996
510(k) Summary of Information Respecting Safety and Effectiveness
A. Submitter:
Biocircuits Corporation
1324 Chesapeake Terrace
Sunnyvale, CA 94089
(408) 745-1961
Contact: Sheila Ramerman
Date Prepared: August 12, 1996
B. Device Names:
Proprietary Name: Biocircuits IOS™ Serum Pregnancy Test Cartridges
Biocircuits IOS™ Immunoassay Controls
(formerly IOS™ Thyroid Controls)
Common Names: Reagents for hCG (human chorionic gonadotropin)
Quality control materials (assayed and unassayed)
Classification Name: human chorionic gonadotropin (hCG) test system
Quality control materials (assayed and unassayed)
C. Legally Marketed Device:
The IOS Serum Pregnancy Test Cartridges are substantially equivalent to the Hybritech ICON II hCG ImmunoConcentration Assay currently manufactured and distributed by Hybritech Incorporated.
D. Device Description:
Serum Pregnancy Test Cartridges: Human chorionic gonadotropin is a glycoprotein that is synthesized by the placenta during pregnancy. hCG appears in both serum and urine relatively soon after implantation of the developing embryo. The presence of hCG, and its rapid rise following conception, is thus the basis of pregnancy testing. hCG testing can aid in the determination of ectopic pregnancy and spontaneous abortion. hCG is also secreted by a wide variety of tumors: gestational trophoblastic tumors, testicular and prostatic tumors, and some breast cancers.
hCG is composed of two noncovalently linked polypeptides: the alpha and beta subunits. The individual subunits lack biological activity, but become active when linked to form the intact hormone. The alpha subunit of hCG is structurally similar to the alpha subunits of LH (luteinizing hormone), FSH (follicle stimulating hormone), and TSH (thyroid stimulating hormone). The beta subunits of each of the hormones are structurally unique and confer the biologic and immunologic specificity to each hormone. The beta subunit of hCG contains a unique chain of 30 carboxy-terminal amino acids that confers its specificity (1). Monoclonal antibodies directed against this portion of the beta subunit permit differentiation between hCG and the other pituitary glycoprotein hormones.
{1}
# Principle of the Test
The IOS™ Serum Pregnancy test is a two-site simultaneous sandwich immunoassay which utilizes two antibodies directed against intact hCG: a polyclonal antibody immobilized on the plastic cartridge surface (capture antibody) and a monoclonal antibody labeled with alkaline phosphatase (detection antibody, 'conjugate'). hCG in the patient sample binds to conjugate in the test cartridge, and this hCG-conjugate complex then binds to the immobilized capture antibody. After a short incubation period, excess hCG-conjugate complex is washed away. Substrate is added, which reacts with the alkaline phosphatase conjugate-hCG-capture antibody 'sandwich' and produces a fluorescent signal. The level of fluorescence is directly proportional to the amount of bound hCG-conjugate, which is directly proportional to the amount of hCG present in the patient sample. All reagents necessary to perform the test are dried in the IOS™ cartridge, and are rehydrated by the addition of patient sample by the operator, or by the addition of buffer by the instrument.
To perform a test, the operator inserts an IOS™ Serum Pregnancy Test Cartridge in the IOS™ instrument. When prompted, the operator adds sample to a sample well. Each cartridge can test two different patient samples, with one patient sample added to the left well and the second to the right well. The operator starts the test sequence when sample addition is complete. Patient sample rehydrates conjugate dried in the sample well; the conjugate binds to hCG present in the patient serum sample. This hCG-conjugate complex then flows into the incubation/reaction chamber where it binds to the immobilized capture antibody. At the end of the incubation time, excess sample-conjugate is washed away by buffer. Buffer is used to rehydrate the dried substrate necessary for signal generation and quantitation of hCG, in a second reaction chamber. Substrate flows into the incubation/reaction chamber where signal is produced as a result of substrate reacting with conjugate bound to captured patient hCG (the 'sandwich'). The fluorescent signal is read as a rate by front-surface fluorometry, compared to the rate for the cut-off value stored in the barcode on the test cartridge, and determined to be higher or lower than the cut-off value.
Immunoassay Controls: The use of materials derived from human blood to monitor quality control of clinical chemistry testing in the clinical laboratory has been widely established over the past several years. The Biocircuits IOS™ Immunoassay Controls are two levels of blood-based material for use with Biocircuits IOS™ Test Cartridges.
To run a control, the operator inserts the Control Cartridge (packaged with the controls) into the IOS™ instrument. The instrument reads the lot number and ranges of acceptable values for the control solutions from the Control Cartridge barcode, and then ejects the Control Cartridge. The operator then inserts a test cartridge and follows the instrument prompts to identify the control level, apply control solutions, and begin the test sequence. The IOS™ instrument performs the required buffer additions to rehydrate assay reagents and perform wash steps as necessary, reads the fluorescence signal generated, and calculates and prints the control result just as it would if the cartridge were used to test a patient sample.
# E. Intended Use:
The IOS™ Serum Pregnancy Test Cartridge is to be used for the qualitative determination of human chorionic gonadotropin (hCG) in serum for the early detection of pregnancy. It is intended to be used with the IOS™ instrument in clinical laboratories, physicians' office laboratories, and other alternate sites of use close to the point of patient care.
The IOS™ Immunoassay Controls Kit is to be used to assist in monitoring accuracy and precision in the IOS™ immunoassay test cartridges.
{2}
F. Comparison with the Predicate Device:
Table I summarizes the comparative features of the IOS and Hybritech assays.
G. Performance Data:
1. Serum Pregnancy Test Cartridges:
Non-clinical testing performed in the manufacturer's laboratories gave the following results:
**Sensitivity (Detection Limit or Cut-off Value)**
The analytical sensitivity of the IOS™ Serum Pregnancy Test is defined as the detection limit, or cut-off value, of the assay. Sixty replicates of a zero calibrator were run using IOS™ Serum Pregnancy Test Cartridges. The mean value and standard deviation of the signal were calculated. The sensitivity was obtained by interpolating two standard deviations above the mean signal of the zero calibrator. The analytical sensitivity of the IOS™ Serum Pregnancy Test is 25 mIU/mL.
The cut-off value was confirmed by spiking serum samples from normal males with different levels of hCG below, at, and above the cut-off value. Samples containing 0, 15, 20, 25, 30, 35, 40 and 100 mIU/mL hCG were tested in IOS™ Serum Pregnancy Test Cartridges. The IOS™ Serum Pregnancy Test Cartridges gave positive results in 100% of the samples containing 25 mIU/mL hCG; all samples containing >25 mIU/mL hCG also tested 100% positive.
**Accuracy (Correlation)**
Serum samples were tested using the IOS Serum Pregnancy Test and the Hybritech ICON II HCG ImmunoConcentration™ Assay, a qualitative test for serum hCG. Results obtained using the IOS Serum Pregnancy Test agreed with the ICON II Assay in 142 (100%) of the positive samples tested (n = 142). The IOS Serum Pregnancy Test agreed with the ICON II Assay in 146 (98%) of the negative samples tested (n = 149).
**Cross-Reactivity**
Potential cross-reacting substances were added to serum samples containing 0 and 100 mIU/mL hCG at the levels specified below, and tested in the IOS Serum Pregnancy Test. None of the substances interfered with the test results.
- LH at 500 mIU/mL
- FSH at 500 mIU/mL
- TSH at 500 uIU/mL
**Interfering Substances**
Potential interfering substances were added to serum samples containing 0 and 100 mIU/mL hCG at the levels specified below, and tested in the IOS Serum Pregnancy Test. None of the substances interfered with the test results.
- Hemoglobin at 500 mg/dL
- Bilirubin (unconjugated) at 20 mg/dL
- Cholesterol at 750 mg/dL
- Triglycerides at 750 mg/dL
{3}
Clinical testing performed in three typical physicians' office laboratories gave the following results:
Samples were tested by users in three typical physicians' office laboratories. Sites of testing included an OB/GYN practice, an Internal Medicine practice, and a Reproductive Endocrinology/Fertility practice. Users included medical technologists, medical assistants, and front-office personnel. A total of 172 patient samples were tested using the IOS™ Serum Pregnancy Test cartridges in the office laboratories; samples tested included those from pregnant females, non-pregnant females, post-menopausal females, and males. The samples were tested in the manufacturer's laboratory using the ICON II assays. Results obtained using the IOS Serum Pregnancy Test agreed with the ICON II Assay in 81 of 82 (98.7%) of the positive samples tested and in 90 of 90 (100%) of the negative samples tested.
## 2. Immunoassay Controls
The following ranges for the IOS™ Immunoassay Controls were determined in studies in the manufacturer's laboratories. To establish the ranges, the controls were tested in at least 40 cartridges each, over at least 10 days, using several IOS™ instruments. These values only apply to this lot of IOS™ Immunoassay Controls. Different lots of Immunoassay Controls will likely have slightly different ranges. Your laboratory should establish its own range for these controls over time.
| Analyte | Control Level 1 | | Control Level 2 | |
| --- | --- | --- | --- | --- |
| | Mean | Range | Mean | Range |
| T4 (ug/dL) | 8.1 | 6.48-9.72 | 12.9 | 10.32-15.4 |
| T-Uptake (%) | 32.4 | 29.25-35.55 | 39.7 | 36.85-42.55 |
| Serum Preg. Test | Negative | | Positive | |
It is self-evident from the data and information presented here that the Biocircuits IOS™ Serum Pregnancy Test Cartridges are as safe, effective, and perform as well as the Hybritech ICON II hCG Immunoconcentration Assay manufactured and distributed by Hybritech Incorporated.
Attachment: Table I: Assay Comparison
{4}
TABLE 1
Hybritech ICON II vs. Biocircuits IOS™
Assay Comparison
| ATTRIBUTE | Hybritech ICON II hCG | IOS™ Serum Pregnancy Test |
| --- | --- | --- |
| Technology | Immunoconcentration immunoassay | Fluorometric enzyme immunoassay |
| Assay format | Sandwich | Sandwich |
| Enzyme label | Alkaline phosphatase | Alkaline phosphatase |
| Substrate | Bromochloroindolyl phosphate (BCIP) | 4-Methylumbelliferyl phosphate (4-MUP) |
| Reagents | | |
| Immobilization Medium | Paper membrane | Plastic cartridge |
| Dry reagents | Monoclonal antibody only | Monoclonal antibody, polyclonal antibody conjugate, substrate |
| Wet reagents | 3 (for urine) or 4 (for serum), each added by operator as needed | 1 (for all assays), always on board, added by instrument as needed |
| Delivery of wet reagents | Manual, dropwise | Fully automated, precisely measured |
| Stability | Three months at room temperature | Four months at room temperature |
| Storage | Room Temperature (15-30°C) or Refrigerated (2-8°C) | Room Temperature (15-30°C) |
| Sample | | |
| Type | Serum or urine | Serum |
| Volume | 450 uL (serum) | 33 uL |
| | 250 uL (urine) | |
| Preparation | Serum diluted by operator in separate container before addition to icon | None; serum diluted by instrument after addition to cartridge |
| Test Performance | | |
| Hands-on Time | 5+ minutes | Approximately one minute |
| Number of Steps | 6, including serum dilution | One (sample addition to cartridge) |
| Test Result | Qualitative color development | Displayed on screen
Printed alphanumeric hard copy |
| Test Interpretation | By operator
Visual comparison to Positive Reference Zone and to Negative Control Zone | By instrument
Comparison to cut-off value in software |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.