This product is intended for in vitro diagnostic use as a control product to assess the analytical performance of immunochemistry systems and clinical chemistry systems using methods for the quantitative measurement of the listed analytes. It is intended to provide the necessary control material required by inspection agencies for the evaluation of system performance. In addition, it will provide assistance when troubleshooting chemistry systems, reagent problems and calibration anomalies.
Device Story
DOCUMENT® CARDIAC Assayed Control is a human serum-based liquid control product. It contains creatine kinase isoenzymes, lactate dehydrogenase isoenzymes 1 and 2, and myoglobin. Used in clinical laboratories on automated and semi-automated immunochemistry and clinical chemistry systems. Laboratory personnel use the control to assess analytical performance, monitor accuracy and precision, troubleshoot system/reagent issues, and satisfy regulatory inspection requirements for quality control. The device provides a stable, ready-to-use liquid matrix for verifying system performance across three levels (I, II, III).
Clinical Evidence
Bench testing only. Equivalence established by comparing inter-assay precision of the subject device against multiple predicate controls for the specified analytes (CK, CK-MB, LD, myoglobin). Results indicate the device behaves similarly to predicate devices and is suitable for use as a control.
Technological Characteristics
Liquid human serum matrix containing creatine kinase isoenzymes, lactate dehydrogenase isoenzymes 1 and 2, and myoglobin. Provided in three levels (I, II, III) in plastic containers with dropper tips. No dilution required. Stability: 45 days open-vial.
Indications for Use
Indicated for use as an assayed control material to monitor the accuracy and precision of immunochemistry and clinical chemistry systems measuring total creatine kinase (CK) activity, CK-MB mass, CK immunoinhibition, CK electrophoresis, total lactate dehydrogenase (LD) activity, LD isoenzyme electrophoresis (LD1, LD2), and myoglobin.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
Dade CK-MB/Myoglobin Immunoassay Control Level 1,2 and 3 (B5200-1, B5200-2, B5200-3)
Abbott IMx STAT CK-MB Controls, L, M, H (7A28-10)
Bio-Rad Unassayed Chemistry Control (Human) Levels 1 and 2 (16030)
Beckman I.D-Zone Liquid CK Isoenzyme Control (667990)
Beckman I.D-Zone Normal LD Isoenzyme Liquid Control (439090)
Submission Summary (Full Text)
{0}
K961750
JUL 22 1996
Exhibit 21. 510(k) Summary
FEDERAL FOOD, DRUG AND COSMETIC ACT
510(k) SUMMARY
DOCUMENT® CARDIAC Assayed Control, Levels I, II, III
1. Submitted by: CASCO Standards
P.O. Box 970
Yarmouth, ME 04096
Attention: Christine V. Beach
Manager, Regulatory Affairs
(207) 878-7550
(207) 878-7578 FAX
May 5, 1996
2. Product Name:
Proprietary Name: DOCUMENT® CARDIAC Assayed Control
Classification Name: Control, multiple analyte, assayed.
Page • 65 •
{1}
Page • 66 •
# 3. Predicate Device:
Product: Dade CK-MB/Myoglobin Immunoassay Control Level 1,2 and 3
Catalog: Level 1 (B5200-1), Level 2 (B5200-2), Level 3 (B5200-3)
Manufacturer: Baxter Diagnostics, Inc., Deerfield, IL 60015-4633
Analytes: CK-MB, Myoglobin
Product: Abbott IMx STAT CK-MB Controls, L, M, H
Catalog: 7A28-10
Manufacturer: Abbott Laboratories, Abbott Park, IL 60064
Analytes: CK-MB
Product: Bio-Rad Unassayed Chemistry Control (Human) Levels 1 and 2
Catalog: 16030
Manufacturer: Bio-Rad, ECS Division, Anaheim, CA 28506
Analytes: CK, LD
Product: Beckman I.D-Zone Liquid CK Isoenzyme Control
Catalog: 667990
Manufacturer: Beckman Instruments, Brea, CA
Analytes: CK
Product: Beckman I.D-Zone Normal LD Isoenzyme Liquid Control
Catalog: 439090
Manufacturer: Beckman Instruments, Brea, CA
Analytes: LD
# 4. Product Description:
**DOCUMENT**® **CARDIAC** Assayed Controls consist of three levels of a human serum matrix containing creatine kinase isoenzymes, lactate dehydrogenase isoenzymes 1 and 2, and myoglobin. The formulation design provides a liquid matrix intended for use on automated and semi-automated immunochemistry systems and clinical chemistry systems for the determination of total creatine kinase activity (CK), creatine kinase MB mass (CKMB), creatine kinase immunoinhibition (CKII), creatine kinase electrophoresis (CKELEC), total lactate dehydrogenase activity (LD), lactate dehydrogenase electrophoresis for isoenzyme 1 and 2 (LD1; LD2), and myoglobin (MYO).
# 5. Intended Use:
This product is intended for *in vitro* diagnostic use as a control product to assess the analytical performance of immunochemistry systems and clinical chemistry systems using methods for the quantitative measurement of the listed analytes. It is intended to provide the necessary control material required by inspection agencies for the evaluation of system performance. In addition, it will provide assistance when troubleshooting chemistry systems, reagent problems and calibration anomalies.
{2}
6. Comparison to the Predicate Device:
| Characteristic | DOCUMENT
CARDIAC Assayed Control | Dade CK-MB/Myoglobin Immunoassay Control | Abbott CK-MB Stat Control | Bio-Rad Liquichek Unassayed Chemistry Control |
| --- | --- | --- | --- | --- |
| Intended Use | Assayed control for monitoring accuracy and precision of total CK and LD activity, CKMB immunoassay, CK Immunoinhibition, CK isoenzyme electrophoresis, LD1 and LD2 isoenzyme electrophoresis, myoglobin | Assayed control for monitoring of the accuracy and precision of CK-MB and myoglobin analysis | None indicated | Unassayed control for the monitoring of the accuracy and precision of automated and manual testing procedures |
| Number of Levels | Three (3): Level I, Level II, Level III | Three (3): Level 1, Level 2, Level 3 | Three (3): Low, Mid, High | Two (2): Level 1, Level 2 |
| Type | Assayed | Assayed | Assayed | Unassayed |
| Analytes | 3 | 2 | 1 | Many |
| Volume | 2 mL | 3 mL | 8 mL | 5 mL |
| Matrix | Human serum, Liquid | Human source, Lyophilized | Bovine serum, Lyophilized | Human serum, Liquid |
| Dilution | None required | None required | None required | None required |
| Unopened Stability | Until Expiration Date | Until Expiration Date | Until Expiration Date | Until Expiration Date |
| Open Stability | 45 Days | CKMB: 20 Days
MYO: 14 Days | None indicated | 15 Days |
| Container | Plastic, Dropper Tip | Glass | Plastic, Dropper Tip | Glass |
Page • 67 •
{3}
6. Comparison to the Predicate Device (Continued):
| Characteristic | DOCUMENT
CARDIAC Assayed Control | Beckman I.D-Zone Liquid CK Isoenzyme Control | Beckman I.D-Zone Liquid LD Isoenzyme Control |
| --- | --- | --- | --- |
| Intended Use | Assayed control for monitoring accuracy and precision of total CK and LD activity, CKMB immunoassay, CK Immunoinhibition, CK isoenzyme electrophoresis, LD1 and LD2 isoenzyme electrophoresis, myoglobin | Assayed control for monitoring the overall reliability of CK isoenzyme procedures | Assayed control for monitoring the overall reliability of LD isoenzyme electrophoretic procedures |
| Number of Levels | Three (3): Level I, Level II Level III | One (1) | One (1) |
| Type | Assayed | Assayed | Assayed |
| Analytes | 3 | 1 | 1 |
| Volume | 2 mL | 2 mL | 2 mL |
| Matrix | Human serum, Liquid | Human serum albumin, Liquid | Human serum, Liquid |
| Dilution | None required | None required | None required |
| Unopened Stability | Until Expiration Date | Until Expiration Date | Until Expiration Date |
| Open Stability | 45 Days | 20 Days | 20 Days |
| Container | Plastic, Dropper Tip | Plastic Dropper Tip | Plastic, Dropper Tip |
Page • 68 •
{4}
Page • 69 •
# 7. Test Results:
The equivalence for this product was carried out by comparing the inter-assay precision of the listed analytes of the DOCUMENT™ CARDIAC Assayed Control to that of Dade CK-MB/Myoglobin Immunoassay Controls, Abbott CK-MB Stat Controls, Bio-Rad Unassayed Chemistry Control, Beckman I.D-Zone Liquid CK Isoenzyme Control and Beckman I.D-Zone Normal LD Isoenzyme Liquid Control. The results show that the DOCUMENT™ CARDIAC Assayed Control behaves in a similar manner compared to the predicate devices and is suitable for use as a control for the listed analytes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.