# MEDLITE/ERBIUM LASER SYSTEM (K961748)

_Continuum Biomedical, Inc. · GEX · Jul 26, 1996 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K961748

## Device Facts

- **Applicant:** Continuum Biomedical, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jul 26, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

For use in dermatology for the incision, excision, ablation, vaporization and hemostasis of soft tissue.

## Device Story

Medlite Erbium Laser System; 2.94μ wavelength laser; 10W average power (2J pulse energy at 5Hz). System comprises power supply, water cooling, and optical bench. Used by dermatologists for soft tissue surgery; incision, excision, ablation, vaporization, hemostasis. Clinical application includes skin resurfacing and dermabrasion.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Erbium laser; 2.94μ wavelength; 10W average power; 2J pulse energy; 5Hz repetition rate. System includes power supply, water cooling, and optical bench. Class 79-GEX surgical laser.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- SEO Medical TriLase 2940 Erbium Laser System ([K954013](/device/K954013.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) PREMARKET NOTIFICATION
Medlite\Erbium Laser System
Continuum Biomedical, Inc.
Revised July 12, 1996
K961748

# 510(k) Summary

JUL 26 1996

|  Submitter: | Continuum Biomedical, Inc.
2456 Armstrong Street
Livermore, CA 94550-7617
Phone: 510-606-3750
Fax: 510-371-1015  |
| --- | --- |
|  Contact: | Robert S. Anderson, Ph. D
President  |
|  Date Summary Prepared: | July 12, 1996  |
|  Device Trade Name: | Medlite\Erbium Laser System  |
|  Common Name: | Dermatology laser system  |
|  Classification Name: | Instrument, surgical, powered, laser
79-GEX
21 CFR 878.48  |
|  Equivalent Device: | SEO Medical TriLase 2940 Erbium Laser System (K954013)  |
|  Device Description: | The Medlite\Erbium Laser is an erbium laser producing laser emissions at 2.94μ. The laser consists of three interconnected sections: the power supply, the water cooling system and the optical bench. At 5 pulses per second and a pulse energy of 2 Joules, the average power from the laser is 10 watts.  |
|  Intended Use: | For use in dermatology for the incision, excision, ablation, vaporization and hemostasis of soft tissue.  |
|  Comparison: | The TriLase 2940 Erbium Laser has the ability to deliver laser energy at 2.94 microns, maximum power of 5 watts at repetition rates of up to 20 pulses per second. These characteristics are very similar to the Medlite\Erbium Laser System.  |
|  Nonclinical Performance Data: | none  |
|  Clinical Performance Data: | none  |
|  Conclusion: | The Medlite\Erbium Laser System is substantially equivalent to other existing surgical laser systems in commercial distribution for dermatology treatments such as skin resurfacing and dermabrasion.  |
|  Additional Information: | None requested at this time.  |

c:\conb\k\mederb\510k.doc:13

---

**Source:** [https://fda-staging.innolitics.com/device/K961748](https://fda-staging.innolitics.com/device/K961748)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).