DRAGER DVE 8080/81/82 WITH DVE 4002 OR DVE 4001 SERIES SYSTEM
K961742 · Alm Surgical Equipment, Inc. · FQO · Jul 26, 1996 · General, Plastic Surgery
Device Facts
Record ID
K961742
Device Name
DRAGER DVE 8080/81/82 WITH DVE 4002 OR DVE 4001 SERIES SYSTEM
Applicant
Alm Surgical Equipment, Inc.
Product Code
FQO · General, Plastic Surgery
Decision Date
Jul 26, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4960
Device Class
Class 1
Indications for Use
The Drager DVE 8080 System is a piece of durable equipment intended to provide ceiling mounted overhead physical support for anaesthesia machines, medical/surgical monitoring and related equipment, storage space, and ceiling mounted dispensing of medical gases and electrical power required in clinical settings, emergency rooms, intensive care, procedure rooms and operating rooms.
Device Story
Ceiling-mounted modular supply system; provides physical support for anesthesia machines, monitoring equipment, and storage; dispenses medical gases and electrical power. System components include ceiling fixture, support arms (fixed, single, or double), pendant column, and equipment carriers. Features pneumatic brakes for arm positioning and an induction motor for motorized vertical adjustment. Used in ORs, ICUs, and procedure rooms; operated by clinical staff. Output is physical support and utility access; facilitates equipment positioning and workspace organization. Benefits include improved clinical workflow and efficient management of medical gases/power in high-acuity environments.
Clinical Evidence
Bench testing only. Performance validated through UL testing and in-house testing to establish maximum load-bearing capacities and electrical safety compliance.
Technological Characteristics
Modular ceiling-mounted pendant system. Materials: metal construction. Energy: 190V/60Hz three-phase induction motor. Connectivity: integrated medical gas outlets and electrical receptacles. Features: pneumatic brakes, 250 kg load capacity, 23.6-inch vertical travel. Compliance: UL listed, ANSI/NFPA National Electrical Code.
Indications for Use
Indicated for use in clinical settings, emergency rooms, intensive care, procedure rooms, and operating rooms to provide overhead physical support for medical equipment and dispensing of medical gases and electrical power.
Regulatory Classification
Identification
Operating tables and accessories and operating chairs and accessories are AC-powered or air-powered devices, usually with movable components, intended for use during diagnostic examinations or surgical procedures to support and position a patient.
AMSCO Ceiling Mounted Surgical ORBITER System (K831086)
Heraeus Hanauport 810 Universal Ceiling Column
Submission Summary (Full Text)
{0}
JUL 26 1996
K961742
# Summary of Safety and Effectiveness
## INTENDED USE
The Drager DVE 8080 System is a piece of durable equipment intended to provide ceiling mounted overhead physical support for anaesthesia machines, medical/surgical monitoring and related equipment, storage space, and ceiling mounted dispensing of medical gases and electrical power required in clinical settings, emergency rooms, intensive care, procedure rooms and operating rooms.
The DVE 8080 will be used in the same indications as the DVE 4000 System and the DVE 8000/1/2 System and the Drager 8032E which provide physical support, storage space, and overhead, ceiling mounted dispensing of medical gases and electrical power. The DVE 4000 has previously been reviewed by FDA (K921297). The Drager 8000/1/2 Series System and the combined DVE 8000/1/2 with DVE 4002 have been previously reviewed by FDA (K942043). The new DVE 8080 Series will be used in indications which are the same as indications for the DVE 4000, DVE 8000, DVE 8032E and for other pendant systems currently offered for commercial distribution in the United States.
## PRODUCT DESCRIPTION
ALM Surgical Equipment Inc. will be the distributor of the Drager DVE 8080 Series in the United States. The DVE 8080 System is manufactured by Dragerwerk, Aktiengesellschaft Lubeck, Federal Republic of Germany.
The DVE 8080 Series System is a ceiling mounted overhead supply system of modular design. The individual components of any modular system are divided into four categories: ceiling fixture, support systems, pendant head or column, and equipment carrier.
The pendant column contains four terminal plates that can be outfitted according to the individual need with various outlets for gas supply, electricity, vacuum, communications, etc.
Add-on terminal boxes are available which allow the user the opportunity to meet individual equipment needs most efficiently. Add-on terminal boxes can be installed at the time of purchase or anytime afterwards.
{1}
The support system consists of either a fixed vertical support tube (DVE 8080 basic version of the series) or a single (DVE 8081) support arm or two (DVE 8082) support arms. The support arms are equipped with pneumatic brakes and are available in three combinable lengths.
The DVE 8080 Series System rotates laterally and is powered for up and down adjustment. The motor used in the DVE 8080 is an induction motor manufactured by Gefeg GmbH & Co KG. The differences between the DVE 8032E Series motor and the DVE 8080 Series motor are:
| | DVE 8080 | DVE 8032E |
| --- | --- | --- |
| Rated voltage: | 190V/60Hz | 110V/60Hz |
| Rated Input: | 1.37A | 1.7A |
| Speed: | 3300 1/min | 103 1/min |
| P1: | 336W | 255W |
| Capacity: | - | 30u/400V |
| | three-phase current | alternating current |
The equipment shelf provides for better positioning of the medical devices required and can be installed at time of purchase or added-on afterwards. Shelves are available in one standard size: 620 mm (width) by 400 mm.
The Drager DVE 8080/81/82 Systems have been tested and registered with Underwriters Laboratories (UL). All parts are in accordance with the National Electrical Code, ANSI/NFPA.
Maximum load-bearing capacities have been established for possible numbers and lengths of carrying arm combinations. These numbers have been derived from both UL testing and in-house testing.
The Drager DVE 8080 Series Systems can be used alone or in combination with the DVE 4000/1/2 Series System.
## RATIONALE FOR SUBSTANTIAL EQUIVALENCE
The Drager DVE 8080 Series System is substantially equivalent in materials, design, and function to pre-enactment devices of similar intended use, as well as to other pendant anaesthesia supply and monitoring support systems currently in commercial distribution in the United States.
{2}
The Drager DVE 8080 Series is substantially equivalent to:
1) The DVE 8031/2 Series (K921297) and,
2) The DVE 8032E Series (powered for vertical adjustment) (K942043)
3) The AMSCO Ceiling Mounted Surgical ORBITER System,
4) The Heraeus Hanauport 810 Universal Ceiling Column
All of these systems are ceiling mounted pendant systems intended to provide overhead support for anaesthesia machines, medical/surgical monitoring and related equipment and to provide physical support for ceiling mounted dispensing equipment of medical gases and electrical power.
All of these systems are modular in design to allow for maximum flexibility in meeting the needs of individual users and sites.
The main differences between the DVE 8080 Series System and the DVE 8032E Series System are:
1) The position of the motor is changed. The 8032E motor is between the end of the first arm and the beginning of the second arm. The 8080 system motor is located directly above the pendant column.
2) The weight load capacity of the DVE 8080 Series (250 kg) is larger than the weight load capacity of the 8032E (70 kg).
The following charts provide a comparison of features of the DVE 8080 system to two other similar commercially marketed systems.
Table I compares the features of the DVE 8080 Series System to the comparable motorized, vertically adjustable DVE 8032E System.
Table II compares the features of the DVE 8080 Series System to the comparable motorized, vertically adjustable ORBITER system offered for commercial distribution by AMSCO in the United States.
{3}
TABLE I
| | DVE 8080 | DVE 8032E |
| --- | --- | --- |
| Marketed By | ALM Surgical | ALM Surgical |
| Intended Use | Support for Medical Equipment, Storage, Electrical and Gas
Support for Equipment | Support for Medical Equipment, Storage, Electrical and Gas
Support for Equipment |
| Components | Factory Installed Add-ons (prefabrication) | Factory Installed Add-ons (prefabrication) |
| | Vertical Support Tube
2 Arms (single or double)
Pendant Column
Ceiling Fixture | Vertical Support Tube
2 Arms (single or double)
Pendant Column
Ceiling Fixture |
| Features and Technical Specifications | 48 pre-installed medical gas outlets and/or electrical receptacles in any combinations and of any manufacturer. | 24 pre-installed medical gas outlets and/or electrical receptacles in any combinations and of any manufacturer. |
| | Quick connect or D.I.S.S. outlets.
Hospital grade electrical receptacles. | Quick connect or D.I.S.S. outlets.
Hospital grade electrical receptacles. |
| | Fully Rotational, Fully Articulated Or Fixed Height | Fully Rotational, Fully Articulated Or Fixed Height |
| | Pendant Column | Pendant Column |
| | 3 Combinable Lengths For Support Arms | 3 Combinable Lengths For Support Arms |
| Motor | 190 v, 60 Hz | 110 v, 60 Hz |
| Compatible Inserts | Any Manufacturer, Including Chemetron, Puritan Bennett, Ohio Med. Prod., Oxyquip. | Any Manufacturer, Including Chemetron, Puritan Bennett, Ohio Med. Prod., Oxyquip. |
| Allowed Vertical Movement | 23.6 inches | 23.6 inches |
Summary Page 4, May 2, 1996
{4}
Table II
| | DVE 8080 | ORBITER |
| --- | --- | --- |
| Marketed By | ALM Surgical | AMSCO |
| Intended Use | Structural Support for Medical, Electrical and Gas Equipment. Conduit for Wires, Hose. | Support for Medical Equipment, Storage, Electrical and Gas Support for Equipment |
| Components | Factory Installed Add-ons or Site Installed | Factory Installed Add-ons (prefabrication) |
| | Vertical Support Tube 2 Arms (single or double) Pendant Column Ceiling Fixture | Vertical Support Tube Arm (single or double) Pendant Column Ceiling Fixture |
| Features and Technical Specifications | 48 pre-installed medical gas outlets and/or electrical receptacles in any combinations and of any manufacturer. | 12 pre-installed medical gas outlets and/or electrical receptacles in any combinations and of any manufacturer. |
| | Quick connect or D.I.S.S. outlets. Hospital grade electrical receptacles. | Quick connect or D.I.S.S. outlets. Hospital grade electrical receptacles. |
| | Fully Rotational, Fully Articulated Height Adjustable | Fully Rotational, Fully Articulated Height Adjustable |
| | Pendant Column or Head | Variable Lengths, 24 to 34 inches 3 Combinable Lengths |
| | 3 Combinable Lengths For Support Arms | Premarket Notification K831086 |
| Compatible Inserts | Any Manufacturer, Including Chemetron, Puritan Bennett, Ohio Med. Prod., Oxyquip. | Any Manufacturer, Including Chemetron, Puritan Bennett, Ohio Med. Prod., Oxyquip. |
| Motor | 190 v, 60 Hz | |
| Allowed Vertical Movement | 23.6 inches | 28 inches |
Summary Page 5, May 2, 1996
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.