AOC MEDICAL SYSTEM MODEL PHASE II 25 SURGICAL DIODE LASER SYSTEM
K961710 · Applied Optronics Corp. · GEX · Nov 1, 1996 · General, Plastic Surgery
Device Facts
Record ID
K961710
Device Name
AOC MEDICAL SYSTEM MODEL PHASE II 25 SURGICAL DIODE LASER SYSTEM
Applicant
Applied Optronics Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 1, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AOC Phase II 25 surgical diode laser system has the same general intended use as the predicate diode lasers. This intended use includes incision, excision, coagulation and vaporization applications in the following surgical specialties: Urology Gynecology Gastroenterology Pulmonology General and Thoracic Surgery Neurosurgery (hemostasis) Ear, Nose, and Throat; Head/Neck Surgery Dermatology/Plastic Surgery
Device Story
The AOC Phase II 25 is a compact, air-cooled, fiber-coupled diode laser system. It utilizes indium gallium aluminum on gallium arsenide semiconductor diodes to generate 980 nm laser energy. The system is operated by physicians in clinical settings using a front-panel interface, footswitch, or handswitch. It features a visible aiming laser for targeting and supports continuous or pulsed output modes. The device delivers laser energy through disposable optical fibers to perform surgical tasks like incision, excision, coagulation, and vaporization. By providing precise, adjustable power, it enables surgeons to manage tissue effects during procedures. The system's modular design and forced-air cooling eliminate the need for water cooling, facilitating reliable, maintenance-free operation.
Clinical Evidence
No clinical test data submitted. Bench testing confirmed rated power and wavelength output. Peer-reviewed literature on 980 nm tissue interaction provided.
Technological Characteristics
Solid-state diode laser; 980 nm wavelength (range 970-1000 nm); indium gallium aluminum on gallium arsenide semiconductor material. Forced-air cooling; SMA-905 fiber connectors. Microprocessor-controlled; menu-driven interface. Continuous, single, repeat, and exponential pulse modes. Safety features include interlocks, aiming beam, and audible/visual indicators.
Indications for Use
Indicated for patients requiring surgical incision, excision, coagulation, or vaporization in urology, gynecology, gastroenterology, pulmonology, general/thoracic surgery, neurosurgery (hemostasis), ENT/head/neck surgery, and dermatology/plastic surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
AOC 25 (980 nm Diode Laser)
Cynosure Phototome (980 nm Diode Laser)
Cytocare DioLase HP-1 (980 nm Diode Laser)
DioMed 25 (810 nm Diode Laser)
Indigo Medical IDE 830E (830 nm Diode Laser)
Sharplan Lasers 6020 (810 nm Diode Laser)
Submission Summary (Full Text)
{0}
APPLIED OPTRONICS CORP.
K 961710
111 Corporate Blvd. • Building J • South Plainfield, NJ 07080
TEL (908) 753-6300 • FAX (908) 753-4041
NOV - 1 1996
# SECTION K
## 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
### 1. Submitter Information
AOC Medical Systems
A Division of Applied Optronics Corporation
111 Corporate Blvd. Building J
South Plainfield, NJ 07080
Telephone: (908) 753-6300
Fax: (908) 753-4041
Contact Person: Douglass Mead
Regulatory Affairs
Date Prepared: April 30, 1996
### 2. Model AOC Phase II 25 Surgical Diode Laser
### 3. Predicate or Legally Marketed Devices:
| Manufacturer | Model | Laser Type |
| --- | --- | --- |
| AOC Medical Sys. | AOC 25 | 980 nm Diode Laser |
| Cynosure | Phototome | 980 nm Diode Laser |
| Cytocare | DioLase HP-1 | 980 nm Diode Laser |
| DioMed | DioMed 25 | 810 nm Diode Laser |
| Indigo Medical | IDE 830E | 830 nm Diode Laser |
| Sharplan Lasers | 6020 | 810 nm Diode Laser |
### 4. General System Description
The AOC Medical Systems Model Phase II 25 Fiber-Coupled Diode Laser System delivers up to 25 watts of CW optical power from a compact, air cooled unit. It is a complete, self-contained instrument, including a high efficiency power supply, an adjustable light output with constant current operation and a visible aiming laser. The system couples the laser light into high numerical aperture delivery fibers fitted with standard SMA-905 fiber connectors.
The diode lasers in this unit were fabricated by Applied Optronics Corp. from single quantum well, strained-layer, indium gallium aluminum on gallium arsenide semiconductor material. The diode lasers are sealed in
{1}
a rugged, factory-aligned, environmentally protective module. High electrical efficiency of the diode lasers eliminate the need for water cooling and assure low maintenance, reliable laser operation for this all-solid-state system. Each individual diode laser emits at a narrow wavelength; however coupling several diodes through a fiber optic bundle results in each diode wavelength contributing to a broader spectrum emitted form the final delivery fiber. The laser energy is centered at 980nm with a total range from approximately 970-1000nm.
The AOC Phase II 25 has a modular design comprised of:
- Laser Diode and Cooling Module containing the laser diodes, the heat sink and fans.
- Optics Module containing a beamsplitter pick-off for display of the total optical power.
- Front panel for simple control of the unit.
- Transformer isolated power supplies and control electronics.
Unique attributes of this system include the following:
- Multiple, individually fiber-coupled, high power diode lasers mounted on a large thermal footprint to allow sustained high power CW operation without water cooling.
- Optical power level continuously adjustable from the front panel, with display of optical output power.
- Low power, visible aiming laser for beam targeting prior to activation of the high power lasers. The aiming laser is controlled independently of the high power lasers.
- Immediate turn-on capability. No warm-up period. No system alignments or maintenance to be made by the user.
- Forced air cooling (no water). Maximum inlet air temperature 30° C.
- Individually replaceable, fiber pigtailed diode lasers. Repairs performed at factory.
The microprocessor controlled AOC Phase II 25 diode laser system is designed to allow easy and safe operation of the laser using controls common to other surgical laser systems. The laser system utilizes a menu driven set-up operation for user prompts and program entry. The laser can be set for continuous output or programmed for single or repeat pulses at output powers up to 25 watts. A footswitch or handswitch, activated by the user, always controls laser output. An optional remote control for use in a sterile plastic dressing (e.g., standard OR microscope drape) allows the user to change the power and activate Ready and Standby Modes.
The following pre-packaged, presterilized disposable optical fibers are available for use with the AOC Phase II 25 laser system:
{2}
| Fiber Model | Type |
| --- | --- |
| AOC DCLF-600 | 600μm Conical tip |
| AOC DCLF-1000 | 1000μm Conical tip |
| AOC DCLF-600B | 600μm Hemispherical Ball tip |
| AOC DCLF-1000B | 1000μm Hemispherical Ball tip |
| AOC DBLF-60S | 600μm Side firing |
| AOC DBLF-100S | 1000μm Side firing |
| AOC DBLF-60SF | 600μm Scatter Free Side firing |
| AOC DBLF-60 | 600μm Free Beam |
| AOC Neos | Conical (conical shaped) |
| AOC Neos | Delta (wedge shaped) Handpiece, Laparoscopic and Bare Fiber |
| AOC HSD | Conical shaped with electrode |
## 5. Intended Use:
The AOC Phase II 25 surgical diode laser system has the same general intended use as the predicate diode lasers. This intended use includes incision, excision, coagulation and vaporization applications in the following surgical specialties:
- Urology
- Gynecology
- Gastroenterology
- Pulmonology
- General and Thoracic Surgery
- Neurosurgery (hemostasis)
- Ear, Nose, and Throat; Head/Neck Surgery
- Dermatology/Plastic Surgery
No new intended use was sought in this Premarket Notification.
The wavelength (980 nm) and tissue effects of the AOC Phase II 25 diode laser are closely matched to the predicate lasers at 805-830 and 980 nm. The 980 nm wavelength absorption in tissue is less dependent on variations in tissue pigment or vascularity than at 805 nm and has a slightly higher water absorption than the 1064 nm wavelength. Other predicate 980 nm diode laser systems have been found substantially equivalent to Nd:YAG and 805 nm diode lasers.
{3}
4
# 6. Comparison of Technological Characteristics
The AOC Phase II 25 laser output is at the same nominal wavelength, 980 nanometers as the AOC 25, Cynosure Phototome, and Cytocare DioLase HP-1 lasers. Its laser tissue interactions would be expected to be identical. The same individual diode laser components are supplied by Applied Optronics for all four laser systems. The AOC Phase II 25 differs in wavelength from the DioMed 25, Indigo 830 and Sharplan 6020 (at 805 -830 nanometers) which has a slightly greater energy absorption in pigment and blood rich tissue. The 980 nm wavelength, like the Nd:YAG wavelength at 1064 nm, is less dependent on variations in tissue pigment and blood content.
The temporal output of all these systems is the same. These lasers are operated most often in the Continuous Wave (steady state) Mode but can be electronically gated to produce time pulses, single or repeated. The AOC Phase II 25 laser has a Continuous, Single Pulse, Countdown Pulse, Repeat Pulse, and Exponential Pulse output which are found on one or more of the predicate lasers. Its range of timed outputs are within the ranges of the above cleared products.
The output power of the AOC Phase II 25 is about the same as the power of the Diomed 25, Indigo 830, and Sharplan 6020. For applications using contact fibers with sculpted tips which typically operate below 25 watts, the difference in output is not clinically significant. The lower output power of the AOC Phase II 25 in comparison to the CytoCare and Cynosure 980 nm lasers can be compensated for by using smaller diameter fibers for equal power density or exposing target tissue for longer duration (equivalent energy).
The AOC Phase II 25 laser system has the same overall characteristics and features as the predicate diode laser systems. Like these lasers, the AOC Phase II 25 provides for adjustable power output, internal monitoring of power output accuracy, prominent power output display, operation in the Continuous Mode and Single Pulse, and other Modes (with adjustable pulse duration), foot switch activation, safety interlocks, a red aiming beam, Ready and Standby Modes, an electrically safe design, visual and audible indicators of operation, as well as a rugged cabinet designed for the clinical environment.
Like The Sharplan 6020, the AOC Phase II 25 is equipped for handswitch activation of the laser output for convenient surgical control. Also, like the Indigo unit, the AOC Phase II 25 has a Decay (e.g., exponential) Pulse mode.
# 7. Performance Data
No clinical test data was submitted in support of this submission. Peer reviewed clinical articles on the tissue interaction at 980 nm were included in the 510 (k). Bench test data was submitted demonstrating that the design is able to produce the rated power and wavelength.
Notice: This Summary was prepared in order to comply with CFR 21 807.87. AOC Medical Systems makes no claim that any disease or other condition can be cured and promotes its laser as a surgical tool only for use by qualified physicians. Laser surgery, like any surgery, entails necessary risks, some of which may result in complications, injury or, in the extreme, death.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.