# ACUFEX TIBIAL ANCHOR SCREW & SPIKED WASHER (K961649)

_Smith & Nephew Endoscopy, Inc. · HWC · Jul 16, 1996 · Orthopedic · SN_

**Canonical URL:** https://fda-staging.innolitics.com/device/K961649

## Device Facts

- **Applicant:** Smith & Nephew Endoscopy, Inc.
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** Jul 16, 1996
- **Decision:** SN
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Acufex® Tibial Anchor Screw & Spiked Washer is indicated for fixation of soft tissue, such as tendon and ligaments, to bone in orthopedic procedures.

## Device Story

Device consists of two-part tibial anchor screw and spiked washer; used for soft tissue-to-bone fixation in orthopedic procedures. Outer screw (9mm diameter, 20-30mm length) features cancellous threads for bone contact; inner screw (7mm diameter, 20-25mm length) secures spiked washer (14-20mm diameter) to outer screw. Spikes at washer periphery provide tissue grip. Device implanted by orthopedic surgeons in clinical/OR settings to secure tendons/ligaments. Provides mechanical fixation to facilitate healing. Materials include titanium and reinforced polyacetal polymer.

## Clinical Evidence

Bench testing only; no clinical data provided.

## Technological Characteristics

Materials: Titanium and reinforced polyacetal polymer. Design: Two-part screw system (outer 9mm, inner 7mm) with spiked washer (14-20mm). Mechanical fixation principle. Sterile, single-use.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Concept® Soft Tissue Anchoring Implants ([K871037](/device/K871037.md))
- Synthes® Spiked Washer

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

JUL 16 1996
K 96/649

# 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is :

This summary was prepared on April 26, 1996

## A. Submitter

Smith &amp; Nephew Endoscopy
130 Forbes Boulevard
Mansfield, MA 02048

## B. Company Contact

Tedd Gosian
Clinical Regulatory Manager

## C. Device Name

Trade Name: Acufex® Tibial Anchor Screw &amp; Spiked Washer
Common Name: Fixation Screw &amp; Washer
Classification Name: Single/Multiple component metallic bone fixation appliances and accessories

## D. Predicate/Legally Marketed Devices

Concept® Soft Tissue Anchoring Implants (K871037)
Synthes® Spiked Washer

## E. Device Description

Acufex® Tibial Anchor Screw &amp; Spiked Washer is a device used in soft tissue to bone fixation. The Tibial Anchor Screw is a two part screw. The outer larger part has cancellous threads, which come in direct contact with bone, and is 9 mm in diameter with a length of 20 - 30 mm. The inner part, 7 mm diameter and 20 - 25 mm length, is used to attach the Spiked Washer directly to the outer part. The spiked washer is 14 - 20 mm in diameter with spikes set at the periphery.

## F. Intended Use

The Acufex® Tibial Anchor Screw &amp; Spiked Washer is indicated for fixation of soft tissue, such as tendon and ligaments, to bone in orthopedic procedures.

{1}

104

# G. Substantial Equivalence

The AcufeX® Tibial Anchor Screw &amp; Spiked Washer is intended for use for fixation of soft tissue, such as tendon and ligaments, to bone in orthopedic procedures. The AcufeX® Tibial Anchor Screw &amp; Spiked Washer is similar in design, function, materials and intended use to other devices currently marketed and in commercial distribution, namely Concept® Soft Tissue Anchoring Implants (K871037) and Synthes® Spiked Washer.

Risks to health have been addressed through the specified materials, processing controls, quality assurance, and compliance to the Medical Device Good Manufacturing Practices regulations.

A summary comparison of the characteristics of the AcufeX® Tibial Anchor Screw &amp; Spiked Washer and the substantially equivalent device is presented in the table below.

|   | Current Product | Substantially Equivalent Product  |   |
| --- | --- | --- | --- |
|  Attribute I | AcufeX® Tibial Anchor Screw & Spiked Washer | Concept® Soft Tissue Anchoring Implants | Synthes® Spiked Washer  |
|  Indication | Soft Tissue Fixation | Soft Tissue Fixation | Ligament Fixation  |
|  Length | Outer: 20 - 30 mm
Inner: 18 - 25 | Screw: 25 - 55 mm |   |
|  Diameter | Outer: 9 mm.
Inner: 7 mm
Washer: 14 - 20 mm O.D. | Screw: 6.5 mm -
Washer: 14 to 20 mm O.D. | Washer: 13.5 mm O.D.  |
|  Materials | Screw: Titanium
Washer: Titanium
Reinforced Polyacetal Polymer | Screw: Titanium
Washer: Titanium | Washer: Stainless Steel reinforced polyacetal  |
|  Labeling | Sterile, Single Use Only | Sterile, Single Use Only | Sterile, Single Use  |

Applicant _Vicki R. Brown_ Date _4/26/96_

---

**Source:** [https://fda-staging.innolitics.com/device/K961649](https://fda-staging.innolitics.com/device/K961649)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).