PICKER MR WORKSTATIONS (VISTAR,MONTAGE AND ONCALL)
K961637 · Philips Medical Systems (Cleveland), Inc. · LLZ · Jul 3, 1996 · Radiology
Device Facts
Record ID
K961637
Device Name
PICKER MR WORKSTATIONS (VISTAR,MONTAGE AND ONCALL)
Applicant
Philips Medical Systems (Cleveland), Inc.
Product Code
LLZ · Radiology
Decision Date
Jul 3, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The MR Workstations (including the ViStar, Montage and OnCall) are designed to provide fast and convenient review of DICOM compatible digital images from all six diagnostic imaging modalities (MR, CT, Nuclear Medicine, X-Ray, CR and Ultrasound) and secondary capture as well as post-processing of MR and CT images including multi planar reformatting, angiographic rendering, surface rendering, segmentation/disarticulation and cine display. When used in conjunction with any of these diagnostic imaging modalities, the indications for use remain unchanged from those defined for that modality.
Device Story
MR Workstations (ViStar, Montage, OnCall) function as image viewing and processing systems for DICOM-compatible digital images. Inputs include MR, CT, Nuclear Medicine, X-Ray, CR, and Ultrasound data. The system transforms these inputs via software-based post-processing, including multi-planar reformatting, angiographic rendering, surface rendering, segmentation/disarticulation, and cine display. ViStar systems are single-monitor workstations; Montage systems are multi-monitor, multi-modality stations; OnCall systems are PC-based for remote viewing. Operated by clinicians in clinical settings, the device provides visual output on high-resolution monitors. Healthcare providers use these outputs to review diagnostic images and perform image manipulation, facilitating clinical decision-making by providing enhanced visualization of soft tissue structures and diagnostic data.
Clinical Evidence
No clinical data provided. Bench testing only.
Technological Characteristics
Hardware includes DEC/Alpha RISC-based workstations (ViStar/Montage) and Pentium-class PCs (OnCall). RAM ranges from 24-256 MB. Connectivity via Ethernet TCP/IP supporting DICOM v3.0. Monitors include 17-21" high-resolution (1280x1024) color/greyscale displays. Software provides graphical user interface for image review, database management, and post-processing (reformatting, rendering, segmentation).
Indications for Use
Indicated for review and post-processing of DICOM-compatible diagnostic images (MR, CT, Nuclear Medicine, X-Ray, CR, Ultrasound) to assist trained physicians in diagnosis. Patient population is defined by the primary imaging modality used in conjunction with the workstation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
K96/637
# SUMMARY OF SAFETY AND EFFECTIVENESS
JUL - 3 1996
## 1. General Information
Classification: Class II
Image Processing System
Common/Usual Name: Image Viewing and Processing Workstation
Proprietary Name: MR Workstations (ViStar, Montage and OnCall)
Establishment Registration: Picker International, Inc.
World Headquarters
595 Miner Road
Cleveland, Ohio 44143
FDA Owner Number: #1580240
FDA Registration Number: #1525965
Performance Standards: Not Applicable
## 2. Intended Uses:
The MR Workstations (including the ViStar, Montage and OnCall) are designed to provide fast and convenient review of DICOM compatible digital images from all six diagnostic imaging modalities (MR, CT, Nuclear Medicine, X-Ray, CR and Ultrasound) and secondary capture as well as post-processing of MR and CT images including multi planar reformatting, angiographic rendering, surface rendering, segmentation/disarticulation and cine display.
When used in conjunction with any of these diagnostic imaging modalities, the indications for use remain unchanged from those defined for that modality.
## 3. Device Description:
The Picker MR Workstations (ViStar, Montage and OnCall) integrate image review and processing software with a variety of hardware configurations. The software consists of a graphical user interface and post-processing capabilities for MR and CT images including: multi planar reformatting, angiographic rendering, surface rendering, segmentation / disarticulation and cine display. The hardware configurations can be divided into three major types, ViStar, Montage and Oncall. In general, the ViStar line of products are single-monitor workstations which can be equipped with all of the image manipulation and reformatting capabilities described above. The Montage line of products are multi-monitor, multi-modality viewing stations which have limited image manipulation and reformatting capabilities. Finally, the OnCall systems are PC-based for remote viewing and provide image review and cine display.
{1}
# 4. Safety and Effectiveness
The Picker MR Workstations (ViStar, Montage and OnCall) are substantially equivalent to the ViStar (K905070) and the software described with the EDGE/VISTA systems (K931544) in safety and effectiveness. The following chart has been compiled to demonstrate the MR Workstations' substantial equivalence to these devices.
## SUBSTANTIAL EQUIVALENCE CHART
| Parameter | MR Workstations (ViStar/Montage/OnCall) | Predicate Device (ViStar K905070, EDGE/VISTA K931544) |
| --- | --- | --- |
| Computer | ViStar/Montage: DEC/Alpha supecomputer with RISC architecture.
OnCall: Pentium class personal computer platform (e.g. IBM Thinkpad 760C) | Stardent 3000 supercomputer using parallel RISC architecture. (K905070) |
| RAM Memory | ViStar/Montage: Ranges from 32 to 256 MBytes. Expandable for all models.
OnCall: 24 MBytes. Expandable to 64 MBytes. | 64 MBytes. Expandable to 128 MBytes. (K905070) |
| Image Storage Capabilities | ViStar/Montage: Disk storage ranges from 0.5 to 2.1 GBytes, Additional Storage, optical disk drive, VCR Interface and 1/2" streaming tape drive available.
OnCall: Minimum 760 MBytes of Disk Storage. | 760 MBytes disk storage, 1/4" cartridge tape drive, magnetic tape drive. Optional optical disk drive. (K905070) |
PICKER INTERNATIONAL, INC. (MRWKST)
4/23/96
I-2
{2}
| Parameter | MR Workstations (ViStar/Montage/OnCall) | Predicate Device (ViStar K905070, EDGE/VISTA K931544) |
| --- | --- | --- |
| Networking | Ethernet TCP/IP. Supports the DICOM v3.0 imaging/networking protocols. | Ethernet TCP/IP. (K905070) |
| Filming | ViStar/Montage: Direct and Remote Filming Capabilities.
OnCall: Limited formats available. | Direct and Remote Filming Capabilities. (K905070) |
| Monitor | ViStar/Montage: 17-21" color or greyscale high resolution (1280 x 1024) monitor. 72 Hz, non-interlaced.
OnCall: Laptop- Color display with 800 x 600 resolution and SVGA graphics. Desktop- same as ViStar/Montage. | 19 inch, 1280 x 1024 high resolution color monitor, 60 Hz, non-interlaced. (K905070) |
| Graphical Interface | Image review, Database Directory, Draw, Filter, and Filming. | Image review, Database Directory, Draw, Filter, and Filming. (K931544) |
| Image Utilities | Archive, Delete, CD Player, CT Import, HPQ Import, HPQ Tape Import, Export, Camera Summary, Film Summary, Log Problem, Image Processing, Video Review, Video Capture, System Analysis, | Archive, Delete, Camera Summary, HPQ Import (K931544) |
PICKER INTERNATIONAL, INC. (MRWKST)
4/24/96
I-3
{3}
| Parameter | MR Workstations (ViStar/Montage/OnCall) | Predicate Device (ViStar K905070, EDGE/VISTA K931544) |
| --- | --- | --- |
| Image Manipulation & Reformatting Capabilities | ViStar: Cine Display, Multi planar Reformatting, Angiographic Rendering, Surface Rendering & Cubeview, Segmentation / Disarticulation (Optional).
Montage: Cine Display, Multi planar Reformatting, Angiographic Rendering (Optional).
OnCall: Cine Display. | Cine Display, Multi planar Reformatting, Angiographic Rendering, Surface Rendering & Cubeview. (K905070) |
| Intended Use | The ViStar systems used with HPQ and 2055HP systems do not change the existing indications which are: The Picker International Vista HPQ and 2055 HP systems are indicated for use as NMR devices that produce images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis. (K905070)
The Picker MR Workstations (including the ViStar, Montage and OnCall) are designed to provide fast and convenient review of DICOM compatible digital images from all six diagnostic imaging modalities (MR, CT, Nuclear Medicine, X-Ray, CR and Ultrasound) and secondary capture as well as post-processing of MR and CT images including multi planar reformatting, angiographic rendering, surface rendering, segmentation/disarticulation and cine display. When used in conjunction with any of these diagnostic imaging modalities, the indications for use remain unchanged from those defined for that modality. | |
PICKER INTERNATIONAL, INC. (MRWKST)
4/23/96
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.