VERSALOK SCREW ASSEMBLY (PROPOSED NAME)

K961572 · Wrightmedicaltechnologyinc · MNH · Sep 17, 1996 · Orthopedic

Device Facts

Record IDK961572
Device NameVERSALOK SCREW ASSEMBLY (PROPOSED NAME)
ApplicantWrightmedicaltechnologyinc
Product CodeMNH · Orthopedic
Decision DateSep 17, 1996
DecisionSN
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The VERSALOK™ Screw Assembly is supplied as an alternative component for use with the WRIGHTLOCK™ Posterior Spinal Fixation System to fix the spine for appropriate indications. Limited indications apply when used as a pedicle screw. A WRIGHTLOCK™ construct with screws attached to the pedicles of the lumbar and sacral spine (L3 to S1) and autogenous bone graft may be used for the treatment of severe spondylolisthesis (Grade 3 and Grade 4) at the fifth lumbar - first sacral (L5-S1) vertebral joint. The device is intended to be implanted using a posterior surgical approach and removed after development of a solid fusion mass. When not used as a pedicle screw fixation system, various combinations of the system are also indicated to provide temporary stability of the thoracic, thoracolumbar, or lumbar spine (T1 to S1) during bony fusion healing secondary to: (1) Unstable spinal fractures (such as fracture dislocations) or spinal tumors; (2) Degenerative disk diseases of the spine (defined as back pain of diskogenic origin with degeneration of the disk confirmed by radiographic studies); (3) Spinal curvatures (such as idiopathic scoliosis, neuromuscular scoliosis/kyphoscoliosis with associated paralysis or spasticity, and secondary to spinal fractures) which are: progressive, despite other forms of treatment; detrimental to cardiopulmonary function; interfering with spinal mechanics or causing severe back pain; or cosmetically unacceptable, progressive, and painful.

Device Story

Polyaxial screw assembly (screws, locking sleeves, caps) designed as component for WRIGHTLOCK™ Posterior Spinal Fixation System; provides spinal stabilization via posterior surgical approach. Features top-loading design; allows 6mm rod offset and 30-degree angulation in all planes prior to locking. Used by orthopedic/neurosurgeons in hospital setting to fix spine during bony fusion healing. Provides temporary stability for fractures, tumors, degenerative disk disease, or spinal deformities; removed after solid fusion mass development. Benefits patient by restoring spinal stability and facilitating fusion.

Clinical Evidence

Bench testing only. Fatigue endurance limit: 353N; axial push strength: 1096N; torsional grip strength: 1.95Nm. Results indicate adequate performance under physiological loading.

Technological Characteristics

Stainless steel (ASTM F 1314). Polyaxial screw assembly with locking sleeves and caps. Top-loading design. 6mm offset capacity; 30-degree angulation range. Mechanical fixation device.

Indications for Use

Indicated for patients requiring spinal fixation for severe spondylolisthesis (Grade 3-4, L5-S1) when used as pedicle screws with autogenous bone graft. Also indicated for temporary stability of thoracic, thoracolumbar, or lumbar spine (T1-S1) for unstable fractures, tumors, degenerative disk disease, or progressive spinal curvatures (scoliosis/kyphoscoliosis) when not used as pedicle screw fixation.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0} 1 1 1 ![img-0.jpeg](img-0.jpeg) # WRIGHT MEDICAL TECHNOLOGY, INC. 5677 AIRLINE ROAD ARLINGTON, TN 38002 901-867-9971 K961572 SEP 17 1996 Contact Person: Judy English Date Prepared: September 8, 1996 # 510(k) Summary Trade Name: VERSALOK™ Screw Assembly, Product line addition to WRIGHTLOCK™ Posterior Spinal Fixation System Common Name: Spondylolisthesis Spinal Fixation Device System Predicate Devices: WRIGHTLOCK™ Posterior Spinal Fixation System (Wright Medical Technology, Inc.) MOSS® Miami System (DePuy) This 510(k) summary is being submitted in accordance with the requirements of 21 CFR §807.92. Description/Intended Use The VERSALOK™ Screw Assembly is a polyaxial screw assembly consisting of screws, locking sleeves, and caps. The system is designed to accommodate the rods of the commercially available WRIGHTLOCK™ Posterior Spinal Fixation System (K950074-SE 08/11/95). The system features a low profile as well as a top loading design. It allows for up to 6mm offset from the rod and up to 30 degrees of angulation in all planes prior to final locking. The VERSALOK™ Screw Assembly is available in a range of sizes to fit varying anatomical requirements. The VERSALOK™ Screw Assembly is supplied as an alternative component for use with the WRIGHTLOCK™ Posterior Spinal Fixation System to fix the spine for appropriate indications. Limited indications apply when used as a pedicle screw. A WRIGHTLOCK™ construct with screws attached to the pedicles of the lumbar and sacral spine (L3 to S1) and autogenous bone graft may be used for the treatment of severe spondylolisthesis (Grade 3 and Grade 4) at the fifth lumbar - first sacral (L5-S1) vertebral joint. The device is intended to be implanted using a posterior surgical approach and removed after development of a solid fusion mass. When not used as a pedicle screw fixation system, various combinations of the system are also indicated to provide temporary stability of the thoracic, thoracolumbar, or lumbar spine (T1 to S1) during bony fusion healing secondary to: (1) Unstable spinal fractures (such as fracture dislocations) or spinal tumors; (2) Degenerative disk diseases of the spine (defined as back pain of diskogenic origin with degeneration of the disk confirmed by radiographic studies); (3) Spinal curvatures (such as idiopathic scoliosis, neuromuscular scoliosis/kyphoscoliosis with associated paralysis or spasticity, and secondary to spinal fractures) which are: progressive, despite other forms of treatment; detrimental to cardiopulmonary function; interfering with spinal mechanics or causing severe back pain; or cosmetically unacceptable, progressive, and painful. Material Components in the VERSALOK™ Screw Assembly are manufactured from stainless steel conforming to ASTM F 1314. Summary of Technological Differences The VERSALOK™ Screw Assembly is a product line addition to the WRIGHTLOCK™ Posterior Spinal Fixation System and there are few differences between the devices. Both devices have the same intended use and are manufactured from the same material. They are different in that the VERSALOK™ Screw Assembly is a polyaxial screw option to the system that previously included only a monoaxial screw. In addition, the design of the VERSALOK™ Screw Assembly differs from the screw assembly of the MOSS® Miami System in minor design features and locking mechanism; however, the differences only address surgeon preferences and do not impact the function or negatively impact the safety of the device. Testing Summary The VERSALOK™ Screw Assembly tested with the following mean results: fatigue endurance limit of 353N, axial push strength of 1096N, and torsional grip strength of 1.95Nm. Based on these results, the VERSALOK™ Screw Assembly should perform adequately when subjected to normal physiological loading. 000003
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...