BIOCIRCUITS IOS TSH TEST CARTRIDGES/THYROID CONTROLS

K961566 · Biocircuits Corp. · JLW · Nov 18, 1996 · Clinical Chemistry

Device Facts

Record IDK961566
Device NameBIOCIRCUITS IOS TSH TEST CARTRIDGES/THYROID CONTROLS
ApplicantBiocircuits Corp.
Product CodeJLW · Clinical Chemistry
Decision DateNov 18, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1690
Device ClassClass 2

Indications for Use

TSH: The IOS™ TSH Test Cartridge is to be used for the quantitative determination of thyroid stimulating hormone (TSH, thyrotropin) levels in serum using the Biocircuits IOS™ system. Thyroid Controls: The IOS™ Thyroid Controls Kit is to be used to assist in monitoring accuracy and precision in the IOS™ thyroid assays.

Device Story

Biocircuits IOS™ TSH Test Cartridge is a single-use, disposable cartridge for quantitative TSH measurement in serum. Device operates on the Biocircuits IOS™ instrument. Principle is a two-site sandwich immunoassay using monoclonal anti-TSH capture antibody and enzyme-labeled polyclonal anti-TSH detection antibody. Operator inserts cartridge, adds serum sample, and initiates sequence. Instrument performs automated incubation, washing, and reagent rehydration (dried reagents in cartridge). Signal generation via front-surface fluorometry measuring enzyme-substrate reaction rate. Instrument compares rate to stored calibration curves to calculate TSH concentration. Output is displayed on LCD and printed. Used in clinical laboratories or physician office settings to aid in thyroid function diagnosis and monitoring. Thyroid Controls are blood-based materials used to monitor assay accuracy and precision.

Clinical Evidence

Bench testing and clinical laboratory comparison. Precision studies performed at manufacturer and physician office laboratory (POL) using three control levels; overall CVs ranged from 9.2% to 13.8%. Accuracy evaluated by comparing 206 samples (manufacturer lab) and 43 samples (POL) against predicate method, yielding correlation coefficients of 0.98 and 0.978 respectively. Regression equations: y = 0.141 + 1.07x and y = 0.255 + 0.922x.

Technological Characteristics

Two-site sandwich immunoassay; fluorometric enzyme immunoassay; alkaline phosphatase label; methylumbelliferyl phosphate substrate. Plastic cartridge form factor with dried reagents. Automated instrument-based processing (rehydration, incubation, washing). Front-surface fluorometry. Microprocessor-controlled data analysis with stored calibration curves.

Indications for Use

Indicated for quantitative determination of TSH in serum to assist in the diagnosis and management of thyroid disorders, including hypothyroidism and hyperthyroidism, in patients requiring thyroid function assessment.

Regulatory Classification

Identification

A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.

Predicate Devices

Submission Summary (Full Text)

{0} K96/566 NOV 18 1996 510(k) Summary of Safety and Effectiveness A. Submitter: Biocircuits Corporation 1324 Chesapeake Terrace Sunnyvale, CA 94089 (408) 745-1961 Contact: Sheila J. Ramerman Date Prepared: October 25, 1996 B. Device Names: Proprietary Name: Biocircuits IOS™ TSH Test Cartridges Biocircuits IOS™ Thyroid Controls Common Name: Reagents for thyroid stimulating hormone assay Quality control materials (assayed and unassayed) Classification Name: Thyroid stimulating hormone test system Quality control materials (assayed and unassayed) C. Legally Marketed Device: The IOS™ TSH Test Cartridges are substantially equivalent to the Stratus ultra-sensitive hTSH test currently manufactured and distributed by Dade International (formerly Baxter Dade). D. Device Description: Thyroid stimulating hormone (TSH) is a glycoprotein with a molecular weight of about 28,000 daltons that is secreted by the anterior pituitary gland. TSH interacts with a specific receptor on the thyroid gland cell surface, stimulating production and secretion of the thyroid hormones thyroxine (T4) and triiodothyronine (T3). Secretion of TSH is regulated by two factors: 1) a hormone secreted by the hypothalamus called thyrotropin releasing hormone (TRH) which stimulates the pituitary to produce and release TSH; and 2) the concentration of unbound thyroid hormones T4 and T3 in the interstitial fluid of the brain. An increase of the concentrations of T4 and T3 inhibits the production and secretion of TSH, while a decrease in T4 and T3 levels stimulates production and secretion of TSH, forming a negative feedback mechanism. Failure at any level of regulation of the hypothalamic-pituitary-thyroid axis will result in either under-production (hypothyroidism) or over-production (hyperthyroidism) of T4 and/or T3. Primary hypothyroidism is associated with low thyroid hormone levels and elevated TSH levels, while secondary (pituitary) and tertiary (hypothalamic) forms of hypothyroidism are associated with both low levels of T4 and/or T3 and low-to-undetectable levels of TSH. All three forms of hypothyroidism can be differentiated by the patient's TSH response to TRH. Primary hyperthyroidism is associated with high levels of thyroid hormones and very low or undetectable levels of TSH. The TRH challenge test has been used to confirm primary hyperthyroidism. More recently, sensitive TSH assays have been developed which can differentiate overtly thyrotoxic patients from euthyroid individuals. {1} TSH is composed of two noncovalently linked subunits, designated "alpha" and "beta". The alpha portion is essentially identical to the alpha subunits of human luteinizing hormone (hLH), human follicle stimulating hormone (hFSH), and human chorionic gonadotropin (hCG), differing only in carbohydrate moieties. The beta subunits of each of these hormones is structurally unique and contains the biologic and immunologic specificity. Both subunits are needed for biological activity. The Biocircuits IOSTM TSH assay uses a monoclonal antibody against the beta subunit of TSH to specifically capture the patient TSH, and a polyclonal antibody also against the beta subunit for the enzyme conjugate. This allows specific determination of the concentration of TSH without cross-reactivity with other glycoprotein hormones. ## Principle of the Test: **TSH:** The IOSTM TSH test is a two-site sandwich immunoassay. TSH in the patient sample binds to monoclonal anti-TSH antibody in the test cartridge. After a short incubation time, excess sample is washed away and enzyme-labeled polyclonal anti-TSH antibody (conjugate) is added, which binds to any antibody-bound TSH, forming an antibody-antigen-labeled antibody sandwich. After another short incubation time, excess conjugate is washed away and substrate is added. The rate of the enzyme-substrate reaction is directly proportional to the amount of conjugate bound, which is directly proportional to the amount of TSH present in the patient sample. All reagents necessary to perform the test are dried in the IOSTM cartridge, and are rehydrated by the addition of patient sample by the operator, or by the addition of buffer by the instrument. To perform a test, the operator inserts an IOSTM TSH cartridge into the IOSTM instrument. When prompted, the operator adds sample to the sample well and starts the test sequence. The instrument draws the cartridge inside to be incubated. Patient sample flows into the incubation/reaction chamber where patient TSH binds to anti-TSH antibody. At the end of the sample incubation time, excess patient sample is washed away using buffer added by the instrument. Buffer is also added by the instrument to rehydrate dried conjugate in a separate reagent chamber for the next step; the rehydrated conjugate is then allowed to enter the incubation/reaction chamber and bind to antibody-bound TSH. At the end of the conjugate incubation time, excess conjugate is washed away by buffer. Buffer is used to rehydrate the substrate necessary for signal generation and quantitation of TSH in a second reagent chamber; rehydrated substrate is then allowed to enter the incubation/reaction chamber. The fluorescent signal produced is read as a rate by front-surface fluorometry, compared to the rates produced by a series of calibrators stored in instrument memory, and the amount of TSH present in the patient sample is calculated from the stored calibration curve. **Thyroid Controls:** The use of materials derived from human blood to monitor quality control of clinical chemistry testing in the clinical laboratory has been widely established over the past several years. (1) The Biocircuits IOSTM Thyroid Controls are two levels of blood-based material for use with Biocircuits IOSTM thyroid assays test cartridges (T4/TU, T4, or TSH). To run a control, the operator inserts the Thyroid Control Cartridge (packaged with the controls) into the IOSTM instrument. The instrument reads the lot number and ranges of acceptable values for the control solutions from the Control Cartridge barcode, and then ejects the Control Cartridge. The operator then inserts a test cartridge and follows the instrument prompts to identify the control level, apply control solutions, and begin the test sequence. The IOSTM instrument performs the required buffer additions to rehydrate assay reagents and perform wash steps as necessary, reads the fluorescence signal generated, and calculates and prints the control result just as it would if the cartridge were used to test a patient sample. {2} E. Intended Use: TSH: The IOS™ TSH Test Cartridge is to be used for the quantitative determination of thyroid stimulating hormone (TSH, thyrotropin) levels in serum using the Biocircuits IOS™ system. Thyroid Controls: The IOS™ Thyroid Controls Kit is to be used to assist in monitoring accuracy and precision in the IOS™ thyroid assays. F. Comparison with the Predicate Device: Table I summarizes the comparative features of both the IOS™ and Stratus TSH assays. G. Performance Data: **TSH Test Cartridges:** Non-clinical testing performed in the manufacturer's laboratories gave the following results: 1. Precision A laboratory study performed at the manufacturer for within-day and between-day precision in the IOS™ TSH assay used three levels of control material. The following data were obtained: | Control Level | 1 | 2 | 3 | | --- | --- | --- | --- | | Mean (uIU/mL) | 0.78 | 12.02 | 23.60 | | SD, overall (uIU/mL) | 0.11 | 1.18 | 2.67 | | % CV, within-day (n=10) | 6.4% | 9.9% | 11.3% | | % CV, between-day (n=40) | 13.6% | 9.2% | 11.3% | | % CV, overall | 13.5% | 9.8% | 11.3% | 2. Accuracy A comparison of methods obtained by testing 206 patient samples in the manufacturer's laboratories using the IOS™ TSH assay and a commercially available fluorescent enzyme immunoassay gave a correlation coefficient (r') of 0.98, with the line of regression described by the equation $y = 0.141 + 1.07x$. Samples tested ranged from ≤ 0.173 uIU/mL to 26.30 uIU/mL. Clinical testing performed at a typical physician's office laboratory gave the following results: 1. Precision A study was performed in a typical physicians' office laboratory for total precision in the IOS™ TSH assay. Data were collected over twelve working days. The following data were obtained: | Control Level | 1 | 2 | | --- | --- | --- | | Number of replicates | 23 | 18 | | Mean (uIU/mL) | 0.773 | 12.93 | | SD, overall (uIU/mL) | 0.071 | 1.78 | | % CV, overall | 9.2% | 13.8% | {3} # 2. Accuracy A comparison of methods was also performed by users in a typical physicians' office laboratory. A total of 43 patient samples were tested using the IOS™ TSH assay in the office laboratory; the samples were split and sent to the manufacturer's laboratory for testing by the predicate method. These studies gave a correlation coefficient ('r') of 0.978, with the line of regression described by the equation $$y = 0.255 + 0.922x$$. Samples tested ranged from ≤ 0.173 uIU/mL to 23.06uIU/mL. # Thyroid Controls: The following ranges for the IOS™ Thyroid Controls were determined in studies in the manufacturer's laboratories. To establish the ranges, the controls were tested in a total of 40 cartridges each, over at least 10 days, using several IOS™ instruments. These values only apply to this lot of IOS™ Thyroid Controls. Different lots of Thyroid Controls will likely have slightly different ranges. Your laboratory should establish its own range for these controls over time. | Analyte | Control Level 1 | | Control Level 2 | | | --- | --- | --- | --- | --- | | | Mean | Range | Mean | Range | | T4 (ug/dL) | 8.1 | 6.48 - 9.72 | 12.9 | 10.32 - 15.4 | | T-Uptake (%) | 32.4 | 29.25 - 35.55 | 39.7 | 36.85 - 42.55 | | TSH (uIU/mL) | 0.78 | 0.45 - 1.11 | 12.02 | 8.48 - 15.56 | It is self-evident from the data and information presented here that the Biocircuits IOS™ TSH Test Cartridges as as safe, effective, and perform as well as the Stratus ultra-sensitive hTSH assay in commercial distribution by Dade International. Attachment: Table I: Assay Comparison {4} TABLE 1 Baxter STRATUS vs. Biocircuits IOS™ Assay Comparison | ATTRIBUTE | STRATUS hTSH | IOS™ TSH | | --- | --- | --- | | Technology | Fluorometric enzyme immunoassay | Fluorometric enzyme immunoassay | | Assay format | Sandwich | Sandwich | | Enzyme label | Alkaline phosphatase | Alkaline phosphatase | | Substrate | Methylumbelliferyl phosphate | Methylumbelliferyl phosphate | | Reagents | | | | Immobilization Medium | Reaction tab | Plastic cartridge | | Dry | Monoclonal antibody only | Monoclonal antibody, polyclonal antibody conjugate, substrate | | Wet | 2 (assay-specific), loaded by operator at start of each run | 1 (used for all assays), continuously on board | | Delivery | Fully automated | Fully automated | | Calibration | User-generated | Factory-generated | | Calibration Stability | 14 days (minimum) | 90 days (minimum) | | Storage | Refrigerated (2-8°C) | Room Temperature (15-30°C) | | Sample | | | | Type | Serum or plasma | Serum | | Volume | 0.2 ml (minimum) | 0.15 mL | | Measurement Needed | Non-precision | Non-precision | | Operating environment | 22°-32° C | 15°-30° C | | Data analysis | Microprocessor-controlled | Microprocessor-controlled | | | Stored standard curves | Stored standard curves | | Data output | LCD display | LCD display | | | Printed alphanumeric hard copy | Printed alphanumeric hard copy |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...