ACS FRT4 IMMUNOASSAY

K961510 · Ciba Corning Diagnostics Corp. · JJX · Jun 17, 1996 · Clinical Chemistry

Device Facts

Record IDK961510
Device NameACS FRT4 IMMUNOASSAY
ApplicantCiba Corning Diagnostics Corp.
Product CodeJJX · Clinical Chemistry
Decision DateJun 17, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

For the quantitative determination of free thyroxine (FT4) in serum using the Ciba Corning Automated Chemiluminescence Systems.

Device Story

Competitive chemiluminescence immunoassay; measures free thyroxine (FT4) in serum samples. FT4 competes with acridinium ester-labeled T4 for polyclonal rabbit anti-T4 antibody coupled to paramagnetic particles. Indirect relationship between FT4 concentration and relative light units (RLUs). Used in clinical laboratory settings with Ciba Corning Automated Chemiluminescence Systems. Output provides quantitative FT4 levels to assist clinicians in evaluating thyroid status.

Clinical Evidence

Bench testing only. Correlation study (n=74) against RIA reference method showed r=0.97, equation: ACS FrT4 = 1.06(RIA) - 0.33. Sensitivity: 0.1 ng/dl (range up to 11.0 ng/dl). Precision: Total % CV 2.81-4.65%.

Technological Characteristics

Competitive chemiluminescence immunoassay. Solid phase: paramagnetic particles with covalently coupled polyclonal rabbit anti-T4 antibody. Tracer: acridinium ester-labeled T4. Analyte: free thyroxine (FT4). Platform: Ciba Corning Automated Chemiluminescence Systems.

Indications for Use

Indicated for quantitative measurement of free thyroxine (FT4) in human serum to assess thyroid function.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Submission Summary (Full Text)

{0} Company Confidential K961510 JUN 17 1996 # Summary of Safety and Effectiveness As required by 21 CFR 807.92, the following 510(k) Summary is provided: ## 1. Submitter Information Contact person: Thomas F. Flynn Address: Ciba Corning Diagnostics Corp. 63 North Street Medfield, MA 02052 Phone: 508 359-3877 e-mail: thomas.flynn@cibadiag.com Date Summary Prepared: April 19. 1996 ## 2. Device Information Proprietary Name: ACS FrT4 Immunoassay Common Name: FrT4 Immunoassay Classification Name: Class II, Free Thyroxine Test Kit, 21 CFR 862.1695 ![img-0.jpeg](img-0.jpeg) ## 3. Predicate Device Information Name: Amerlex-MAB FT4 Manufacturer: Amerlex 510(k) Number: Not Known ## 4. Device Description The Ciba Corning ASC FrT4 Immunoassay is a competitive immunoassay. FT4 in the sample competes with the acridinium ester (AE-labeled T4 (Lite Reagent) for a limited amount of polyclonal rabbit anti-T4 antibody, which is covalently coupled to paramagnetic particles (solid phase). An indirect relationship exists between the FT4 in a sample and the relative light units (RLUs) detected. ## 5. Statement of Intended Use For the quantitative determination of free thyroxine (FT4) in serum using the Ciba Corning Automated Chemiluminescence Systems. ## 6. Summary of Technological Characteristics The Ciba Corning ACS FrT4 procedure is a competitive chemiluminescence immunoassay. ACS FrT4 510(k) Notification April 19, 1996 2 {1} Company Confidential # 7. Performance Data ## Sensitivity The ACS FrT4 immunoassay measures FT4 concentrations up to 11.0 ng/dl with a minimum detectable concentration of 0.1 ng/dl. ## Accuracy For 74 samples in the range of 0.62 to 4.26 ng/dl, the correlation between the ACS FrT4 immunoassay and the RIA reference method assay is described by the equation: $$ \text{ACS FrT4} = 1.06 \, (\text{RIA}) - 0.33 $$ Correlation coefficient (r) = 0.97 ## Precision Total precision (Total % CV) ranged from 2.81 to 4.65. ACS FrT4 510(k) Notification April 19, 1996 3
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...