← Product Code [GCJ](/productcode/GCJ) · K961463

# HEARTPORT THORACIC SURGICAL TROCARS: (TOST-4-3, TOSTL-6-3, TON-4-2 AND TONL-6-2 (K961463)

_Heartport, Inc. · GCJ · Oct 8, 1996 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K961463

## Device Facts

- **Applicant:** Heartport, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Oct 8, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Heartport Thoracic Surgical Trocars are intended to provide access to the thoracic cavity through ports in the intercostal spaces, with minimal to no rib retraction. The Heartport Thoracic Trocar allows endoscopic instruments, thoracoscopes and accessory devices to be inserted and used during cardiac surgery procedures. The trocar cannula may also be used for suture organization.

## Device Story

Device provides thoracic cavity access via intercostal spaces; facilitates insertion of endoscopic instruments, thoracoscopes, and accessory devices; enables suture organization. Used in cardiac surgery; operated by surgeons. Device integrates functions of multiple predicate trocars and suture guides into single unit. Compatible with Heartport Oval Light for surgical site illumination. Benefits include minimal to no rib retraction during endoscopic procedures.

## Clinical Evidence

Bench testing (in vitro) evaluated compatibility with endoscopic instruments and suture organization. In vivo studies conducted to confirm performance. Results demonstrated device is safe, effective, and meets clinical requirements.

## Technological Characteristics

Thoracic trocar cannula and obturator system. Materials are identical to predicate devices with established biocompatibility. Designed for intercostal access. Compatible with external illumination accessory (Oval Light).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- U.S. Surgical - Auto Suture Thoracoport 11.5 mm Disposable Trocar
- Origin Medsystems (a Guidant Corp.) - Surgical Trocar Thoracic
- Ethicon (a Johnson & Johnson co.) - Endopath Thoracic Trocar Sleeve with Rounded Tip Obturator
- Snowden-Pencer - Reusable Thora-Port
- Deknatel - Gabby-Frater Suture Guide

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Heartport™ Thoracic Surgical Trocars
Appendices

# APPENDIX A. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: **K961463**
OCT - 8 1996

## Applicant Information:

Date Prepared: April 16, 1996

Name: Heartport, Inc.
Address: 200 Chesapeake Drive
Redwood City, CA 94063

Contact Person: Isabella Abati
Phone Number: (415) 306-7900
Fax Number: (415) 306-7905

## Device Information:

Classification: Class II
Trade Name: Heartport™ Thoracic Surgical Trocar
Common Name: Trocars
Classification Name: Trocars: Cardiovascular Devices - Trocar, 21 CFR 870.1390

## Equivalent Devices:

The Heartport Thoracic Surgical Trocars are substantially equivalent in intended use and/or method of operation to a combination of the following predicate devices:

1. U.S. Surgical - Auto Suture Thoracoport 11.5 mm Disposable Trocar
2. Origin Medsystems (a Guidant Corp.) - Surgical Trocar Thoracic
3. Ethicon (a Johnson &amp; Johnson co.) - Endopath Thoracic Trocar Sleeve with Rounded Tip Obturator
4. Snowden-Pencer - Reusable Thora-Port
5. Deknatel - Gabby-Frater Suture Guide

Heartport, Inc.
CONFIDENTIAL
Page 20

{1}

Heartport™ Thoracic Surgical Trocars
Appendices

# 510(k) Summary of Safety and Effectiveness (continued)

## Intended Use:

Heartport Thoracic Surgical Trocars are intended to provide access to the thoracic cavity through ports in the intercostal spaces, with minimal to no rib retraction. The Heartport Thoracic Trocar allows endoscopic instruments, thoracoscopes and accessory devices to be inserted and used during cardiac surgery procedures. The trocar cannula may also be used for suture organization.

## Comparison To Predicate Devices:

The Heartport Thoracic Surgical Trocars are equivalent in intended use and operational characteristics to the following devices: (1) U.S. Surgical - Auto Suture Thoracoport 11.5 mm Disposable Trocar, (2) Origin Medsystems (a Guidant Corp.) - Surgical Trocar Thoracic, (3) Ethicon (a Johnson &amp; Johnson co.) - Endopath Thoracic Trocar Sleeve with Rounded Tip Obturator, (4) Snowden-Pencer - Reusable Thora-Port and (5) Deknatel - Gabby-Frater Suture Guide.

The Heartport Thoracic Surgical Trocars combine the same functions as the individual predicate devices into a single device. Heartport Thoracic Surgical Trocars may be used with the Heartport Oval Light accessory device to provide illumination to the surgical site.

## Non-Clinical Test Results:

### Performance

The Heartport Thoracic Surgical Trocar was evaluated *in vitro* for performance including compatibility with endoscopic instruments and suture organization ability. In addition, the Thoracic Surgical Trocar was evaluated during *in vivo* studies. Performance testing demonstrated that the Heartport Surgical Thoracic Trocar was safe and effective, while meeting the anticipated clinical requirements for its intended use.

### Biocompatibility

The materials used to fabricate the Heartport Thoracic Surgical Trocar and Oval Light are the same as, and in most cases identical to, the predicate devices. The materials used in the Heartport Thoracic Surgical Trocar and Oval Light have proven biocompatibility.

## Summary:

Based on the intended use, product information, performance and biocompatibility data provided in this Notification, Heartport Thoracic Surgical Trocars have been shown to be substantially equivalent to currently marketed predicate devices.

Heartport, Inc.
CONFIDENTIAL
Page 21

---

**Source:** [https://fda-staging.innolitics.com/device/K961463](https://fda-staging.innolitics.com/device/K961463)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).