← Product Code [MGP](/productcode/MGP) · K961319

# EPIVIEW (K961319)

_Convatec, A Division of E.R. Squibb & Sons · MGP · Jul 8, 1996 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K961319

## Device Facts

- **Applicant:** Convatec, A Division of E.R. Squibb & Sons
- **Product Code:** [MGP](/productcode/MGP.md)
- **Decision Date:** Jul 8, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

EpiVIEW Thin Film Dressing is indicated as a securement device and dressing for peripheral and central IV catheters. It may be used as a wound dressing on minor abrasions, superficial pressure ulcers, over eschar to facilitate debridement, skin graft donor sites, closed clean surgical wounds and first and second degree burns. EpiVIEW may also be used to prevent skin breakdown from chafing or from exposure to continuous moisture and friction. EpiVIEW Thin Film Dressing is not intended for use over infected catheter sites or infected wounds, or over arterial sites. The dressing is not intended to replace sutures.

## Device Story

Thin film dressing; polyurethane film coated with acrylic adhesive; applied to skin/wounds by clinician; provides securement for IV catheters; protects wounds from external environment; facilitates debridement; prevents skin breakdown; non-invasive; passive barrier function.

## Clinical Evidence

Bench testing only. Evaluated peel adhesion (N/cm), moisture vapor transmission rate, elasticity, and coefficient of friction compared to predicate. Biocompatibility testing performed per ISO 10993 Part I (non-toxic, non-cytotoxic, non-hemolytic, negligible irritant, weak allergenic potential).

## Technological Characteristics

Thin polyurethane film with acrylic adhesive wound contact layer. Biocompatibility per ISO 10993 Part I. Passive barrier dressing.

## Predicate Devices

- 3M's Tegaderm Transparent Dressing ([K811291](/device/K811291.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) Premarket Notification
EpiVIEW™ Thin Film Dressing

K961319

JUL - 8 1996

# ITEM 8: 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

The purpose of this 510(k) Premarket Notification is to request clearance to market EpiVIEW Thin Film Dressing.

EpiVIEW Thin Film Dressing is indicated as a securement device and dressing for peripheral and central IV catheters. It may be used as a wound dressing on minor abrasions, superficial pressure ulcers, over eschar to facilitate debridement, skin graft donor sites, closed clean surgical wounds and first and second degree burns. EpiVIEW may also be used to prevent skin breakdown from chafing or from exposure to continuous moisture and friction.

EpiVIEW Thin Film Dressing is not intended for use over infected catheter sites or infected wounds, or over arterial sites. The dressing is not intended to replace sutures.

EpiVIEW Thin Film Dressing is substantially equivalent to 3M's Tegaderm Transparent Dressing, K811291. Both devices have essentially the same intended uses, contraindications, precautions and observations. EpiVIEW Thin Film Dressing is similar in construction and design to Tegaderm; both dressings consist of a thin polyurethane film coated with an acrylic adhesive wound contact layer.

Both dressings have been tested and have been shown to have the same peel adhesion (N/cm), moisture vapor transmission rate, and elasticity. The EpiVIEW dressing compared with Tegaderm is shown to have a lower coefficient of friction (COF) value, which would mean a lower incidence of shearing during human wear. Results for moisture vapor transmission rate show both dressings have substantially equivalent MVTR profiles. In addition, both dressings showed no significant differences in elasticity.

EpiVIEW Thin Film Dressing has been subjected to biocompatibility testing utilizing the ISO 10993 Part I "Biological Evaluation of Medical Devices" with FDA modified matrix (Guidance effective July 1, 1995). The results of this testing demonstrate that EpiVIEW is considered to be non-toxic, non-cytotoxic, non-hemolytic, a negligible irritant and having a weak allergenic potential. Test reports appear in Item 5 of this application.

000163

ConraTec - A Division of E.R. Squibb &amp; Sons, Inc.
EpiVIEW™ - Registered Trademark of E.R. Squibb &amp; Sons, Inc.
Tegaderm™ - Registered Trademark of 3M

---

**Source:** [https://fda-staging.innolitics.com/device/K961319](https://fda-staging.innolitics.com/device/K961319)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).