PUNCTUR-GUARD WINGED SET

K961251 · Bio-Plexus, Inc. · FMI · Jun 27, 1996 · General Hospital

Device Facts

Record IDK961251
Device NamePUNCTUR-GUARD WINGED SET
ApplicantBio-Plexus, Inc.
Product CodeFMI · General Hospital
Decision DateJun 27, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2

Indications for Use

The PUNCTUR-GUARD® Winged Set is a sterile, single use blood collection set that performs the following functions. (i) provides sterile vascular access for the collection of blood utilizing a vacuum tube system or syringe. (ii) allows vascular access with a low entry angle and firm needle tip control. (iii) permits collection of blood via a flexible tubing, effectively dissociating the puncture site from the proximal end of the device.

Device Story

Sterile, single-use blood collection set; provides vascular access for blood collection via vacuum tube or syringe. Device features sharp, lubricated stainless steel front needle with internal blunt cannula; flexible winged body; third wing actuates blunt cannula forward of needle point to cover tip. Flexible tubing connects to female luer connector; available with male luer cap or Multi-Sample Luer Adapter (MSLA). Used by clinicians in clinical settings for venipuncture; blunting mechanism reduces needle-stick risk. Output is blood sample collected into vacuum tube or syringe.

Clinical Evidence

Bench testing only; simulated use tests confirmed performance similarities to equivalent marketed products.

Technological Characteristics

Stainless steel needle; flexible winged body; blunt inner cannula mechanism; female luer connector; optional Multi-Sample Luer Adapter (MSLA). Sterile, single-use.

Indications for Use

Indicated for sterile vascular access for blood collection via vacuum tube system or syringe in patients requiring venipuncture.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} K961251 JUN 27 1996 # 510(k) SUMMARY | Proprietary Name: | PUNCTUR-GUARD® Winged Set | | --- | --- | | Classification Name: | Single Lumen Hypodermic Needle | | Regulation #: | 880.5570 | | Class: | II | | Panel Code: | 80 | | Product Code: | FMI | | Reason for submission: | new device | This document demonstrates the substantial equivalence of the PUNCTUR-GUARD® Winged Set to currently marketed Winged Blood Collection Sets, such as the Vacutainer® Brand Blood Collection Set, reorder number 7250, manufactured by Becton Dickinson, the SurFlo® Winged Infusion Set, reorder number SV*21BLK, manufactured by Terumo, and the Butterfly® Infusion Set, reorder number 4492, manufactured by Abbott Laboratories. This summary provides the following safety and effectiveness information upon which a determination of substantial equivalence is based. The PUNCTUR-GUARD® Winged Set is a sterile, single use blood collection set that performs the following functions. (i) provides sterile vascular access for the collection of blood utilizing a vacuum tube system or syringe. (ii) allows vascular access with a low entry angle and firm needle tip control. (iii) permits collection of blood via a flexible tubing, effectively dissociating the puncture site from the proximal end of the device. ## Device Description: The PUNCTUR-GUARD® Winged Set is comprised of a sharp, lubricated stainless steel front needle which is attached to a flexible winged body. A blunt inner cannula is contained inside the front needle and provides the fluid path. A third wing is provided to actuate the blunting cannula forward of the needle point. The third wing is connected to flexible tubing at the end of which is a female luer connector. The device will be available in two configurations, with a male luer cap covering the female luer connector or with a Multi-Sample Lure Adapter (MSLA) attached to the female luer connector. The MSLA is comprised of a valved needle, hub with female luer connector and threaded section for the attachment of a holder. BIO-PLEXUS manufactures a one piece Needle Holder. PAGE 6' {1} PAGE 6A' Labels and Labeling: Labeling meets the requirements of the 21 CFR Part 801 as relates to the indications for use of the device. Descriptive Comparison to Legally Marketed Devices: The comparison of the intended use, technological features and specifications (including material) of the PUNCTUR-GUARD® Winged Set to currently marketed winged blood collection sets demonstrates the device's substantial equivalence. The blunting feature is compared to the PUNCTUR-GUARD® Blood Collection Needle as they utilize a similar technology. Performance Testing: The bench tests and the simulated use tests confirmed the similarities of the PUNCTUR-GUARD® Winged Set to the equivalent currently marketed products. Accessories to the Device: The PUNCTUR-GUARD® Winged Set with a MSLA is by design compatible with PUNCTUR-GUARD® Needle Holders.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...