SJM SEGUIN ANNULOPLASTY RING MODEL SAR-M

K961246 · St. Jude Medical, Inc. · KRH · Feb 14, 1997 · Cardiovascular

Device Facts

Record IDK961246
Device NameSJM SEGUIN ANNULOPLASTY RING MODEL SAR-M
ApplicantSt. Jude Medical, Inc.
Product CodeKRH · Cardiovascular
Decision DateFeb 14, 1997
DecisionST
Submission TypeTraditional
Regulation21 CFR 870.3800
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Seguin Annuloplasty Ring is indicated for use in repair of diseased or damaged mitral heart valves that are determined by the Physician to be repairable and do not require replacement. The Seguin Annuloplasty Ring provides support to the mitral heart valve restricting expansion of the annulus.

Device Story

Semi-rigid annuloplasty ring for mitral valve repair; kidney-shaped to match mitral annulus. Fabricated from ultra-high molecular weight polyethylene (PE) core covered with Dialine polyester sewing ring; knitted polyester exterior facilitates attachment to heart annulus and promotes tissue ingrowth. Used by surgeons in cardiac procedures to restrict annular expansion. Device provides structural support to mitral valve; prevents further dilation. Sizing obturators used to select appropriate prosthesis size (26mm-40mm). Benefits include restoration of valve function and avoidance of valve replacement.

Clinical Evidence

Bench testing included computational structural analysis, compressive/tensile failure testing, and suture pull-out tests. Biocompatibility testing (USP systemic injection, implantation, Ames assay, sensitization, pyrogen, subchronic toxicity) passed. Long-term animal study (up to two years) evaluated surgical handling, in-vivo performance, tissue ingrowth, and histological outcomes; results were excellent and comparable to the predicate.

Technological Characteristics

Semi-rigid annuloplasty ring; kidney-shaped. Core: Ultra-high molecular weight polyethylene (PE). Covering: Dialine polyester fabric. Sizes: 26mm-40mm. Sizing obturators included. Biocompatibility per USP standards. Non-active, implantable device.

Indications for Use

Indicated for patients of all ages with diseased or damaged mitral valves caused by acquired or congenital processes resulting in mitral insufficiency and/or stenosis, where the valve is determined repairable by a physician.

Regulatory Classification

Identification

An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} 02/14/97 FRI 17:39 FAX 6124817673 ST JUDE MEDICAL K961246 002 # 510(k) Summary (As required by 21 CFR 807.92) FEB 14 1997 ## A. Submitter Information Submitter's Name: St. Jude Medical Division (SJMD) Address: St. Jude Medical, Inc. One Lillehei Plaza St. Paul, MN 55117 Contact Person: Jonas A. Runquist Telephone Number: (612) 483-2000 Submission Prepared: March 28, 1996 ## B. Device Information Proprietary Name: SJM® Seguin Annuloplasty Ring Model SAR-M Common or Usual Name: Annuloplasty Ring Valvuloplasty Ring Mitral Valve Support Ring Classification Name: Pre-amendment Class III CFR §870.3800 Cardiovascular Prosthetic Devices, Annuloplasty Ring Predicate Device: St. Jude Medical considers the Seguin Annuloplasty ring to be "substantially equivalent" to the Carpentier-Edwards Physio™ annuloplasty ring manufactured by Baxter Healthcare Corporation. The Physio annuloplasty ring was previously found by the Food and Drug Administration to be substantially equivalent to the Carpentier-Edwards ring which was marketed prior to May, 1976. Device Description: The Seguin Annuloplasty Ring is a semi-rigid annuloplasty ring fabricated from an ultra-high molecular weight polyethylene (PE) core surrounded by 510(K) Summary Seguin Annuloplasty Ring {1} 02/14/97 FRI 17:39 FAX 6124817673 ST JUDE MEDICAL 003 a custom Dialine® polyester sewing ring. The PE core is covered with a knitted polyester material providing a means for attaching the ring to the heart annulus as well as a suitable surface for tissue ingrowth. **Intended Use:** The Seguin Annuloplasty Ring is indicated for use in repair of diseased or damaged mitral heart valves that are determined by the Physician to be repairable and do not require replacement. The Seguin Annuloplasty Ring provides support to the mitral heart valve restricting expansion of the annulus. **C. Comparison of Required Technological Characteristics** SJM considers the Seguin Annuloplasty Ring substantially equivalent in general configuration, function and intended use to the Carpentier-Edwards Physio™ annuloplasty ring. The table below addresses equivalency characteristics for the Seguin Annuloplasty Ring and the Physio™ annuloplasty ring. Each equivalency characteristic is then further addressed. | Characteristic | Equivalency | | --- | --- | | a. Product Labeling | Substantially Equivalent | | b. Intended Use | Identical | | c. Physical Characteristics | Substantially Equivalent | | d. Anatomical Sites | Identical | | e. Target Population | Identical | | f. Performance Testing | Substantially Equivalent | | g. Safety Characteristics | Substantially Equivalent | **Product Labeling** Product labeling provided with the Seguin Annuloplasty Ring and the Physio™ annuloplasty ring are substantially equivalent as both devices provide similar Indications for Use, contraindications, warnings, precautions, etc. 510(K) Summary Seguin Annuloplasty Ring {2} 02/14/97 FRI 17:40 FAX 6124817673 ST JUDE MEDICAL 004 ## Intended Use The intended use for the Seguin Annuloplasty Ring and the Physio™ annuloplasty ring are identical. Both devices are intended for the repair of diseased or damaged mitral valves determined by the Physician to be repairable. ## Physical Characteristics The physical characteristics of the Seguin Annuloplasty ring and the Physio™ annuloplasty ring are substantially equivalent as both devices are kidney shaped to match the mitral heart annulus. Both devices utilizes a core that provides a combination of flexibility and rigidity. The Seguin Annuloplasty Ring utilizes a polymer core and the Physio™ annuloplasty ring utilizes a core of metal bands to provide the semi-flexible properties. Both devices are covered in polyester fabric providing a means to attach the device to the heart annulus. The Seguin annuloplasty ring is available in mitral sizes 26mm - 40mm and the Physio™ annuloplasty ring is available in mitral sizes 24mm - 40mm. Both devices utilize a set of sizing obturators to select the proper size of prosthesis. ## Anatomical Sites Both the Seguin Annuloplasty Ring and the Physio™ annuloplasty ring are intended for use in the mitral valve. ## Target Population The target population for the Seguin Annuloplasty Ring and the Physio™ annuloplasty ring are identical. Both devices are targeted for patients of all ages, etc. with diseased or damaged mitral valves caused by acquired congenital processes resulting in mitral insufficiency and or stenosis. ## Performance Testing Performance testing on the Seguin Annuloplasty Ring has demonstrated that the prosthesis is substantially equivalent to the Physio™ annuloplasty ring. Animal testing (and clinical evaluations) have clearly demonstrated that the Seguin Annuloplasty Ring, like the Physio™ annuloplasty ring is capable of adequately repairing diseased or damaged mitral valves. ## Safety Characteristics Safety testing on the Seguin Annuloplasty Ring has demonstrated that the prosthesis is substantially equivalent to the Physio™ annuloplasty ring. *in-vitro* testing has illustrated that the Seguin Annuloplasty Ring, like the Physio™ annuloplasty ring is capable of withstanding stresses without failure well beyond those that may be experience *in-vivo*. Biocompatibility testing has also established that the Seguin ring like the Physio™ annuloplasty ring, is biocompatible and non-toxic. 510(K) Summary Seguin Annuloplasty Ring {3} 02/14/97 FRI 17:41 FAX 6124817673 ST JUDE MEDICAL 005 ## D. Summary of Nonclinical Tests The following outlines the testing performed to demonstrate substantial equivalence to the Physio™ annuloplasty ring. ### 1. Evaluation of the mitral annulus to examine deflections of the Seguin Annuloplasty Ring based on literature references. A thorough review of the literature was performed to obtain estimates for physiological loads that may be placed on a cardiovascular prosthesis such as an annuloplasty ring *in-vivo*. The literature indicates that a worst case physiological load may be approximately 1.8 Lbs. and 200 mm Hg. The literature states that the maximum orifice reduction based on these loads is approximately 10% for a flexible annuloplasty ring, and 2.9% for a rigid annuloplasty ring. SJM utilized the 10% value for the flexible ring as the worst case condition in evaluating structural integrity of the Seguin Annuloplasty Ring. In reality, the Seguin Annuloplasty Ring like the Physio™ annuloplasty ring is a semi-rigid prosthesis and would be expected to have an orifice reduction of between 10% and 2.9% during a severe cardiac cycle. ### 2. Theoretical failure analysis a. Stress magnitudes b. Stress location The Computational Structural Analysis addressed the structural stress magnitudes and stress locations on the Seguin Annuloplasty Ring. The results of this testing demonstrated that the expected worst case physiological loads based on literature references are less than half the force required to yield the Ultra High Molecular Weight polyethylene core of the Seguin Annuloplasty Ring. This testing has demonstrated that the Seguin Annuloplasty Ring has similar safety characteristics as the Physio™ annuloplasty ring. ### 3. Physical testing a. Compressive failure mechanism based on theoretical loads and deflections Compressive testing was designed to demonstrate that the force required to reduce the area of the Seguin Annuloplasty Ring by 10% (representing extreme conditions) did not result in damage or compromise the performance of the prosthesis. Data 510(K) Summary Seguin Annuloplasty Ring {4} 02/14/97 FRI 17:41 FAX 6124817673 ST JUDE MEDICAL 006 indicates that the 26mm ring (smallest) represents worst case conditions as it requires the greatest load to reduce the orifice of the ring by 10%. Decreased loads are required to achieve the 10% orifice reduction for the larger ring sizes. ## b. Tensile failure mechanism based on theoretical loads and deflections Tensile tests were performed to illustrate that the yield strength and the ultimate tensile strength of the Seguin Annuloplasty Ring are extremely high and are well above expected *in-vivo* stresses. ## c. Suture pull out tests The suture pull out test was designed to evaluate the potential of ring dehiscence with the Seguin Annuloplasty Ring as a result of sewing ring fabric failure. These tests investigated the strength of the hand sewn seam on the sewing ring as well as the tensile strength of the fabric. The strength of the sewing ring seam and fabric were evaluated by placing sutures through the sewing cuff and performing tensile or pull tests. This testing demonstrated that the hand sewn seam is stronger than the fabric as all sutures tore the fabric before failure of the seam occurred. The study also demonstrated that the strength of the fabric far exceeds the forces expected *in-vivo*. The physical testing performed on the Seguin Annuloplasty Ring has demonstrated that it has similar safety characteristics as the Physio™ annuloplasty ring. ## 4. Biocompatibility testing The materials used for the Seguin Annuloplasty Ring have an extensive clinical history with no adverse biocompatible response. The tests performed and the results are as follows: | Test | Result | | --- | --- | | USP Systemic Injection | Pass | | USP Implantation (14 day) - Ultra High Molecular Weight Polyethylene | Pass | | USP Implantation (14 day) - Polyester Fabric | Pass | | Salmonella Mutagenicity Plate Assay (Ames Assay) | Pass | | USP Intracutaneous Test | Pass | | Guinea Pig Delayed Contact Sensitization Test | Pass | | (Extracted in Cottonseed Oil) | | | Guinea Pig Delayed Contact Sensitization Test | Pass | | (Extracted in Sodium Chloride) | | 510(K) Summary Seguin Annuloplasty Ring {5} 02/14/97 FRI 17:42 FAX 6124817673 007 USP Rabbit Pyrogen Test ... Pass Subchronic Toxicity with Histopathology (14 day) ... Pass The biocompatibility tests were conducted at ViroMed Laboratories in Minneapolis MN. and were performed under Good Laboratory Practices per 21 CFR Part 58. In addition to the biocompatibility tests performed above, physical testing, chemical testing, and animal implant tests were performed on Dialine polyester fabric by an outside expert. This testing was published in a scientific journal. (King, et. al: *Evaluating the Dialine Vascular Prosthesis knitted from an alternative source of polyester yarns* Journal of Biomedical Materials Research, Vol 29, 595-610 (1995)). The results of these tests demonstrated that the Dialine polyester fabric was essentially equivalent to Dupont based (Dacron®) polyester fabrics in terms of physical, chemical, and tissue response characteristics. 5. Long term animal testing a. Surgical handling characteristics b. *in-vivo* performance - Tissue in-growth properties - Macroscopic evaluations - Histological examinations Animal testing on the Seguin Annuloplasty Ring was designed to evaluate the performance of the prosthesis *in-vivo* for up to two years. Animals were monitored for evidence of excessive stenosis, regurgitation or other complications. Following explant, the annuloplasty rings were subjected to complete macroscopic, microscopic and histological examinations to access the prosthesis for evidence structural failure and to evaluate healing characteristics. All results from the animal study were considered excellent. The animal testing has demonstrated that the performance of the Seguin Annuloplasty Ring is similar to the Physio™ annuloplasty ring. 510(K) Summary Seguin Annuloplasty Ring
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...