STARSWAB CULTURE COLLECTION AND TRANSPORT SYSTEM

K960997 · Starplex Scientific, Inc. · LIO · Apr 3, 1996 · Microbiology

Device Facts

Record IDK960997
Device NameSTARSWAB CULTURE COLLECTION AND TRANSPORT SYSTEM
ApplicantStarplex Scientific, Inc.
Product CodeLIO · Microbiology
Decision DateApr 3, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2900
Device ClassClass 1

Indications for Use

The Starswab Culture Collection and Transport System is self-contained, sterile and intended for maintaining the viability of clinical specimens during transportation to the laboratory.

Device Story

Self-contained, sterile specimen collection and transport system; utilizes medium containing sodium thioglycollate as reducing agent; maintains moisture and media depth to support survival of aerobic and anaerobic microorganisms during transport to laboratory; used by clinical staff for specimen collection; ensures viability of microorganisms for subsequent laboratory culture and analysis; benefits patient by preserving specimen integrity for accurate diagnostic testing.

Clinical Evidence

Bench testing only. Comparative study of Starswab vs. Difco transport media using known inocula of ATCC type cultures and clinically significant microorganisms. Samples stored at room temperature for 24 and 48 hours, then subcultured to assess growth characteristics. Results demonstrated equivalent recovery between subject and predicate devices.

Technological Characteristics

Sterile, self-contained transport system. Medium contains sodium thioglycollate as a reducing agent. Designed for aerobic and anaerobic microorganism survival via moisture and media depth control.

Indications for Use

Indicated for the collection and transport of clinical specimens to maintain the viability of aerobic and anaerobic microorganisms.

Regulatory Classification

Identification

A microbiological specimen collection and transport device is a specimen collecting chamber intended for medical purposes to preserve the viability or integrity of microorganisms in specimens during storage of specimens after their collection and during their transport from the collecting area to the laboratory. The device may be labeled or otherwise represented as sterile. The device aids in the diagnosis of disease caused by pathogenic microorganisms.

Predicate Devices

Submission Summary (Full Text)

{0} K960997 # SUMMARY OF SAFETY AND EFFECTIVENESS ## 1. CONTACT NAME AND ADDRESS APR - 3 1996 MEHDI KARAMCHI 50 STEINWAY BLVD. ETOBICOKE, ONTARIO, CANADA M9W 6Y3 TELEPHONE: (416) 674-7474 ## 2. DEVICE NAME Classification Name: Transport System, Aerobic/Anaerobic Common Name: Sterile Microbiological Specimen Collection and Transport System Proprietary Name: Starswab Culture Collection and Transport System ## 3. PREDICATE DEVICE Amies CultureSwab™ Transport Systems with and without charcoal are marketed by Difco Laboratories, P.O. Box 331058, Detroit MI, 48232-7058, U.S.A. ## 4. DEVICE DESCRIPTION The Starswab Culture Collection and Transport System is self-contained, sterile and intended for maintaining the viability of clinical specimens during transportation to the laboratory. The medium contains sodium thioglycollate as a reducing agent and the moisture and media depth provide an ideal environment for the survival of both aerobic and anaerobic microorganisms. ## 5. SUBSTANTIAL EQUIVALENCE DETERMINATION The Starswab Culture Collection and Transport System is similar in design, manufacturing and intended usage to Amies Medium and Amies Medium without charcoal currently manufactured and marketed by Difco Laboratories. This substantial equivalence claim is supported by manufacturing, medium formulation, packaging and clinical evaluations. Known inocula of ATCC type cultures and clinically significant microorganisms were inoculated in the Starswab and Difco transport media and stored at room temperature for 24 and 48 hours prior to subculturing onto an appropriate culture media then checked for growth characteristics. No significant differences in recovery were detected between samples obtained from Starswab vs. Difco transport medium and the results were judged to be equivalent.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...