← Product Code [KPS](/productcode/KPS) · K960948

# SCATTER CORRECTION VOLUME SOFTWARE MODEL NSSC-020A (K960948)

_Toshiba Medical Systems · KPS · Dec 23, 1996 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K960948

## Device Facts

- **Applicant:** Toshiba Medical Systems
- **Product Code:** [KPS](/productcode/KPS.md)
- **Decision Date:** Dec 23, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

This program can be utilized for all traditional and accepted gamma camera applications. It is expected that its benefits will be realized primarily when the system is being used to acquire tomographic studies. The intended use of the gamma camera system is unchanged by this program. Toshiba believes that users will find the T.E.W method to be beneficial when performing dual-nuclide acquisitions, by enhancing the system's ability to detect the energy peaks of each nuclide without interference from the other nuclide. This determination is left to the user, however, and no specific claims regarding dual nuclide uses are being made. Due to the absence of new intended uses, it follows that the overall safety and effectiveness of the gamma camera system is not changed by this program.

## Device Story

Optional software for Toshiba DI-series gamma cameras; performs Compton-scatter correction on acquired image data. Uses Triple Energy Window (T.E.W.) method to calculate scatter location and intensity per pixel; subtracts scatter from patient data. Operated by clinicians/technicians in clinical settings. Enhances image quality by reducing scatter interference, particularly useful in dual-nuclide acquisitions. Output is processed tomographic image data; assists clinicians in diagnostic interpretation by improving signal clarity.

## Clinical Evidence

No clinical data provided; relies on established scientific principles of Compton-scatter correction documented in peer-reviewed literature.

## Technological Characteristics

Computational software module; implements Triple Energy Window (T.E.W.) algorithm for scatter correction. No physical hardware changes to host gamma camera. Standalone software application.

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- Toshiba GCA-7200A ([K931297](/device/K931297.md)/A)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

DEC 23 1996
K960948

510(k) Summary

Submitted by: Javad Seyedzadeh, Director of Regulatory Affairs
Toshiba America Medical Systems, Inc.
2441 Michelle Drive
Tustin, CA 92681
Tel.: 714-730-5000

Contact: Mark Stauffer, Regulatory Affairs Specialist
Toshiba America Medical Systems, Inc.
Tel.: 714-730-5000

Date: March 6, 1996

- Device Name
Scatter Correction Volume Software for Toshiba Gamma Cameras, Model NSSC-020A
Common name: Gamma Camera Processing Software
Classification name: Emission Computed Tomography System

- Predicate Device
Toshiba GCA-7200A (K931297/A)

- Description of Device
This product is an optional software program available for the Toshiba DI-series devices. It is intended to give a user the ability to acquire scatter data during acquisitions, then remove the unwanted scatter data from the processed image data. Based on the Triple Energy Window method of Compton - Scatter Correction, this program calculates the location and amount of scatter data in each pixel and subtracts it from the patient data.

- Intended Uses
This program can be utilized for all traditional and accepted gamma camera applications. It is expected that its benefits will be realized primarily when the system is being used to acquire tomographic studies. The intended use of the gamma camera system is unchanged by this program. Toshiba believes that users will find the T.E.W method to be beneficial when performing dual-nuclide acquisitions, by enhancing the system's ability to detect the energy peaks of each

{1}

nuclide without interference from the other nuclide. This determination is left to the user, however, and no specific claims regarding dual nuclide uses are being made. Due to the absence of new intended uses, it follows that the overall safety and effectiveness of the gamma camera system is not changed by this program.

- Comparison of technological characteristics

There is no change to the technology of the host device when the optional program Scatter Correction Volume, or T.E.W., is applied. This product is computational only and does not affect the physical characteristics of the host device. Determination of and correction for Compton-scatter effects are well understood and documented in peer-reviewed scientific publications.

---

**Source:** [https://fda-staging.innolitics.com/device/K960948](https://fda-staging.innolitics.com/device/K960948)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
