KARL STORZ INST. FOR ENDOSCOPIC SELECTIVE PERF. VEIN DISCISSION & PARATIBIAL FASCIOTOMY

K960903 · KARL STORZ Endoscopy-America, Inc. · GCJ · Aug 22, 1996 · Gastroenterology, Urology

Device Facts

Record IDK960903
Device NameKARL STORZ INST. FOR ENDOSCOPIC SELECTIVE PERF. VEIN DISCISSION & PARATIBIAL FASCIOTOMY
ApplicantKARL STORZ Endoscopy-America, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateAug 22, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The KSEA Instrument Set for Endoscopic Surgery of Superficial Veins and Fascia of the Lower Extremities is intended for use by qualified surgeons during endoscopic surgery of superficial veins and fascia of the lower extremities, including vein discission and paratibial fasciotomy.

Device Story

Manually operated, reusable surgical instrument set; includes rigid endoscope, sheath, scissors, forceps, knives, and adapters. Used by surgeons in clinical settings for endoscopic access to lower extremity veins and fascia. Instruments facilitate visualization and mechanical tissue manipulation (cutting, dissecting, grasping) during vein discission and paratibial fasciotomy. Provides direct visualization of surgical site to assist surgeon in performing minimally invasive procedures; improves surgical access compared to open techniques.

Clinical Evidence

Bench testing only; no clinical data provided. Safety established via history of use of body contact materials.

Technological Characteristics

Manually operated, reusable surgical instruments. Materials have established history of medical use. Rigid endoscope, sheath, scissors, forceps, knives, and adapters. Non-powered, mechanical operation.

Indications for Use

Indicated for qualified surgeons performing endoscopic surgery on superficial veins and fascia of the lower extremities, specifically for vein discission and paratibial fasciotomy.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Submission Summary (Full Text)

{0} Karl Storz Endoscopy-America, Inc. 600 Corporate Pointe Culver City, California 90230-7600 Phone 310 558 1500 Toll Free 800 421 0837 Fax 310 410 5527 K960903 AUG 22 1996 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document are accurate and complete to the best of KSEA’s knowledge. **Applicant:** Karl Storz Endoscopy - America, Inc. 600 Corporate Pointe Culver City, CA 90230 (310) 558-1500 **Contact:** Marika Anderson Regulatory Affairs Specialist **Device Identification:** Common Name Endoscope Telescope sheaths Knives Scissors Dissectors Bipolar Forceps and Cable Adapter Trade Name KSEA Instrument Set for Endoscopic Surgery of Superficial Veins and Fascia of the Lower Extremities **Indication:** The KSEA Instrument Set for Endoscopic Surgery of Superficial Veins and Fascia of the Lower Extremities is intended for use by qualified surgeons during endoscopic surgery of superficial veins and fascia of the lower extremities, including vein discission and paratibial fasciotomy. **Device Description:** The KSEA Instrument Set for Endoscopic Surgery of Superficial Veins and Fascia of the Lower Extremities are manually operated, reusable surgical devices consisting of a rigid endoscope, sheath, scissors, forceps, knives, and adapters. The instruments and endoscope are long enough to gain access to the surgical site. The body contact materials have a long history of use in medical devices and do not pose any new issues of safety and effectiveness. **Substantial Equivalence:** The KSEA Instrument Set for Endoscopic Surgery of Superficial Veins and Fascia of the Lower Extremities is substantially equivalent to the predicate devices since the basic features, design and intended uses are the same. The minor differences in design and dimensions between the KSEA Instrument Set for Endoscopic Surgery of Superficial Veins and Fascia of the Lower Extremities and the predicate devices raise no new issues of safety and effectiveness, as these design differences have no effect on the performance, function or intended use of these devices. Signed: Marika Anderson Regulatory Affairs Specialist 074
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...