PHACOSCOPE

K960811 · Saint John'S Plaza Medical Group · HQC · Oct 2, 1996 · Ophthalmic

Device Facts

Record IDK960811
Device NamePHACOSCOPE
ApplicantSaint John'S Plaza Medical Group
Product CodeHQC · Ophthalmic
Decision DateOct 2, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4670
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Provide illumination at surgical site.

Device Story

Phacoscope™ combines phacofragmentation handpiece with integrated fiber optic light guides; guides routed through fluid channels and infusion sleeve. Device provides illumination at surgical site during ophthalmic procedures. Operates using existing electrical power and light sources identical to predicate phacofragmentation systems. Used by ophthalmologists in surgical settings. Benefits include simultaneous fragmentation and illumination, reducing need for separate light pipe instruments.

Clinical Evidence

Bench testing only; no clinical data provided. Equivalence based on identical materials and design to legally marketed predicate devices.

Technological Characteristics

Phacofragmentation handpiece with integrated quartz or plastic fiber optic light guides. Materials identical to legally marketed phacofragmentation and fiber optic devices. Sterilization via autoclave or standard fiber optic instrument sterilizers.

Indications for Use

Indicated for use during ophthalmic surgery requiring phacofragmentation and intraocular illumination.

Regulatory Classification

Identification

A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.

Predicate Devices

Submission Summary (Full Text)

{0} MICHAEL REYNARD, M.D. Saint John's Plaza Eye Medical Group 1301 - 20th Street #260 Santa Monica, California 90404 (310) 453-0551 OCT - 2 1996 K960811 # 510(k) Summary per section 807.92(c) Submitters Name: Michael Reynard, M.D. Address: 1301 - 20th Street #260, Santa Monica, CA 90404 Telephone: 310-453-0551 FAX: 310-315-0133 Contact Person: Michael Reynard, M.D. Date of Summary: 02/12/1996 **Name of Device:** Fiber Optic Phacofragmentation System **Tradename of Device:** Phacoscope™ **Equivalence:** The subject device combines predicate phacofragmentation system and fiber optic light guides for transmitting visible light. **Description of Device:** Phacofragmentation system containing fiber optic light guides directed through fluid channels, and through infusion sleeve. **Intended Use of Device:** Provide illumination at surgical site. **Comparison with predicate devices:** Medical grade fiber optic guides consistent with Class II classification devices used for intraocular illumination (Classification number 876-1500; Product Code 78FCW) are combined with a phacofragmentation handpiece. Electrical power and light sources for the fiber optic phacofragmentation handpiece (Class II; Classification number 886.4670; Product Code 86HQC) are identical with those currently used. Construction materials are identical with phacofragmentation handpieces and fiber optic devices that are legally marketed devices currently in use. Colors are identical to those in legally marketed devices currently in use. **Sterilization Method:** 1) Autoclave for fiber optic phacofragmentation handpiece incorporating quartz fiber optics, or 2) sterilizers customarily used for fiber optic instruments incorporating quartz or plastic fiber optics. RECEIVED 02/12/1996 9661 811
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...